Quality assurance tool for determination of position and transit time of a Co-60 source in high dose rate brachytherapy

被引:0
作者
Piriz, Gustavo H. [1 ]
Ortega-Spina, Henry G. [1 ]
Gonzalez-Sprinberg, Gabriel A. [1 ]
机构
[1] Univ Republica, Fac Ciencias, Montevideo 11400, Uruguay
关键词
Brachytherapy; Cobalt-60; Quality assurance; hdr; IR-192; BRACHYTHERAPY;
D O I
10.1016/j.apradiso.2021.109971
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In this study, three holders were designed, constructed and characterized to perform quality assurance on the source position and transit time in remote afterloading systems with Co-60 sources for high dose rate brachytherapy. The holders design focused on achieving accuracy, low cost, and a time efficient tool for use in clinical settings. Sensitivities greater than 0.6%/mm and maximum precisions better than 0.14 mm for the source position were obtained. The transit time was determined for the holders with a relative precision better than 19%.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] Verification of High Dose Rate 192Ir Source Position During Brachytherapy Treatment Using Silicon Pixel Detectors
    Batic, Matej
    Burger, Janez
    Cindro, Vladimir
    Kramberger, Gregor
    Mandic, Igor
    Mikuz, Marko
    Studen, Andrej
    Zavrtanik, Marko
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2011, 58 (05) : 2250 - 2256
  • [42] Three-dimensional source position verification in image-guided high-dose-rate brachytherapy using an XCT-based gel dosimeter
    Yonemura, Miki
    Tachibana, Hidenobu
    Kojima, Toru
    Seki, Kazuya
    Nakaichi, Tetsu
    Rachi, Toshiya
    Tachibana, Rie
    Akimoto, Tetsuo
    MEDICAL PHYSICS, 2025, 52 (02) : 1243 - 1255
  • [43] Dosimetric characterizations of GZP6 60Co high dose rate brachytherapy sources: application of superimposition method
    Toossi, Mohammad Taghi Bahreyni
    Ghorbani, Mahdi
    Mowlavi, Ali Asghar
    Meigooni, Ali Soleimani
    RADIOLOGY AND ONCOLOGY, 2012, 46 (02) : 170 - 178
  • [44] A comparative study on the accuracy of an automated multi-parameter high-dose-rate brachytherapy quality assurance system
    Zhang, Jing
    Jia, Mengyu
    Yuan, Zhiyong
    Qian, Shaowen
    Zhang, Daguang
    Wang, Wei
    RADIATION PHYSICS AND CHEMISTRY, 2024, 217
  • [45] A comparative study on the accuracy of an automated multi-parameter high-dose-rate brachytherapy quality assurance system
    Zhang, Jing
    Jia, Mengyu
    Yuan, Zhiyong
    Qian, Shaowen
    Zhang, Daguang
    Wang, Wei
    Radiation Physics and Chemistry, 2024, 217
  • [46] Air-kerma strength determination of a 169Yb high dose rate brachytherapy source
    VanDamme, J. J.
    Culberson, W. S.
    DeWerd, L. A.
    Micka, J. A.
    MEDICAL PHYSICS, 2008, 35 (09) : 3935 - 3942
  • [47] Application of a color scanner for 60Co high dose rate brachytherapy dosimetry with EBT radiochromic film
    Ghorbani, Mahdi
    Toossi, Mohammad Taghi Bahreyni
    Mowlavi, Ali Asghar
    Roodi, Shahram Bayani
    Meigooni, Ali Soleimani
    RADIOLOGY AND ONCOLOGY, 2012, 46 (04) : 363 - 369
  • [48] Design and testing of a phantom and instrumented gynecological applicator based on GaN dosimeter for use in high dose rate brachytherapy quality assurance
    Guiral, P.
    Ribouton, J.
    Jalade, P.
    Wang, R.
    Galvan, J. -M.
    Lu, G. -N.
    Pittet, P.
    Rivoire, A.
    Gindraux, L.
    MEDICAL PHYSICS, 2016, 43 (09) : 5240 - 5251
  • [49] The feasibility study and characterization of a two-dimensional diode array in "magic phantom" for high dose rate brachytherapy quality assurance
    Espinoza, A.
    Beeksma, B.
    Petasecca, M.
    Fuduli, I.
    Porumb, C.
    Cutajar, D.
    Corde, S.
    Jackson, M.
    Lerch, M. L. F.
    Rosenfeld, A. B.
    MEDICAL PHYSICS, 2013, 40 (11)
  • [50] Performance evaluation of a direct-conversion flat-panel detector system in imaging and quality assurance for a high-dose-rate 192Ir source
    Miyahara, Yoshinori
    Hara, Yuki
    Nakashima, Hiroto
    Nishimura, Tomonori
    Itakura, Kanae
    Inomata, Taisuke
    Kitagaki, Hajime
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (05)