An adaptive regularization parameter choice strategy for multispectral bioluminescence tomography

被引:19
作者
Feng, Jinchao [3 ]
Qin, Chenghu [1 ]
Jia, Kebin [3 ]
Han, Dong [1 ]
Liu, Kai [1 ]
Zhu, Shouping [1 ]
Yang, Xin [1 ]
Tian, Jie [1 ,2 ]
机构
[1] Chinese Acad Sci, Inst Automat, Med Image Proc Grp, Beijing 100190, Peoples R China
[2] Xidian Univ, Sch Life Sci & Technol, Xian 710071, Peoples R China
[3] Beijing Univ Technol, Coll Elect Informat & Control Engn, Beijing 100124, Peoples R China
基金
中国国家自然科学基金; 北京市自然科学基金;
关键词
bioluminescence tomography (BLT); reconstruction algorithm; regularization; model function; DIFFUSE OPTICAL TOMOGRAPHY; ILL-POSED PROBLEMS; RECONSTRUCTION; MOUSE; PROJECTION; SELECTION;
D O I
10.1118/1.3635221
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Bioluminescence tomography (BLT) provides an effective tool for monitoring physiological and pathological activities in vivo. However, the measured data in bioluminescence imaging are corrupted by noise. Therefore, regularization methods are commonly used to find a regularized solution. Nevertheless, for the quality of the reconstructed bioluminescent source obtained by regularization methods, the choice of the regularization parameters is crucial. To date, the selection of regularization parameters remains challenging. With regards to the above problems, the authors proposed a BLT reconstruction algorithm with an adaptive parameter choice rule. Methods: The proposed reconstruction algorithm uses a diffusion equation for modeling the bioluminescent photon transport. The diffusion equation is solved with a finite element method. Computed tomography (CT) images provide anatomical information regarding the geometry of the small animal and its internal organs. To reduce the ill-posedness of BLT, spectral information and the optimal permissible source region are employed. Then, the relationship between the unknown source distribution and multiview and multispectral boundary measurements is established based on the finite element method and the optimal permissible source region. Since the measured data are noisy, the BLT reconstruction is formulated as l(2) data fidelity and a general regularization term. When choosing the regularization parameters for BLT, an efficient model function approach is proposed, which does not require knowledge of the noise level. This approach only requests the computation of the residual and regularized solution norm. With this knowledge, we construct the model function to approximate the objective function, and the regularization parameter is updated iteratively. Results: First, the micro-CT based mouse phantom was used for simulation verification. Simulation experiments were used to illustrate why multispectral data were used rather than monochromatic data. Furthermore, the study conducted using an adaptive regularization parameter demonstrated our ability to accurately localize the bioluminescent source. With the adaptively estimated regularization parameter, the reconstructed center position of the source was (20.37, 31.05, 12.95) mm, and the distance to the real source was 0.63 mm. The results of the dual-source experiments further showed that our algorithm could localize the bioluminescent sources accurately. The authors then presented experimental evidence that the proposed algorithm exhibited its calculated efficiency over the heuristic method. The effectiveness of the new algorithm was also confirmed by comparing it with the L-curve method. Furthermore, various initial speculations regarding the regularization parameter were used to illustrate the convergence of our algorithm. Finally, in vivo mouse experiment further illustrates the effectiveness of the proposed algorithm. Conclusions: Utilizing numerical, physical phantom and in vivo examples, we demonstrated that the bioluminescent sources could be reconstructed accurately with automatic regularization parameters. The proposed algorithm exhibited superior performance than both the heuristic regularization parameter choice method and L-curve method based on the computational speed and localization error. (C) 2011 American Association of Physicists in Medicine. [DOI: 10.1118/1.3635221]
引用
收藏
页码:5933 / 5944
页数:12
相关论文
共 51 条
[1]   Tomographic bioluminescence imaging by use of a combined optical-PET (OPET) system: a computer simulation feasibility study [J].
Alexandrakis, G ;
Rannou, FR ;
Chatziioannou, AF .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (17) :4225-4241
[2]  
[Anonymous], 2006, Deblurring images: matrices, spectra, and filtering
[3]   A FINITE-ELEMENT APPROACH FOR MODELING PHOTON TRANSPORT IN TISSUE [J].
ARRIDGE, SR ;
SCHWEIGER, M ;
HIRAOKA, M ;
DELPY, DT .
MEDICAL PHYSICS, 1993, 20 (02) :299-309
[4]   Large-scale active-set box-constrained optimization method with spectral projected gradients [J].
Birgin, EG ;
Martínez, JM .
COMPUTATIONAL OPTIMIZATION AND APPLICATIONS, 2002, 23 (01) :101-125
[5]   Image reconstruction for diffuse optical tomography using sparsity regularization and expectation-maximization algorithm [J].
Cao, Nannan ;
Nehorai, Arye ;
Jacob, Mathews .
OPTICS EXPRESS, 2007, 15 (21) :13695-13708
[6]   Hyperspectral and multispectral bioluminescence optical tomography for small animal imaging [J].
Chaudhari, AJ ;
Darvas, F ;
Bading, JR ;
Moats, RA ;
Conti, PS ;
Smith, DJ ;
Cherry, SR ;
Leahy, RM .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (23) :5421-5441
[7]   A Semismooth Newton Method for L1 Data Fitting with Automatic Choice of Regularization Parameters and Noise Calibration [J].
Clason, Christian ;
Jin, Bangti ;
Kunisch, Karl .
SIAM JOURNAL ON IMAGING SCIENCES, 2010, 3 (02) :199-231
[8]   Iterative method for bioluminescence tomography based on the radiative transport equation [J].
Cong, Wenxiang ;
Wang, Ge .
DEVELOPMENTS IN X-RAY TOMOGRAPHY V, 2006, 6318
[9]   Practical reconstruction method for bioluminescence tomography [J].
Cong, WX ;
Wang, G ;
Kumar, D ;
Liu, Y ;
Jiang, M ;
Wang, LV ;
Hoffman, EA ;
McLennan, G ;
McCray, PB ;
Zabner, J ;
Cong, A .
OPTICS EXPRESS, 2005, 13 (18) :6756-6771
[10]   Advances in vivo bioluminescence imaging of gene expression [J].
Contag, CH ;
Bachmann, MH .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2002, 4 :235-260