Gadolinium-Enhanced Liver Magnetic Resonance Imaging Using a 2-Point Dixon Fat-Water Separation Technique: Impact Upon Image Quality and Lesion Detection

被引:20
作者
Rosenkrantz, Andrew B. [1 ]
Mannelli, Lorenzo [1 ]
Kim, Sungheon [1 ]
Babb, James S. [1 ]
机构
[1] NYU, Dept Radiol, Langone Med Ctr, New York, NY 10016 USA
关键词
liver; VIBE; Dixon; fat suppression; BREATH-HOLD EXAMINATION; HEPATOCELLULAR-CARCINOMA; PULSE SEQUENCE; MR; RECONSTRUCTION; DECOMPOSITION; SUPPRESSION; TIME; EXPERIENCE; EFFICIENT;
D O I
10.1097/RCT.0b013e3181f3d57e
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To compare image quality and lesion detection in postcontrast liver magnetic resonance imaging (MRI) between volumetric interpolated breath-hold examination (VIBE) sequences that achieve fat suppression via chemically selective fat saturation (FS-VIBE) and a 2-point Dixon water-fat separation method (Dixon-VIBE). Materials and Methods: Thirty patients underwent contrast-enhanced liver MRI at 1.5 T in which Dixon-VIBE was performed immediately after a delayed FS-VIBE. Two radiologists in consensus reviewed the sequences for a variety of qualitative and quantitative image quality measures and for lesion detection. Results: Dixon-VIBE received nearly perfect scores for strength and homogeneity of fat suppression that were significantly better than scores for FS-VIBE, with an associated significant improvement in liver-fat contrast (P < 0.0001 for all comparisons). Dixon-VIBE also received significantly better scores for sharpness of intrahepatic vessels (P = 0.0029) and overall image quality (P < 0.0001). Despite a slightly longer acquisition time for Dixon-VIBE, there was no significant difference in motion artifact (P = 0.3877). There was no significant difference for sensitivity, positive predictive value, or contrast relative to background liver for focal lesions (P = 0.448, P = 0.347, and P = 0.2312, respectively). Conclusions: For postcontrast liver MRI, Dixon-VIBE demonstrated significantly improved fat suppression. Various assessments of lesion detection showed no significant difference between sequences.
引用
收藏
页码:96 / 101
页数:6
相关论文
共 35 条
[31]   Abdominal MR imaging with a volumetric interpolated breath-hold examination [J].
Rofsky, NM ;
Lee, VS ;
Laub, G ;
Pollack, MA ;
Krinsky, GA ;
Thomasson, D ;
Ambrosino, MM ;
Weinreb, JC .
RADIOLOGY, 1999, 212 (03) :876-884
[32]   Multiecho Time-Resolved Acquisition (META): A High Spatiotemporal Resolution Dixon Imaging Sequence for Dynamic Contrast-Enhanced MRI [J].
Saranathan, Manojkumar ;
Rettmann, Dan ;
Bayram, Ersin ;
Lee, Christine ;
Glockner, James .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2009, 29 (06) :1406-1413
[33]   Suppression of intravascular signal on fat-saturated contrast-enhanced thoracic MR arteriograms [J].
Siegelman, ES ;
Charafeddine, R ;
Stolpen, AH ;
Axel, L .
RADIOLOGY, 2000, 217 (01) :115-118
[34]   Multiecho reconstruction for simultaneous water-fat decomposition and T2*estimation [J].
Yu, Huanzhou ;
McKenzie, Charles A. ;
Shimakawa, Ann ;
Vu, Anthony T. ;
Brau, Anja C. S. ;
Beatty, Philip J. ;
Pineda, Angel R. ;
Brittain, Jean H. ;
Reeder, Scott B. .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2007, 26 (04) :1153-1161
[35]   Field map estimation with a region growing scheme for iterative 3-point water-fat decomposition [J].
Yu, HZ ;
Reeder, SB ;
Shimakawa, A ;
Brittain, JH ;
Pelc, NJ .
MAGNETIC RESONANCE IN MEDICINE, 2005, 54 (04) :1032-1039