Carotid intima-media thickness and hemodynamic parameters: reproducibility of manual measurements with Doppler ultrasound

被引:15
作者
Plasencia Martinez, Juana M. [1 ]
Garcia Santos, Jose M. [1 ]
Paredes Martinez, Maria L. [1 ]
Moreno Pastor, Ana [1 ]
机构
[1] Univ Gen Hosp Morales Meseguer, Dept Radiol, Murcia, Spain
关键词
ultrasonography; Doppler; carotid intima-media thickness; observer variation; reproducibility; CORONARY-ARTERY-DISEASE; CARDIOVASCULAR-DISEASE; MYOCARDIAL-INFARCTION; DIFFERENT SITES; RISK; PLAQUE; ATHEROSCLEROSIS; EVENTS; VARIABILITY; MEDICINE;
D O I
10.11152/mu.2013.2066.172.ci-m
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Aims: To evaluate the carotid ultrasound intra-and interobserver agreements in a common clinical scenario when making manual measurements of the intima-media thickness (IMT) and peak systolic (PSV) and end diastolic (EDV) velocities in the common (CCA) and the internal carotid (ICA) arteries. Material and methods: Three different experienced operators performed two time-point carotid ultrasounds in 21 patients with cardiovascular risk factors. Each operator measured freehand the CCA IMT three consecutive times in each examination. The CCA and ICA hemodynamic parameters were acquired just once. For our purpose we took the average (IMTmean) and maximum (IMTmax) IMT values. Quantitative variables were analyzed with the t-student, and ANOVA test. Agreements were evaluated with the Intraclass Correlation Coefficient (ICC). Results: IMTmean intraobserver agreement was better on the left (ICC: 0.930-0.851-0.916, operators 1-2-3) than on the right (ICC: 0.789-0.580-0.673, operators 1-2-3). IMTmax agreements (Left ICC: 0.821-0.723-0.853, operators 1-2-3; Right ICC: 0.669-0.421-0.480, operators 1-2-3) were lower and more variable. Interobserver agreements for IMTmean (ICC: 0.852-0.860; first-second ultrasound) and IMTmax (ICC: 0.859-0.835; first-second ultrasound) were excellent on the left, but fair-good and more variable on the right (IMTmean; ICC: 0.680-0.809; first-second ultrasound; IMTmax; 0.694-0.799; first-second ultrasound). Intraobserver agreements were fair-moderate for PSVs and good-excellent for EDVs. Interobserver agreements were good-excellent for both PSVs and EDVs. Overall, 95% confidence intervals were narrower for the left IMTmean and CCA velocities. Conclusions: Intra and interobserver agreements in carotid ultrasound are variable. In order to improve carotid IMT agreements, IMTmean is preferable over IMTmax.
引用
收藏
页码:167 / 174
页数:8
相关论文
共 50 条
  • [1] Carotid Intima-Media Thickness Manual Measurements Intraoperator and Interoperator Agreements Under A Strict Protocol in a Large Sample
    Plasencia-Martinez, Juana M.
    Maria Garcia-Santos, Jose
    Lozano-Herrero, Jesus
    Julia Hernandez-Vidal, M.
    ULTRASOUND QUARTERLY, 2017, 33 (01) : 28 - 36
  • [2] Ultrasound semiautomatic versus manual estimation of carotid intima-media thickness: reproducibility and cardiovascular risk stratification
    Monti, Caterina Beatrice
    Ali, Marco
    Capra, Davide
    Wiedenmann, Federico
    Lastella, Giulia
    Secchi, Francesco
    Sardanelli, Francesco
    MEDICAL ULTRASONOGRAPHY, 2020, 22 (04) : 402 - 408
  • [3] Carotid Intima-Media Thickness Measurements Are Not Affected by the Ultrasound Frequency
    Bhagirath, Kapil M.
    Lester, Steven J.
    Humphries, Julie
    Hentz, Joseph G.
    Hurst, R. Todd
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2012, 29 (03): : 354 - 357
  • [4] Carotid Artery Intima-Media Thickness Measurements in the Youth: Reproducibility and Technical Considerations
    Tierney, Elif Seda Selamet
    Gauvreau, Kimberlee
    Jaff, Michael R.
    Gal, Dana
    Nourse, Susan E.
    Trevey, Shari
    O'Neill, Stephen
    Baker, Annette
    Newburger, Jane W.
    Colan, Steven D.
    JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2015, 28 (03) : 309 - 316
  • [5] Reproducibility of Sonographic Measurements of Carotid Intima-Media Thickness
    Velazquez, F.
    Berna, J. D.
    Abellan, J. L.
    Serrano, L.
    Escribano, A.
    Canteras, M.
    ACTA RADIOLOGICA, 2008, 49 (10) : 1162 - 1166
  • [6] Ultrasound assessment of atherosclerotic vessel wall changes -: Reproducibility of intima-media thickness measurements in carotid and femoral arteries
    Srámek, A
    Bosch, JG
    Reiber, JHC
    Van Oostayen, JA
    Rosendaal, FR
    INVESTIGATIVE RADIOLOGY, 2000, 35 (12) : 699 - 706
  • [7] Variability of automated carotid intima-media thickness measurements by novice operators
    Nichols, S.
    Milner, M.
    Meijer, R.
    Carroll, S.
    Ingle, L.
    CLINICAL PHYSIOLOGY AND FUNCTIONAL IMAGING, 2016, 36 (01) : 25 - 32
  • [8] Reproducibility of in vivo carotid intima-media thickness measurements - A review
    Kanters, SDJM
    Algra, A
    vanLeeuwen, MS
    Banga, JD
    STROKE, 1997, 28 (03) : 665 - 671
  • [9] Carotid intima-media thickness for the practicing lipidologist
    Liviakis, Lea
    Pogue, Bryan
    Paramsothy, Pathmaja
    Bourne, Alicia
    Gill, Edward A.
    JOURNAL OF CLINICAL LIPIDOLOGY, 2010, 4 (01) : 24 - 35
  • [10] Snoring and Carotid Artery Intima-Media Thickness
    Deeb, Robert
    Judge, Paul
    Peterson, Ed
    Lin, Judith C.
    Yaremchuk, Kathleen
    LARYNGOSCOPE, 2014, 124 (06) : 1486 - 1491