Reliability of a New Swept-Source Optical Coherence Tomography Biometer in Healthy Children, Adults, and Cataract Patients

被引:18
|
作者
Huang, Jinhai [1 ,2 ,3 ]
Zhao, Yuyu [1 ,2 ]
Savini, Giacomo [4 ]
Yu, Guanshun [1 ,2 ]
Yu, Jinjin [1 ,2 ]
Chen, Zhongxing [1 ,2 ]
Tu, Ruixue [1 ,2 ,3 ]
Zhao, Yune [1 ,2 ,3 ]
机构
[1] Wenzhou Med Univ, Sch Ophthalmol & Optometry, Wenzhou, Zhejiang, Peoples R China
[2] Wenzhou Med Univ, Eye Hosp, Wenzhou, Zhejiang, Peoples R China
[3] Minist Hlth China, Key Lab Vis Sci, Wenzhou, Zhejiang, Peoples R China
[4] GB Bietti Fdn IRCCS, Rome, Italy
关键词
INTRAOCULAR-LENS POWER; CALCULATION FORMULAS; MYOPIC PROGRESSION; AXIAL LENGTH; REPEATABILITY; AGREEMENT; ACCURACY; ANTERIOR; REPRODUCIBILITY; PARAMETERS;
D O I
10.1155/2020/8946364
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To comprehensively assess the reliability of a new optical biometer (IOLMaster 700), based on swept-source optical coherence tomography (SS-OCT) and comparison with a standard biometer (IOLMaster 500), in healthy children, adults, and cataract patients. Methods. A total of 301 eyes from 301 consecutive subjects were enrolled prospectively. Two experienced operators measured each eye three times consecutively with the IOLMaster 700. The axial length (AL), keratometry (K), anterior chamber depth (ACD), lens thickness (LT), central corneal thickness (CCT), and white-to-white (WTW) distance were recorded. Intraoperator repeatability and interoperator reproducibility of the IOLMaster 700 were analyzed using the test-retest (TRT), coefficients of variation (CoV), and intraclass correlation coefficients (ICCs). The agreement between the two devices was evaluated using the Bland-Altman method. Results. The repeatability and reproducibility of the SS-OCT optical biometer were high for all ocular biometry parameters in all groups, except for the WTW in cataract patients (TRT, 0.27-0.44 mm; ICC, 0.86-0.95). The reproducibility of averaged measurements from three consecutive readings (TRT : AL = 0.02 mm, CCT = 5.41 mu m, ACD = 0.03 mm, LT = 0.03 mm, Km = 0.17 D, and WTW = 0.22 mm) was higher than the reproducibility of single measurements (TRT : AL = 0.04 mm, CCT = 7.43 mu m, ACD = 0.06 mm, LT = 0.05 mm, Km = 0.26 D, and WTW = 0.35 mm) in the three groups. The consistency in the data between the two biometers was high, with narrow 95% LoAs in the three groups. Conclusion. Repeatability and reproducibility of the new SS-OCT optical biometer were excellent and consistent with that of the standard biometer with respect to healthy children, healthy adults, and cataract patients.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Axial length measurement and detection rates using a swept-source optical coherence tomography-based biometer in the presence of a dense vitreous hemorrhage
    Wang, Qinmei
    Huang, Yingying
    Gao, Rongrong
    Chen, Hao
    Song, Benhao
    Tu, Ruixue
    Zhao, Yune
    McAlinden, Colm
    Huang, Jinhai
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2020, 46 (03) : 360 - 364
  • [42] Intraobserver Repeatability of Tomographic, Pachymetric, and Anatomical Measurements in Healthy Eyes Using a New Swept-Source Optical Coherence Topographer
    Escolano Serrano, Jaime
    Tarazona Jaimes, Claudia Patricia
    Monera Lucas, Carlos Enrique
    Romero Valero, Daniel
    Moya Martinez, Alejandro
    Martinez Toldos, Jose Juan
    CORNEA, 2022, 41 (05) : 598 - 603
  • [43] Repeatability and agreement in optical biometry of a new swept-source optical coherence tomography-based biometer versus partial coherence interferometry and optical low-coherence reflectometry
    Kunert, Kathleen S.
    Peter, Monika
    Blum, Marcus
    Haigis, Wolfgang
    Sekundo, Walter
    Schuetze, Juliane
    Bueehren, Tobias
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (01) : 76 - 83
  • [44] Effect of pharmacological pupil dilation on measurements and iol power calculation made using the new swept-source optical coherence tomography-based optical biometer
    Arriola-Villatobos, P.
    Almendral-Gomez, J.
    Garzon, N.
    Ruiz-Medrano, J.
    Fernandez-Perez, C.
    Martinez-de-la-Casa, J. M.
    Diaz-Valled, D.
    JOURNAL FRANCAIS D OPHTALMOLOGIE, 2016, 39 (10): : 859 - 865
  • [45] Observation of anterior chamber volume after cataract surgery with swept-source optical coherence tomography
    Minjie Chen
    Hailin Hu
    Wenwen He
    Yi Lu
    Xiangjia Zhu
    International Ophthalmology, 2019, 39 : 1837 - 1844
  • [46] Topographic Variations of Choroidal Thickness in Healthy Eyes on Swept-Source Optical Coherence Tomography
    Touhami, Sara
    Philippakis, Elise
    Mrejen, Sarah
    Couturier, Aude
    Casteran, Celine
    Levent, Priscille
    Tadayoni, Ramin
    Gaudric, Alain
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2020, 61 (03)
  • [47] Reliability and repeatability of swept-source Fourier-domain optical coherence tomography and Scheimpflug imaging in keratoconus
    Szalai, Eszter
    Berta, Andras
    Hassan, Ziad
    Modis, Laszlo, Jr.
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2012, 38 (03) : 485 - 494
  • [48] Observation of anterior chamber volume after cataract surgery with swept-source optical coherence tomography
    Chen, Minjie
    Hu, Hailin
    He, Wenwen
    Lu, Yi
    Zhu, Xiangjia
    INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (08) : 1837 - 1844
  • [49] Comparative Analysis of Swept-Source Optical Coherence Tomography and Partial Coherence Interferometry Biometers in the Prediction of Cataract Surgery Refractive Outcomes
    Cummings, Arthur B.
    Naughton, Stephanie
    Coen, Ann-Marie
    Brennan, Elizabeth
    Kelly, Gabrielle E.
    CLINICAL OPHTHALMOLOGY, 2020, 14 : 4209 - 4220
  • [50] REPEATABILITY OF CHOROIDAL THICKNESS MEASUREMENTS ASSESSED WITH SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY IN HEALTHY AND DIABETIC INDIVIDUALS
    Abadia, Beatriz
    Calvo, Pilar
    Bartol-Puyal, Francisco
    Verdes, Guayente
    Sunen, Ines
    Ferreras, Antonio
    RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2019, 39 (04): : 786 - 793