Need for Re-validation of Automated Blood Pressure Devices for use in Unstable Conditions

被引:0
作者
Zheng, Dingchang [1 ,2 ]
Liu, Chengyu [1 ,3 ]
Amoore, John [4 ]
Mieke, Stephan [5 ]
Murray, Alan [1 ]
机构
[1] Newcastle Univ, Inst Cellular Med, Newcastle Upon Tyne, Tyne & Wear, England
[2] Anglia Ruskin Univ, Hlth & Wellbeing Acad, Chelmsford, England
[3] Shandong Univ, Sch Control Sci & Engn, Jinan, Peoples R China
[4] NHS Ayrshire & Arran, Dept Med Phys, Ayr, Scotland
[5] Phys Tech Bundesanstalt, Berlin, Germany
来源
2015 COMPUTING IN CARDIOLOGY CONFERENCE (CINC) | 2015年 / 42卷
基金
英国工程与自然科学研究理事会;
关键词
TALKING; HYPERTENSION; VALIDATION; SIMULATOR; BODY;
D O I
暂无
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
The current validation of non-invasive blood pressure (NIBP) device is performed under resting conditon. However, NIBPs are often used without giving much consideration about the measurement conditions. This study aimed to provide scientific data on the use of BP devices in unstable conditions. BP measurements were performed on 20 healthy subjects under both resting and regular deep breathing conditions. During the measurement the oscillometric cuff pressure waveforms were recorded digitally. They were then regenerated by a specially designed BP simulator and presented to two clinically validated hospital grade automatic NIBP devices to obtain automated BPs. Automated BPs obtained from the two conditions were finally compared between the two devices. Under resting condition, there was no significant difference in both automated SBP and DBP between the two devices. However, under regular deep breathing condition, significant SBP and DBP differences were observed between the two devices ( both P<0.01; mean +/- SD: 118.8 +/- 10.6 vs 115.1 +/- 11.6 mmHg for SBP; 68.5 +/- 8.6 vs 65.3 +/- 8.9 mmHg for DBP). For the effect of deep breathing on BP measurement, significant SBP decrease was observed only from device 2 (P<0.05, with a mean difference +/- SD of 3.8 +/- 6.2 mmHg), indicating inconsistent measurements between the two devices under unstable conditions. Our results provide scientific evidence that automated BP devices can be used only under the condition for which the validation was performed.
引用
收藏
页码:17 / 20
页数:4
相关论文
共 13 条
  • [1] Can a simulator that regenerates physiological waveforms evaluate oscillometric non-invasive blood pressure devices?
    Amoore, JN
    Vacher, E
    Murray, IC
    Mieke, S
    King, ST
    Smith, FE
    Murray, A
    [J]. BLOOD PRESSURE MONITORING, 2006, 11 (02) : 63 - 67
  • [2] Validation of oscillometric noninvasive blood pressure measurement devices using simulators
    Amoore, John N.
    Lemesre, Yann
    Murray, Ian C.
    Vacher, Emilie
    Mieke, Stephan
    King, Susan T.
    Smith, Fiona E.
    Murray, Alan
    [J]. BLOOD PRESSURE MONITORING, 2007, 12 (04) : 251 - 253
  • [3] [Anonymous], 2013, 810602 ISO
  • [4] ABC of hypertension - Blood pressure measurement - Part I - Sphygmomanometry: factors common to all techniques
    Beevers, G
    Lip, GYH
    O'Brien, E
    [J]. BRITISH MEDICAL JOURNAL, 2001, 322 (7292): : 981 - 985
  • [5] The Accoson Greenlight 300™, the first non-automated mercury-free blood pressure measurement device to pass the International Protocol for blood pressure measuring devices in adults
    Graves, JW
    Tibor, M
    Murtagh, B
    Klein, L
    Sheps, SG
    [J]. BLOOD PRESSURE MONITORING, 2004, 9 (01) : 13 - 17
  • [6] Le Pailleur C, 2001, BEHAV MED, V26, P149
  • [7] The effects of talking, reading, and silence on the "white coat" phenomenon in hypertensive patients
    Le Pailleur, C
    Helft, G
    Landais, P
    Montgermont, P
    Feder, JM
    Metzger, JP
    Vacheron, A
    [J]. AMERICAN JOURNAL OF HYPERTENSION, 1998, 11 (02) : 203 - 207
  • [8] Influence of body and arm position on blood pressure readings: an overview
    Netea, RT
    Lenders, JWM
    Smits, P
    Thien, T
    [J]. JOURNAL OF HYPERTENSION, 2003, 21 (02) : 237 - 241
  • [9] European Society of hypertension recommendations for conventional, ambulatory and home blood pressure measurement
    O'Brien, E
    Asmar, R
    Beilin, L
    Imai, Y
    Mallion, JM
    Mancia, G
    Mengden, T
    Myers, M
    Padfield, P
    Palatini, P
    Parati, G
    Pickering, T
    Redon, J
    Staessen, J
    Stergiou, G
    Verdecchia, P
    [J]. JOURNAL OF HYPERTENSION, 2003, 21 (05) : 821 - 848
  • [10] Pickering Thomas G., 2005, Hypertension (Baltimore), V45, P142