Quality assurance tool for determination of position and transit time of a Co-60 source in high dose rate brachytherapy

被引:0
作者
Piriz, Gustavo H. [1 ]
Ortega-Spina, Henry G. [1 ]
Gonzalez-Sprinberg, Gabriel A. [1 ]
机构
[1] Univ Republica, Fac Ciencias, Montevideo 11400, Uruguay
关键词
Brachytherapy; Cobalt-60; Quality assurance; hdr; IR-192; BRACHYTHERAPY;
D O I
10.1016/j.apradiso.2021.109971
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
In this study, three holders were designed, constructed and characterized to perform quality assurance on the source position and transit time in remote afterloading systems with Co-60 sources for high dose rate brachytherapy. The holders design focused on achieving accuracy, low cost, and a time efficient tool for use in clinical settings. Sensitivities greater than 0.6%/mm and maximum precisions better than 0.14 mm for the source position were obtained. The transit time was determined for the holders with a relative precision better than 19%.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Radiochromic film-based quality assurance for CT-based high-dose-rate brachytherapy
    Asgharizadeh, Saeid
    Bekerat, Hamed
    Syme, Alasdair
    Aldelaijan, Saad
    DeBlois, Francois
    Te Vuong
    Evans, Michael
    Seuntjens, Jan
    Devic, Slobodan
    BRACHYTHERAPY, 2015, 14 (04) : 578 - 585
  • [22] Quality assurance measurements for high-dose-rate brachytherapy without film
    Jursinic, Paul A.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2014, 15 (01): : 246 - 261
  • [23] Borosilicate glass 60Co high dose rate brachytherapy thermoluminescence dosimetry
    Nazeri, A. A. Z. Ahmad
    Sani, S. F. Abdul
    Ung, N. M.
    Almugren, K. S.
    Alkallas, F. H.
    Bradley, D. A.
    APPLIED RADIATION AND ISOTOPES, 2021, 176
  • [24] Design of a phantom for the quality control of high dose rate 192Ir source used in brachytherapy
    Ochoa, Ricardo
    Gomez, Fredy
    Ferreira, Ivaldo H.
    Gutt, Federico
    de Almeida, Carlos E.
    RADIOTHERAPY AND ONCOLOGY, 2007, 82 (02) : 222 - 228
  • [25] Verification of high dose rate 192Ir source position during brachytherapy treatment
    Batic, M.
    Burger, J.
    Cindro, V.
    Kramberger, G.
    Mandic, I.
    Mikuz, M.
    Studen, A.
    Zavrtanik, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 617 (1-3) : 206 - 208
  • [26] Quality assurance of treatment plans for interstitial and intracavitary high-dose-rate brachytherapy
    Das, Rupak K.
    Bradley, Kristin A.
    Nelson, Ian A.
    Patel, Rakesh
    Thomadsen, Bruce R.
    BRACHYTHERAPY, 2006, 5 (01) : 56 - 60
  • [27] On the potential of 2D ion chamber arrays for high-dose rate remote afterloading brachytherapy quality assurance
    Lekatou, Aristea
    Peppa, Vasiliki
    Karaiskos, Pantelis
    Pantelis, Evangelos
    Papagiannis, Panagiotis
    PHYSICS IN MEDICINE AND BIOLOGY, 2022, 67 (08)
  • [28] The evaluation of a 2D diode array in "magic phantom" for use in high dose rate brachytherapy pretreatment quality assurance
    Espinoza, A.
    Petasecca, M.
    Fuduli, I.
    Howie, A.
    Bucci, J.
    Corde, S.
    Jackson, M.
    Lerch, M. L. F.
    Rosenfeld, A. B.
    MEDICAL PHYSICS, 2015, 42 (02) : 663 - 673
  • [29] Determination of the tissue inhomogeneity correction in high dose rate Brachytherapy for Iridium-192 source
    Ravikumar, Barlanka
    Lakshminarayana, S.
    JOURNAL OF MEDICAL PHYSICS, 2012, 37 (01) : 27 - 31
  • [30] Study of the dosimetric differences between Ir-192 and Co-60 sources of high dose rate brachytherapy for breast interstitial implant
    Sinnatamby, Mourougan
    Nagarajan, Vivekanandan
    Sathyanarayana, Reddy Kanipakam
    Karunanidhi, Gunaseelan
    Singhavajala, Vivekanandam
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2016, 21 (05) : 453 - 459