MINIMUM DETECTABLE DIFFERENCE OF CT ANGIOGRAPHY SCANS AT VARIOUS CARDIAC BEATS: EVALUATION VIA A CUSTOMIZED OBLIQUE V-SHAPED LINE GAUGE AND PMMA PHANTOM

被引:7
作者
Chiang, Chao-Yu [1 ,2 ]
Chen, Yi-Hua [1 ]
Pan, Lung-Fa [1 ,3 ]
Cho, Chien-Chou [1 ,4 ]
Peng, Bing-Ru [1 ,5 ]
Pan, Lung-Kwang [1 ]
机构
[1] Cent Taiwan Univ Sci & Technol, Dept Med Imaging & Radiol Sci, Taichung 406, Taiwan
[2] Taichung Armed Force Gen Hosp, Procurement Team, Taichung, Taiwan
[3] Taichung Armed Forces Gen Hosp, Dept Cardiol, Taichung 411, Taiwan
[4] Changhua Christian Med Fdn Yuanlin Christian Hosp, Dept Radiol, Changhua 500, Taiwan
[5] Taichung Armed Forces Gen Hosp, Dept Radiol, Taichung 411, Taiwan
关键词
Minimum detectable difference; CT angiography; beats/min; PMMA phantom; figure of merit; TAGUCHI ANALYSIS; QUALITY CHARACTERISTICS; IMAGING QUALITY; OPTIMIZATION;
D O I
10.1142/S0219519421400662
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
The minimum detectable difference (MDD) at various beats/min (BPM) of CT angiography (CTA) was evaluated using an oblique V-shape line gauge and poly methyl methacrylate (PMMA) phantom in this study. The customized phantom with the size of 30 x 30 x 30 cm(3) was made from a 1 cm-thick PMMA. The reciprocating mechanism in the phantom was run by a step motor with an eccentric gear connected to a crank rod to provide a stable harmonic motion, simulating the cardiac beats. The MDD has a unique feature in defining the quality characteristic of CT-scanned images and provides more information than simple line pair/cm in the previous studies. The derived MDD was quantified according to various BPM, and the CTA factor combination was preset following either the conventional recommendation or the optimal one. In doing so, the performance was substantiated by the Taguchi-based signal-to-noise ratio and integrated by another index, namely, figure of merit (FOM). The MDD and corresponding eta (dB) changed from 1.70 +/- 0.01mm to 2.66 +/- 0.01mm and from 16.7dB to 14.2dB, respectively, for conventional settings; while those obtained for the optimal preset changed from 1.71 +/- 0.03mm to 2.58 +/- 0.02mm and from 12.2 dB to 16.4 dB, respectively of CTA at 0-90 BPM. The integrated FOM values for conventional or optimal cases were 1240 and 1337, respectively. The MDD proved to be a useful technique in justifying the CTA-scanned images. For compliance with previous studies, MDD results can be converted to the line pair/cm results, but it is more informative than the quantized number of line pairs.
引用
收藏
页数:13
相关论文
共 19 条
[1]  
Chang CF, 2019, HELL J NUCL MED, V22, P49, DOI 10.1967/s002449910959
[2]   Optimizing the TLD-100 readout system for various radiotherapy beam doses using the Taguchi methodology [J].
Chen, Chien Yi ;
Liu, Kuo Chi ;
Chen, Hsin Hsiung ;
Pan, Lung Kwang .
APPLIED RADIATION AND ISOTOPES, 2010, 68 (03) :481-488
[3]   Optimizing quality of digital mammographic imaging using Taguchi analysis with an ACR accreditation phantom [J].
Chen, Ching-Yuan ;
Pan, Lung-Fa ;
Chiang, Fu-Tsai ;
Yeh, Da-Ming ;
Pan, Lung-Kwang .
BIOENGINEERED, 2016, 7 (04) :226-234
[4]  
Chiang FT, 2016, HELL J NUCL MED, V19, P118, DOI 10.1967/s002449910364
[5]   Optimization of Imaging Parameters for SPECT scans of [99mTc] TRODAT-1 Using Taguchi Analysis [J].
Huang, Cheng-Kai ;
Wu, Jay ;
Cheng, Kai-Yuan ;
Pan, Lung-Kwang .
PLOS ONE, 2015, 10 (03)
[6]  
Keevil S, 2014, CT SCANNERS CORONARY
[7]   The statistical analysis of data from small groups [J].
Kenny, DA ;
Mannetti, L ;
Pierro, A ;
Livi, S ;
Kashy, DA .
JOURNAL OF PERSONALITY AND SOCIAL PSYCHOLOGY, 2002, 83 (01) :126-137
[8]   The Optimization of the Single Photon Emission Computed Tomography Image Quality via Taguchi Analysis: A Feasibility Study of a V -Shaped Phantom [J].
Kittipayak, Samrit ;
Pan, Lung-Fa ;
Chiang, Fu-Tsai ;
Pan, Lung-Kwang ;
Lin, Cheng-Hsun .
JOURNAL OF MEDICAL IMAGING AND HEALTH INFORMATICS, 2017, 7 (01) :143-148
[9]   Enhancement of Thermoelectric Figure-of-Merit by a Bulk Nanostructuring Approach [J].
Lan, Yucheng ;
Minnich, Austin Jerome ;
Chen, Gang ;
Ren, Zhifeng .
ADVANCED FUNCTIONAL MATERIALS, 2010, 20 (03) :357-376
[10]   INTEGRATION OF TAGUCHI ANALYSIS WITH PHANTOM AND INNOVATIVE GAUGES: OPTIMIZATION OF THE CT SCAN PROTOCOL FOR PERIPHERAL ARTERIAL OCCLUSIVE DISEASE (PAOD) SYNDROME [J].
Lee, Tsung-Min ;
Lin, Chien-Chung ;
Peng, Bing-Ru ;
Pan, Lung-Fa ;
Pan, Lung-Kwang .
JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2020, 20 (09)