Effect of sterilization on cyclic fatigue of rotary nickel-titanium endodontic instruments

被引:58
|
作者
Mize, SB [1 ]
Clement, DJ [1 ]
Pruett, JP [1 ]
Carnes, DL [1 ]
机构
[1] Univ Texas, Hlth Sci Ctr, Sch Dent, Dept Endodont, San Antonio, TX 78284 USA
关键词
D O I
10.1016/S0099-2399(98)80015-0
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The ability of heat treatment as a result of autoclave sterilization to extend the life of nickel-titanium rotary endodontic instruments by reducing the effect of cyclic fatigue was evaluated using 280 size 40 Lightspeed instruments. Instruments were cycled in artificial canals with angles of curvature of 30 degrees and either 2 or 5 mm radii of curvature. In a pilot study, instruments were sterilized or not sterilized and cycled to failure to obtain mean cycles-to-failure values for each group. In the first experimental protocol, instruments were cycled to either 25%, 50%, or 75% of the mean cycles-to-failure limit determined in the pilot study, then sterilized or not sterilized before being cycled to failure. In the second experimental protocol, instruments were cycled to 25% of the mean cycles-to-failure determined in the pilot study, and sterilized or not sterilized. This sequence of cycling to 25% of the predetermined cycles-to-failure limit followed by sterilization was repeated until the instruments failed. No significant increases in cycles to failure were observed between groups for either experimental protocol when instruments were evaluated at a similar radius. Significant differences in cycles to failure were only observed when instruments cycled to failure in the artificial canal with 2 mm radius were compared with instruments cycled to failure in the artificial canal of 5 mm radius. Scanning electron microscopic photos showed crack initiation and propagation in all instruments that were cycled to a percentage of the predetermined cycles-to-failure limit. It is concluded that heat treatment as a result of autoclave sterilization does not extend the useful life of nickel-titanium instruments.
引用
收藏
页码:843 / 847
页数:5
相关论文
共 50 条
  • [31] Cyclic fatigue of endodontic nickel titanium rotary instruments: Static and dynamic tests
    Li, UM
    Lee, BS
    Shih, CT
    Lan, WH
    Lin, CP
    JOURNAL OF ENDODONTICS, 2002, 28 (06) : 448 - 451
  • [32] Fatigue and mechanical properties of nickel-titanium endodontic instruments
    Kuhn, G
    Jordan, L
    JOURNAL OF ENDODONTICS, 2002, 28 (10) : 716 - 720
  • [33] Fracture resistance of electropolished rotary nickel-titanium endodontic instruments
    Anderson, Margot E.
    Price, John W. H.
    Parashos, Peter
    JOURNAL OF ENDODONTICS, 2007, 33 (10) : 1212 - 1216
  • [34] CHARACTERIZATION OF NEW NICKEL-TITANIUM WIRE FOR ROTARY ENDODONTIC INSTRUMENTS
    Brantley, William A.
    Liu, Jie
    Clark, William A. T.
    Kovarik, Libor
    Buie, Caesar
    Iijima, Masahiro
    Alapati, Satish B.
    Ben Johnson, William
    ADVANCES IN BIOMEDICAL AND BIOMIMETIC MATERIALS, 2009, 206 : 49 - +
  • [35] The corrosion of nickel-titanium rotary endodontic instruments in sodium hypochlorite
    Novoa, X. R.
    Martin-Biedma, B.
    Varela-Patino, P.
    Collazo, A.
    Macias-Luaces, A.
    Cantatore, G.
    Perez, M. C.
    Magan-Munoz, F.
    INTERNATIONAL ENDODONTIC JOURNAL, 2007, 40 (01) : 36 - 44
  • [36] Bending properties of nickel-titanium rotary endodontic instruments.
    Lausten, L
    Luebke, N
    Brantley, W
    JOURNAL OF DENTAL RESEARCH, 1996, 75 : 2935 - 2935
  • [37] Cyclic fatigue resistance of two nickel-titanium rotary instruments in interrupted rotation
    Pedulla, E.
    Lizio, A.
    Scibilia, M.
    Grande, N. M.
    Plotino, G.
    Boninelli, S.
    Rapisarda, E.
    Lo Giudice, G.
    INTERNATIONAL ENDODONTIC JOURNAL, 2017, 50 (02) : 194 - 201
  • [38] Effect of different axial speed patterns on cyclic fatigue resistance of rotary nickel-titanium instruments
    Thu, Myint
    Ebihara, Arata
    Maki, Keiichiro
    Nishijo, Miki
    Kimura, Shunsuke
    Nakatsukasa, Taro
    Kyaw, Moe Sandar
    Okiji, Takashi
    BMC ORAL HEALTH, 2022, 22 (01)
  • [39] Response: Cyclic fatigue resistance of ProTaper Next nickel-titanium rotary instruments
    Elnaghy, A. M.
    INTERNATIONAL ENDODONTIC JOURNAL, 2015, 48 (11) : 1101 - 1101
  • [40] Effect of different axial speed patterns on cyclic fatigue resistance of rotary nickel-titanium instruments
    Myint Thu
    Arata Ebihara
    Keiichiro Maki
    Miki Nishijo
    Shunsuke Kimura
    Taro Nakatsukasa
    Moe Sandar Kyaw
    Takashi Okiji
    BMC Oral Health, 22