Feasibility of Using a Factory-Calibrated Continuous Glucose Monitoring System to Diagnose Type 2 Diabetes

被引:4
作者
Frank, Spencer [1 ,2 ]
Hames, Kazanna C. [1 ]
Jbaily, Abdulrahman [1 ]
Park, Jee Hye [1 ]
Stroyeck, Chuck [1 ]
Price, David [1 ]
机构
[1] Dexcom Inc, R&D Dept, San Diego, CA USA
[2] Dexcom Inc, 6340 Sequence Dr, San Diego, CA 92121 USA
关键词
Continuous glucose monitoring; Type; 2; diabetes; Glucose variability; Diagnostic tests; TOLERANCE TEST; GLYCEMIC VARIABILITY; REPRODUCIBILITY; QUALITY; HBA(1C);
D O I
10.1089/dia.2022.0189
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Context: Plasma glucose or A1C criteria can be used to establish the diagnosis of type 2 diabetes (T2D).Objective: We examined whether continuous glucose monitoring (CGM) data from a single 10-day wear period could form the basis of an alternative diagnostic test for T2D.Design: We developed a binary classification diagnostic CGM (dCGM) algorithm using a dataset of 716 individual CGM sensor sessions from 563 participants with associated A1C measurements from seven clinical trials. Data from 470 participants were used for training and 93 participants for testing (49 normoglycemic [A1C <5.7%], 27 prediabetes, and 17 T2D [A1C >= 6.5%] not using pharmacotherapy). dCGM performance was evaluated against the accompanying A1C measurement, which was assumed to provide the correct diagnosis.Results: The dCGM algorithm's overall sensitivity, specificity, positive predictive value, and negative predictive value were 71%, 93%, 71%, and 93%, respectively. At other clinically relevant A1C thresholds, dCGM specificity among normoglycemic participants was 98% (48/49 correctly classified), and for participants with suboptimally controlled diabetes (A1C >= 7%, above the American Diabetes Association recommended A1C goal) the sensitivity was 100% (8/8 participants correctly diagnosed with T2D).Conclusions: Classifications based on the dCGM algorithm were in good agreement with traditional methods based on A1C. The dCGM algorithm may provide an alternative method for screening and diagnosing T2D, and warrants further investigation.
引用
收藏
页码:907 / 914
页数:8
相关论文
共 23 条
[1]  
Acciaroli Giada, 2018, J Diabetes Sci Technol, V12, P105, DOI 10.1177/1932296817710478
[3]   Validation of Time in Range as an Outcome Measure for Diabetes Clinical Trials [J].
Beck, Roy W. ;
Bergenstal, Richard M. ;
Riddlesworth, Tonya D. ;
Kollman, Craig ;
Li, Zhaomian ;
Brown, Adam S. ;
Close, Kelly L. .
DIABETES CARE, 2019, 42 (03) :400-405
[4]   Continuous Glucose Monitoring Versus Usual Care in Patients With Type 2 Diabetes Receiving Multiple Daily Insulin Injections A Randomized Trial [J].
Beck, Roy W. ;
Riddlesworth, Tonya D. ;
Ruedy, Katrina ;
Ahmann, Andrew ;
Haller, Stacie ;
Kruger, Davida ;
McGill, Janet B. ;
Polonsky, William ;
Price, David ;
Aronoff, Stephen ;
Aronson, Ronnie ;
Toschi, Elena ;
Kollman, Craig ;
Bergenstal, Richard .
ANNALS OF INTERNAL MEDICINE, 2017, 167 (06) :365-+
[5]   The Fallacy of Average: How Using HbA1c Alone to Assess Glycemic Control Can Be Misleading [J].
Beck, Roy W. ;
Connor, Crystal G. ;
Mullen, Deborah M. ;
Wesley, David M. ;
Bergenstal, Richard M. .
DIABETES CARE, 2017, 40 (08) :994-999
[6]   REPRODUCIBILITY OF THE ORAL GLUCOSE-TOLERANCE TEST IN PREGNANT-WOMEN [J].
CATALANO, PM ;
AVALLONE, DA ;
DRAGO, NM ;
AMINI, SB .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1993, 169 (04) :874-881
[7]  
Centers for Disease Control and Prevention, National Health and Nutrition Examination Survey: Questionnaires, Datasets, and Related Documentation
[8]   Including Continuous Glucose Monitoring to Provide Personalized Glycemic Profiles as Part of a Pilot Worksite Health Screening [J].
Crawford, Margaret ;
Johnson, Matthew ;
Klein, Ilene J. ;
Hames, Kazanna C. ;
Norman, Gregory J. .
JOURNAL OF DIABETES SCIENCE AND TECHNOLOGY, 2021, 15 (02) :515-516
[9]   Variation of Interstitial Glucose Measurements Assessed by Continuous Glucose Monitors in Healthy, Nondiabetic Individuals [J].
Fox, Larry A. ;
Beck, Roy W. ;
Xing, Dongyuan ;
Chase, H. Peter ;
Gilliam, Lisa K. ;
Hirsch, Irl ;
Kollman, Craig ;
Laffel, Lori ;
Lee, Joyce ;
Ruedy, Katrina J. ;
Tamborlane, William V. ;
Tansey, Michael ;
Wilson, Darrell M. .
DIABETES CARE, 2010, 33 (06) :1297-1299
[10]   REPRODUCIBILITY OF THE ORAL GLUCOSE-TOLERANCE TEST IN PREGNANCY [J].
HARLASS, FE ;
BRADY, K ;
READ, JA .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1991, 164 (02) :564-568