Porous cups with and without hydroxylapatite-tricalcium phosphate coating -: 23 matched pairs evaluated with radiostereometry

被引:36
作者
Thanner, J [1 ]
Kärrholm, J [1 ]
Herberts, P [1 ]
Malchau, H [1 ]
机构
[1] Sahlgrens Univ Hosp, Dept Orthopaed, S-41345 Gothenburg, Sweden
关键词
acetabular component; porous coating; hydroxyapatite/tricalcium;
D O I
10.1016/S0883-5403(99)90050-5
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Migration wear, and presence of radiolucencies were studied in 23 marched pairs of patients operated with porous-coated acetabular cups with additional screw fixation. All implants had the same type of titanium fiber mesh. In each pair, one of the cups was plasma-sprayed with a coating consisting of 70% hydroxylapatite (HA) and 30% tricalcium phosphate (TCP). Radiostereometric analysis up to 2 years after the operation revealed smaller rotations around the horizontal axis in cups with HA/TCP coating. The migration of the cup center was not significantly influenced. Evaluation of femoral head penetration in 12 of the matched pairs did not reveal any significant difference. Immediately after operation, implants with HA/TCP coating had more central radiolucencies, which, despite minimal migration, disappeared during the follow-up. The clinical results did not differ between the 2 groups. The findings of less tilting and diminishing radiolucencies in the cups with HA/TCP coating suggest a more complete ingrowth of bone and a better sealing of the interface.
引用
收藏
页码:266 / 271
页数:6
相关论文
共 31 条
[1]  
Berger RA, 1996, CLIN ORTHOP RELAT R, P134
[2]  
BLOEBAUM RD, 1993, J ARTHROPLASTY, V8, P203, DOI 10.1016/S0883-5403(09)80014-4
[3]  
Collier J P, 1993, J Arthroplasty, V8, P389, DOI 10.1016/S0883-5403(06)80037-9
[4]   Modeling of bone around hydroxyapatite-coated femoral stems [J].
DAntonio, JA ;
Township, M ;
Capello, WN ;
Manley, MT .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1996, 78A (08) :1226-1234
[5]   Effects of hydroxyapatite tricalcium phosphate coating and intracancellous placement on bone ingrowth in titanium fibermetal implants [J].
Dean, JC ;
Tisdel, CL ;
Goldberg, VM ;
Parr, J ;
Davy, D ;
Stevenson, S .
JOURNAL OF ARTHROPLASTY, 1995, 10 (06) :830-838
[6]  
DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
[7]   6-YEAR RESULTS OF HYDROXYAPATITE-COATED TOTAL HIP-REPLACEMENT [J].
GEESINK, RGT ;
HOEFNAGELS, NHM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1995, 77B (04) :534-547
[8]  
Incavo S J, 1993, J Arthroplasty, V8, P573, DOI 10.1016/0883-5403(93)90002-L
[9]   Cementless hemispheric acetabular components - A 4- to 8-year follow-up report [J].
Incavo, SJ ;
Ames, SE ;
DiFazio, FA ;
Howe, JG .
JOURNAL OF ARTHROPLASTY, 1996, 11 (03) :298-303
[10]  
JARCHO M, 1981, CLIN ORTHOP RELAT R, P259