Effect of various tapping diameters on insertion of thoracic pedicle screws: A biomechanical analysis

被引:61
作者
Kuklo, TR
Lehman, RA
机构
[1] Walter Reed Army Med Ctr, Dept Orthopaed Surg & Rehabil, Washington, DC 20307 USA
[2] Uniformed Serv Univ Hlth Sci, Dept Surg, Bethesda, MD 20814 USA
[3] Union Mem Hosp, Baltimore, MD USA
关键词
biomechanics; tapping; bone mineral density; insertional torque; pedicle screw; thoracic spine; spinal fixation;
D O I
10.1097/01.BRS.0000084665.31967.02
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Study Design. A biomechanical cadaver study to assess the effect of various tapping diameters on thoracic pedicle screw insertional torque. Summary of Background Data. Thoracic pedicle screws are now commonly used for deformity and nondeformity cases. The optimal insertion techniques, however, have not been determined. Purpose. To investigate the effect of various tapping techniques before insertion of thoracic pedicle screws in terms of maximal insertional torque (MIT) or screw pullout. Materials and Methods. Thirty-four fresh cadaveric thoracic vertebrae were harvested and evaluated with dual-energy radiograph absorptiometry (DEXA) to assess bone mineral density (BMD). Twenty-three matched, fixed head, 5.0-mm pedicle screws (group 1) were placed using the straight-forward (ST) trajectory (paralleling the endplate) at various thoracic levels after random side selection using either line-to-line tapping (5.0-mm tap) or 1-mm undertapping (4.0-mm tap) under direct and fluoroscopic visualization. After this, 11 matched 5.0-mm pedicle screws (group 2) were placed comparing undertapping by 0.5-mm (4.5-mm tap) with 1 mm undertapping (4.0-mm tap). MIT was recorded for each screw revolution with a digital torque wrench. Results. BMD averaged 0.732 g/cm(2) (0.620-0.884 g cm(2)) for group 1, and 614 g/cm(2) (0.533-0.697 g/cm(2)) for group 2. In group 1, the average MIT was 0.153+/-0.009 (SE) Nm for line-to-line taping and 0.295+/-0.021 (SE) Nm for 1 mm undertapping, a 93% increase in MIT (P<0.0005). In group 2, the average MIT was 0.138±0.009 (SE) Nm for 0.5 mm undertapping and 0.202±0.018 (SE) Nm for undertapping by 1 mm, a 47% increase in MIT (P=0.03). BMD correlated with undertapping by 1 mm in group 1 (P<0.0005), but not with undertapping by 0.5 mm (P=0.087), although there appeared to be a trend in osteoporotic specimens. There were no noted differences in MIT between thoracic regions/levels, despite small differences in thoracic pedicle widths (P=0.193). Discussion and Conclusion. Undertapping the thoracic pedicle by 1-mm increases MIT by 47% (P=0.03) when compared with undertapping by 0.5 mm, and by 93% (P<0.0005) when compared with tapping line-to-line.
引用
收藏
页码:2066 / 2071
页数:6
相关论文
共 26 条
[1]  
BANTA CJ, 1989, ORTHOPEDICS, V12, P939
[2]   THE EFFECTS OF PEDICLE SCREW FIT - AN IN-VITRO STUDY [J].
BRANTLEY, AGU ;
MAYFIELD, JK ;
KOENEMAN, JB ;
CLARK, KR .
SPINE, 1994, 19 (15) :1752-1758
[3]  
CARLSON GD, 1992, SPINE S, V17, P196
[4]   Factors affecting the pullout strength of cancellous bone screws [J].
Chapman, JR ;
Harrington, RM ;
Lee, KM ;
Anderson, PA ;
Tencer, AF ;
Kowalski, D .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1996, 118 (03) :391-398
[5]   CORRELATIONS BETWEEN SCREW HOLE PREPARATION, TORQUE OF INSERTION, AND PULLOUT STRENGTH FOR SPINAL SCREWS [J].
DAFTARI, TK ;
HORTON, WC ;
HUTTON, WC .
JOURNAL OF SPINAL DISORDERS, 1994, 7 (02) :139-145
[6]  
GEORGE DC, 1991, SPINE, V16, P181, DOI 10.1097/00007632-199116020-00015
[7]   EFFECTS OF BONE-MINERAL DENSITY ON PEDICLE SCREW FIXATION [J].
HALVORSON, TL ;
KELLEY, LA ;
THOMAS, KA ;
WHITECLOUD, TS ;
COOK, SD .
SPINE, 1994, 19 (21) :2415-2420
[8]   Structural characteristics of the pedicle and its role in screw stability [J].
Hirano, T ;
Hasegawa, K ;
Takahashi, HE ;
Uchiyama, S ;
Hara, T ;
Washio, T ;
Sugiura, T ;
Yokaichiya, M ;
Ikeda, M .
SPINE, 1997, 22 (21) :2504-2509
[9]  
KRAG MH, 1986, CLIN ORTHOP RELAT R, P75
[10]  
Krag MH, 1991, SPINE S, V16, p84s