Pneumopericardium and pneumothorax contralateral to venous access site after permanent pacemaker implantation

被引:24
作者
Srivathsan, K [1 ]
Byrne, RA [1 ]
Appleton, CP [1 ]
Scott, LRP [1 ]
机构
[1] Mayo Clin Scottsdale, Div Cardiovasc Dis, Phoenix, AZ USA
来源
EUROPACE | 2003年 / 5卷 / 04期
关键词
contralateral pneumothorax; dual chamber permanent pacemaker; pacemaker insertion complications;
D O I
10.1016/S1099-5129(03)00093-X
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
A 77-year-old female underwent implantation of a left-sided dual chamber permanent pacemaker for symptomatic bradycardia with active fixation leads. Eight hours after the procedure, the patient complained of shortness of breath and was found to have a 30% right pneumothorax on chest X-ray. Immediately, a chest tube was inserted, promptly relieving the symptoms. A CT scan of the chest revealed extrusion of the helix of the screw-in atrial lead, through the wall of the right atrial appendage. The helix was abutting a bulla in the right lung, the likely cause for pneumothorax and pneumopericardium. The atrial lead was explanted without incident. (C) 2003 The European Society of Cardiology. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:361 / 363
页数:3
相关论文
共 13 条
  • [1] AGGARWAL RK, 1995, BRIT HEART J, V73, P571
  • [2] EARLY COMPLICATIONS AFTER DUAL-CHAMBER VERSUS SINGLE CHAMBER PACEMAKER IMPLANTATION
    CHAUHAN, A
    GRACE, AA
    NEWELL, SA
    STONE, DL
    SHAPIRO, LM
    SCHOFIELD, PM
    PETCH, MC
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (11): : 2012 - 2015
  • [3] Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes
    Connolly, SJ
    Kerr, CR
    Gent, M
    Roberts, RS
    Yusuf, S
    Gillis, AM
    Sami, MH
    Talajic, M
    Tang, ASL
    Klein, GJ
    Lau, C
    Newman, DM
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (19) : 1385 - 1391
  • [4] PERICARDIAL COMPLICATIONS OF ENDOCARDIAL AND EPICARDIAL PACING
    DEANFIELD, J
    JONATHAN, A
    FOX, K
    [J]. BRITISH MEDICAL JOURNAL, 1981, 283 (6292) : 635 - 636
  • [5] Delayed complications following pacemaker implantation
    Ellenbogen, KA
    Wood, MA
    Shepard, RK
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2002, 25 (08): : 1155 - 1158
  • [6] CLINICAL SURVEILLANCE OF AN ACTIVE FIXATION, BIPOLAR, POLYURETHANE INSULATED PACING LEAD .1. THE ATRIAL LEAD
    GLIKSON, M
    VONFELDT, LK
    SUMAN, VJ
    HAYES, DL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1994, 17 (08): : 1399 - 1404
  • [7] REAL-TIME, TWO-DIMENSIONAL ECHOCARDIOGRAPHIC FEATURES OF PACEMAKER PERFORATION
    GONDI, B
    NANDA, NC
    [J]. CIRCULATION, 1981, 64 (01) : 97 - 106
  • [8] Right pneumothorax resulting from an endocardial screw-in atrial lead
    Ho, WJ
    Kuo, CT
    Lin, KH
    [J]. CHEST, 1999, 116 (04) : 1133 - 1134
  • [9] Quality of life and clinical outcomes in elderly patients treated with ventricular pacing as compared with dual-chamber pacing
    Lamas, GA
    Orav, J
    Stambler, BS
    Ellenbogen, KA
    Sgarbossa, EB
    Huang, SKS
    Marinchak, RA
    Estes, NAM
    Mitchell, GF
    Lieberman, EH
    Mangione, CM
    Goldman, L
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1998, 338 (16) : 1097 - 1104
  • [10] Complications of dual chamber pacemaker implantation in the elderly
    Link, MS
    Estes, NAM
    Griffin, JJ
    Wang, PJ
    Maloney, JD
    Kirchhoffer, JB
    Mitchell, GF
    Orav, J
    Goldman, L
    Lamas, GA
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 1998, 2 (02) : 175 - 179