共 50 条
Quality assurance tool for determination of position and transit time of a Co-60 source in high dose rate brachytherapy
被引:0
作者:
Piriz, Gustavo H.
[1
]
Ortega-Spina, Henry G.
[1
]
Gonzalez-Sprinberg, Gabriel A.
[1
]
机构:
[1] Univ Republica, Fac Ciencias, Montevideo 11400, Uruguay
关键词:
Brachytherapy;
Cobalt-60;
Quality assurance;
hdr;
IR-192;
BRACHYTHERAPY;
D O I:
10.1016/j.apradiso.2021.109971
中图分类号:
O61 [无机化学];
学科分类号:
070301 ;
081704 ;
摘要:
In this study, three holders were designed, constructed and characterized to perform quality assurance on the source position and transit time in remote afterloading systems with Co-60 sources for high dose rate brachytherapy. The holders design focused on achieving accuracy, low cost, and a time efficient tool for use in clinical settings. Sensitivities greater than 0.6%/mm and maximum precisions better than 0.14 mm for the source position were obtained. The transit time was determined for the holders with a relative precision better than 19%.