Magnetic resonance imaging interactions with a sacral neuromodulation system

被引:4
|
作者
Huang, Xuechen [1 ]
Jiang, Guangqiang [1 ]
机构
[1] Axonics Inc, 26 Technol Dr, Irvine, CA 92618 USA
关键词
implantable neurostimulator; MRI physics; 1.5; TESLA; PACEMAKER; SAFETY; MRI; FAILURES; PATIENT; RISKS; LEADS;
D O I
10.1002/nau.24756
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Aims Sacral neuromodulation (SNM) has successfully treated patients with functional urinary and/or bowel disorders for more than two decades. Historically, patients with the InterStim system (Medtronic) were contraindicated for Magnetic Resonance Imaging (MRI) scans. In 2012, Medtronic obtained Food and Drug Administration (FDA) approval for allowing 1.5 Tesla (T) MRI head scans. In September 2019, the Axonics System (Axonics) received FDA approval for 1.5 T full-body MR Conditional labeling and then 3 T full-body MR Conditional labeling in July 2020. In August 2020, Medtronic received 1.5 and 3 T full-body MR Conditional labeling from the FDA for their new SNM systems (InterStim II and Micro devices with SureScan(TM) leads). With the advancements in MRI technology and availability of full-body MRI eligible SNM systems, it is important for physicians to better understand MRI safety for these systems. Methods This paper explains the fundamentals of MRI physics, its interactions with active implantable medical devices (AIMDs), the subsequent potential safety hazards with emphasis on radio frequency (RF)-related safety, and the risks associated with "Off-label" scans, including abandoned and broken leads. Results MRI guidelines provided by the AIMD device manufacturer should be followed to ensure MRI scan safety and avoid any unnecessary risk to patients. Conclusions MRI guidelines provided by the device manufacturer are the best resource for guidance for performing safe MRI scanning. Specific conditions should be fully understood and generalizations on MRI safety claims based on partial analysis or case studies should be avoided.
引用
收藏
页码:1862 / 1868
页数:7
相关论文
共 50 条
  • [1] REMOVAL OF SACRAL NEUROMODULATION DEVICES FOR MAGNETIC RESONANCE IMAGING
    Quirouet, Adrienne
    Goldman, Howard
    NEUROUROLOGY AND URODYNAMICS, 2015, 34 : S59 - S60
  • [2] Safety of Magnetic Resonance Imaging in patients under Sacral Neuromodulation with an InterStim Neuromodulator
    Karrer-Warzinek, Elisabeth
    Abt, Dominik
    Kim, Olaf Chan-Hi
    Schmid, Hans-Peter
    Engeler, Daniel Stephan
    Muellhaupt, Gautier
    UROLOGY, 2021, 154 : 115 - 119
  • [3] CHANGES IN BRAIN ACTIVITY ON FUNCTIONAL MAGNETIC RESONANCE IMAGING DURING SACRAL NEUROMODULATION FOR OVERACTIVE BLADDER
    Pizarro-Berdichevsky, J.
    Gill, B.
    Bhattacharyya, P.
    Marks, B.
    Quirouet, A.
    Vasavada, S.
    Jones, S. E.
    Goldman, H.
    NEUROUROLOGY AND URODYNAMICS, 2017, 36 : S176 - S177
  • [4] Waiting for Godot: What to Do When a Patient with a Sacral Neuromodulation Device Needs Magnetic Resonance Imaging
    Siegel, Steven W.
    JOURNAL OF UROLOGY, 2018, 200 (05): : 954 - 955
  • [5] Sacral Neuromodulation Device Heating During Lumbar and Pelvic Magnetic Resonance Imaging-a Phantom Study
    Quirouet, Adrienne
    Bhattacharyya, Pallab K.
    Dielubanza, Elodi J.
    Gill, Bradley C.
    Jones, Stephen E.
    Goldman, Howard B.
    UROLOGY, 2017, 107 : 61 - 66
  • [6] A functional magnetic resonance imaging study of the effect of sacral neuromodulation on brain responses in women with Fowler's syndrome
    Kavia, Rajesh
    DasGupta, Ranan
    Critchley, Hugo
    Fowler, Clare
    Griffiths, Derek
    BJU INTERNATIONAL, 2010, 105 (03) : 366 - 372
  • [7] Can Lumbosacral Magnetic Resonance Imaging be Performed Safely in Patients with a Sacral Neuromodulation Device? An In Vivo Prospective Study
    Guzman-Negron, Juan M.
    Pizarro-Berdichevsky, Javier
    Gill, Bradley C.
    Goldman, Howard B.
    JOURNAL OF UROLOGY, 2018, 200 (05): : 1088 - 1092
  • [8] Can lumbosacralmagnetic resonance imaging be performed safely in patients with a sacral neuromodulation device?
    Pizarro-Berdichevsky, J.
    Guzman-Negron, J. M.
    Gill, B.
    Goldman, H. B.
    INTERNATIONAL UROGYNECOLOGY JOURNAL, 2018, 29 : S30 - S31
  • [9] HEATING OF THE INTERSTIM SACRAL NEUROMODULATION DEVICE IN A SIMULATED PHANTOM MODEL DURING LUMBAR AND PELVIC MAGNETIC RESONANCE IMAGING (MRI).
    Quirouet, Adrienne
    Jones, Stephen
    Bhattacharyya, Pallab
    Goldman, Howard
    JOURNAL OF UROLOGY, 2015, 193 (04): : E39 - E40
  • [10] CAN LUMBOSACRAL MAGNETIC RESONANCE IMAGING BE PERFORMED SAFELY IN PATIENTS WITH A SACRAL NEUROMODULATION DEVICE? AN IN-VIVO PROSPECTIVE STUDY
    Guzman-Negron, J. M.
    Pizarro-Berdichevsky, J.
    Gill, B. C.
    Goldman, H. B.
    NEUROUROLOGY AND URODYNAMICS, 2018, 37 : S285 - S285