Accuracy of intensity-modulated radiation therapy dose calculations: verification based on-board conebeam CT imaging

被引:0
|
作者
Tuntipumiamorn, Lalida [1 ]
Iampongpaiboon, Porntip [1 ]
Damrongkijudom, Nuanpen [2 ]
Krongyuth, Khummook [1 ]
Polwatsatian, Vinai [1 ]
Jaikreng, Satja [1 ]
机构
[1] Mahidol Univ, Siriraj Hosp, Fac Med, Div Radiat Oncol,Dept Radiol, Bangkok 10700, Thailand
[2] Mahidol Univ, Dept Radiol Technol, Fac Med Technol, Bangkok 10700, Thailand
关键词
Cone-beam CT; dose verification; IMRT; on-board imager; BEAM COMPUTED-TOMOGRAPHY; ELECTRON-DENSITY; RADIOTHERAPY; LOCALIZATION; QUALITY;
D O I
10.5372/1905-7415.0504.071
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Background: Patient's conebeam computer tomography (CBCT) images have suggested a possibility for adaptive radiotherapy although the dose delivery is of structural complexity. It is of practical importance to verify and test the intensity-modulated radiation (IMRT) planning system for radiation therapy. Objective: Verify accuracy of dose calculations based on CBCT imaging. Materials and methods: Electron density calibration curve was generated for planning CT and CBCT data set using two CT phantoms (Gammex RMI (R) and Catphan (R) 600). Anthropomorphic head and neck phantom images were acquired from planning CT and CBCT. The routine IMRT technique was generated on the planning CT, which was applied to the CBCT. Dose distributions were computed. All LiF TLD-100 dosimeters were calibrated with gamma-ray. Forty-eight TLD measuring points were chosen in five different slices of the phantom. Measurements were repeated four times, and the average dose was compared to the reading doses on both CT and CBCT plans. Dose volume histograms (DV H) of various structures were generated, and dose statistics were analyzed. Results: Hounsfield unit obtained from Catphan phantom was similar between planning CT and CBCT. IMRT dose calculations based on the planning CT and CBCT agreed well with reading doses at 48 points. Statistical point doses by DVH calculation on CBCT were about 3% lower than those by the conventional CT. Dose ratios calculated over measured ones ranged from 0.82 to 1.09. Conclusion: Point doses and DVH calculations based on the planning CT and on-board CBCT were in acceptable agreement. CT phantom specifically designed for CBCT is recommended to improve accuracy of IMRT dose calculation on CBCT images.
引用
收藏
页码:537 / 541
页数:5
相关论文
共 50 条
  • [21] Accuracy and sources of error of out-of field dose calculations by a commercial treatment planning system for intensity-modulated radiation therapy treatments
    Huang, Jessie Y.
    Followill, David S.
    Wang, Xin A.
    Kry, Stephen F.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (02): : 186 - 197
  • [22] Development and commissioning of a multileaf collimator model in Monte Carlo dose calculations for intensity-modulated radiation therapy
    Jang, SY
    Vassiliev, ON
    Liu, HH
    Mohan, R
    Siebers, JV
    MEDICAL PHYSICS, 2006, 33 (03) : 770 - 781
  • [23] Accuracy requirements for head and neck intensity-modulated radiation therapy based on observed dose response of the major salivary glands
    Kapanen, Mika
    Collan, Juhani
    Saarilahti, Kauko
    Heikkonen, Jorma
    Kairemo, Kalevi
    Tenhunen, Mikko
    RADIOTHERAPY AND ONCOLOGY, 2009, 93 (01) : 109 - 114
  • [24] Comparison of dose accuracy between film and two-dimensional detectors in intensity-modulated radiation therapy
    Yuichi Onishi
    Shinichi Nakayama
    Shinsaku Watanabe
    Souichirou Kaneshige
    Hajime Monzen
    Kenji Matsumoto
    Naoya Shintani
    Takeshi Kamomae
    Journal of the Korean Physical Society, 2015, 67 : 89 - 95
  • [25] Comparison of Dose Accuracy between Film and Two-dimensional Detectors in Intensity-modulated Radiation Therapy
    Onishi, Yuichi
    Nakayama, Shinichi
    Watanabe, Shinsaku
    Kaneshige, Souichirou
    Monzen, Hajime
    Matsumoto, Kenji
    Shintani, Naoya
    Kamomae, Takeshi
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 67 (01) : 89 - 95
  • [26] Accuracy of dose calculation algorithms for virtual heterogeneous phantoms and intensity-modulated radiation therapy in the head and neck
    Onizuka R.
    Araki F.
    Ohno T.
    Nakaguchi Y.
    Kai Y.
    Tomiyama Y.
    Hioki K.
    Radiological Physics and Technology, 2016, 9 (1) : 77 - 87
  • [27] ICRU reference dose in an era of intensity-modulated radiation therapy clinical trials: Correlation with planning target volume mean dose and suitability for intensity-modulated radiation therapy dose prescription
    Yaparpalvi, Ravindra
    Hong, Linda
    Mah, Dennis
    Shen, Jin
    Mutyala, Subhakar
    Spierer, Marnee
    Garg, Madhur
    Guha, Chandan
    Kalnicki, Shalom
    RADIOTHERAPY AND ONCOLOGY, 2008, 89 (03) : 347 - 352
  • [28] Database energy fluence verification and the importance of on-board CT imaging in dose reconstruction
    Kapatoes, JM
    Olivera, GH
    Ruchala, KJ
    Reckwerdt, PJ
    Smilowitz, JB
    Balog, JP
    Pearson, DW
    Mackie, TR
    USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 294 - 296
  • [29] Dose verification of a clinical intensity-modulated radiation therapy eye case by the magnetic resonance imaging of N-isopropylacrylamide gel dosimeters
    Chen, Yen-Li
    Hsieh, Bor-Tsung
    Chiang, Chih-Ming
    Shih, Cheng-Ting
    Cheng, Kai-Yuan
    Hsieh, Ling-Ling
    RADIATION PHYSICS AND CHEMISTRY, 2014, 104 : 188 - 191
  • [30] Intensity-Modulated Proton Therapy with Dose Painting based on Hypoxia Imaging for Nasopharyngeal Cancer
    Yasuda, K.
    Takao, S.
    Matsuo, Y.
    Yoshimura, T.
    Tamura, M.
    Minatogawa, H.
    Dekura, Y.
    Matsuura, T.
    Onimaru, R.
    Shiga, T.
    Shimizu, S.
    Umegaki, K.
    Shirato, H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2018, 102 (03): : E378 - E378