Reliability of a New Swept-Source Optical Coherence Tomography Biometer in Healthy Children, Adults, and Cataract Patients

被引:18
|
作者
Huang, Jinhai [1 ,2 ,3 ]
Zhao, Yuyu [1 ,2 ]
Savini, Giacomo [4 ]
Yu, Guanshun [1 ,2 ]
Yu, Jinjin [1 ,2 ]
Chen, Zhongxing [1 ,2 ]
Tu, Ruixue [1 ,2 ,3 ]
Zhao, Yune [1 ,2 ,3 ]
机构
[1] Wenzhou Med Univ, Sch Ophthalmol & Optometry, Wenzhou, Zhejiang, Peoples R China
[2] Wenzhou Med Univ, Eye Hosp, Wenzhou, Zhejiang, Peoples R China
[3] Minist Hlth China, Key Lab Vis Sci, Wenzhou, Zhejiang, Peoples R China
[4] GB Bietti Fdn IRCCS, Rome, Italy
关键词
INTRAOCULAR-LENS POWER; CALCULATION FORMULAS; MYOPIC PROGRESSION; AXIAL LENGTH; REPEATABILITY; AGREEMENT; ACCURACY; ANTERIOR; REPRODUCIBILITY; PARAMETERS;
D O I
10.1155/2020/8946364
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose. To comprehensively assess the reliability of a new optical biometer (IOLMaster 700), based on swept-source optical coherence tomography (SS-OCT) and comparison with a standard biometer (IOLMaster 500), in healthy children, adults, and cataract patients. Methods. A total of 301 eyes from 301 consecutive subjects were enrolled prospectively. Two experienced operators measured each eye three times consecutively with the IOLMaster 700. The axial length (AL), keratometry (K), anterior chamber depth (ACD), lens thickness (LT), central corneal thickness (CCT), and white-to-white (WTW) distance were recorded. Intraoperator repeatability and interoperator reproducibility of the IOLMaster 700 were analyzed using the test-retest (TRT), coefficients of variation (CoV), and intraclass correlation coefficients (ICCs). The agreement between the two devices was evaluated using the Bland-Altman method. Results. The repeatability and reproducibility of the SS-OCT optical biometer were high for all ocular biometry parameters in all groups, except for the WTW in cataract patients (TRT, 0.27-0.44 mm; ICC, 0.86-0.95). The reproducibility of averaged measurements from three consecutive readings (TRT : AL = 0.02 mm, CCT = 5.41 mu m, ACD = 0.03 mm, LT = 0.03 mm, Km = 0.17 D, and WTW = 0.22 mm) was higher than the reproducibility of single measurements (TRT : AL = 0.04 mm, CCT = 7.43 mu m, ACD = 0.06 mm, LT = 0.05 mm, Km = 0.26 D, and WTW = 0.35 mm) in the three groups. The consistency in the data between the two biometers was high, with narrow 95% LoAs in the three groups. Conclusion. Repeatability and reproducibility of the new SS-OCT optical biometer were excellent and consistent with that of the standard biometer with respect to healthy children, healthy adults, and cataract patients.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Effect of orthokeratology on precision and agreement assessment of a new swept-source optical coherence tomography biometer
    Shu, Bao
    Bao, Fangjun
    Savini, Giacomo
    Lu, Weicong
    Tu, Ruixue
    Chen, Haisi
    Song, Benhao
    Wang, Qinmei
    Huang, Jinhai
    EYE AND VISION, 2020, 7 (01)
  • [2] Comparison of a new biometer using swept-source optical coherence tomography and a conventional biometer using partial coherence interferometry
    Higashiyama, Tomoaki
    Mori, Hazuki
    Nakajima, Fumi
    Ohji, Masahito
    PLOS ONE, 2018, 13 (04):
  • [3] Repeatability and reproducibility of a new spectral-domain optical coherence tomography biometer and agreement with swept-source optical coherence tomography based biometer
    Li, Xin
    Lei, Chak Seng
    Ning, Rui
    Liu, Luze
    Chen, Aodong
    Yang, Xinning
    Savini, Giacomo
    Schiano-Lomoriello, Domenico
    Zhou, Xingtao
    Huang, Jinhai
    EYE AND VISION, 2025, 12 (01)
  • [4] Repeatability of a fully automated swept-source optical coherence tomography biometer and agreement with a low coherence reflectometry biometer
    Dominguez-Vicent, Alberto
    Venkataraman, Abinaya Priya
    Dalin, Amanda
    Brautaset, Rune
    Montes-Mico, Robert
    EYE AND VISION, 2023, 10 (01)
  • [5] Comparison of a new optical biometer using swept-source optical coherence tomography and a biometer using optical low-coherence reflectometry
    Hoffer, Kenneth J.
    Hoffmann, Peter C.
    Savini, Giacomo
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2016, 42 (08) : 1165 - 1172
  • [6] Clinical Evaluation of a New Swept-Source Optical Coherence Biometer That Uses Individual Refractive Indices to Measure Axial Length in Cataract Patients
    Tamaoki, Akeno
    Kojima, Takashi
    Hasegawa, Asato
    Yamamoto, Mana
    Kaga, Tatsushi
    Tanaka, Kiyoshi
    Ichikawa, Kazuo
    OPHTHALMIC RESEARCH, 2019, 62 (01) : 11 - 23
  • [7] Ocular biometry with swept-source optical coherence tomography
    Montes-Mico, Robert
    Pastor-Pascual, Francisco
    Ruiz-Mesa, Ramon
    Tana-Rivero, Pedro
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2021, 47 (06) : 802 - 814
  • [8] Comparison of a New Optical Biometer That Combines Scheimpflug Imaging With Partial Coherence Interferometry With That of an Optical Biometer Based on Swept-Source Optical Coherence Tomography and Placido-Disk Topography
    Chen, Shihao
    Zhang, Qiaoyue
    Savini, Giacomo
    Zhang, Shuangzhe
    Huang, Xiaomin
    Yu, Jinjin
    Wang, Yirang
    Ning, Rui
    Huang, Jinhai
    Tu, Ruixue
    FRONTIERS IN MEDICINE, 2022, 8
  • [9] Precision of a new ocular biometer in eyes with cataract using swept source optical coherence tomography combined with Placido-disk corneal topography
    Wang, Wenwen
    Miao, Yaxin
    Savini, Giacomo
    McAlinden, Colm
    Chen, Hao
    Hu, Qingjie
    Wang, Qinmei
    Huang, Jinhai
    SCIENTIFIC REPORTS, 2017, 7
  • [10] Repeatability of a fully automated swept-source optical coherence tomography biometer and agreement with a low coherence reflectometry biometer
    Alberto Domínguez-Vicent
    Abinaya Priya Venkataraman
    Amanda Dalin
    Rune Brautaset
    Robert Montés-Micó
    Eye and Vision, 10