Quality assurance tool for determination of position and transit time of a Co-60 source in high dose rate brachytherapy

被引:0
作者
Piriz, Gustavo H. [1 ]
Ortega-Spina, Henry G. [1 ]
Gonzalez-Sprinberg, Gabriel A. [1 ]
机构
[1] Univ Republica, Fac Ciencias, Montevideo 11400, Uruguay
关键词
Brachytherapy; Cobalt-60; Quality assurance; hdr; IR-192; BRACHYTHERAPY;
D O I
10.1016/j.apradiso.2021.109971
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In this study, three holders were designed, constructed and characterized to perform quality assurance on the source position and transit time in remote afterloading systems with Co-60 sources for high dose rate brachytherapy. The holders design focused on achieving accuracy, low cost, and a time efficient tool for use in clinical settings. Sensitivities greater than 0.6%/mm and maximum precisions better than 0.14 mm for the source position were obtained. The transit time was determined for the holders with a relative precision better than 19%.
引用
收藏
页数:8
相关论文
共 50 条
  • [31] Dose assessment in high dose rate brachytherapy with cobalt-60 source for cervical cancer treatment: a phantom study
    Bour, Bright Kwadwo
    Inkoom, Stephen
    Tagoe, Samuel Nii Adu
    Amuasi, John Humphrey
    Sasu, Evans
    Hasford, Francis
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2020, 26 (04): : 243 - 250
  • [32] Effect of inhomogeneities and source position on dose distribution of nucletron high dose rate Ir-192 brachytherapy source by Monte Carlo simulation
    Chandola, R. M.
    Tiwari, S.
    Kowar, M. K.
    Choudhary, V.
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2010, 6 (01) : 54 - 57
  • [33] Treatment planning study of the 3D dosimetric differences between Co-60 and Ir-192 sources in high dose rate (HDR) brachytherapy
    Palmer, Antony
    Hayman, Orla
    Muscat, Sarah
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2012, 4 (01) : 52 - 59
  • [34] A user-oriented procedure for the commissioning and quality assurance testing of treatment planning system dosimetry in high-dose-rate brachytherapy
    Peppa, Vasiliki
    Pantelis, Evaggelos
    Pappas, Eleftherios
    Lahanas, Vasileios
    Loukas, Constantinos
    Papagiannis, Panagiotis
    BRACHYTHERAPY, 2016, 15 (02) : 252 - 262
  • [35] Dosimetric characterization of the 60Co BEBIG Co0.A86 high dose rate brachytherapy source using PENELOPE
    Guerrero, Rafael
    Almansa, Julio F.
    Torres, Javier
    Lallena, Antonio M.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2014, 30 (08): : 960 - 967
  • [36] Technical Note: Scintillation markers for real-time visual source tracking during skin high dose rate brachytherapy
    Huynh, Elizabeth
    Bhagwat, Mandar S.
    MEDICAL PHYSICS, 2020, 47 (11) : 5490 - 5495
  • [37] Feasibility Study of Patient-specific Quality Assurance System for High-dose-rate Brachytherapy in Patients with Cervical Cancer
    Lee, Boram
    Ahn, Sung Hwan
    Kim, Hyeyoung
    Han, Youngyih
    Huh, Seung Jae
    Kim, Jin Sung
    Kim, Dong Wook
    Sim, Jina
    Yoon, Myonggeun
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2016, 68 (08) : L1029 - L1036
  • [38] Feasibility study of patient-specific quality assurance system for high-dose-rate brachytherapy in patients with cervical cancer
    Boram Lee
    Sung Hwan Ahn
    Hyeyoung Kim
    Youngyih Han
    Seung Jae Huh
    Jin Sung Kim
    Dong Wook Kim
    Jina Sim
    Myonggeun Yoon
    Journal of the Korean Physical Society, 2016, 68 : 1029 - 1036
  • [39] Source position measurement by Cherenkov emission imaging from applicators for high-dose-rate brachytherapy
    Yogo, Katsunori
    Noguchi, Yumiko
    Okudaira, Kuniyasu
    Nozawa, Marika
    Ishiyama, Hiromichi
    Okamoto, Hiroyuki
    Yasuda, Hiroshi
    Oguchi, Hiroshi
    Yamamoto, Seiichi
    MEDICAL PHYSICS, 2021, 48 (01) : 488 - 499
  • [40] Assessment of a source position checking tool for the quality assurance of transfer tubes used in HDR 192Ir brachytherapy treatments
    Awunor, O. A.
    BRACHYTHERAPY, 2018, 17 (03) : 628 - 633