Dosimetric properties of equivalent-quality flattening filter-free (FFF) and flattened photon beams of Versa HD linear accelerator

被引:8
|
作者
Meshram, Mukesh N. [1 ]
Pramanik, Srimanta [1 ]
Ranjith, C. P. [1 ]
Gopal, Saravana K. [1 ]
Dobhal, Rishabh [1 ]
机构
[1] CIMS Hosp Pvt Ltd, Ahmadabad 380060, Gujarat, India
来源
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS | 2016年 / 17卷 / 03期
关键词
equivalent-quality beam; flattening filter free; dosimetric properties; photon beam data; Versa HD linac; FREE CLINICAL ACCELERATOR; MONTE-CARLO; COLLIMATOR; THERAPY;
D O I
10.1120/jacmp.v17i3.6173
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
This study presents the basic dosimetric properties of photon beams of a Versa HD linear accelerator (linac), which is capable of delivering flattening filter-free (FFF) beams with a beam quality equivalent to the corresponding flattened beams based on comprehensive beam data measurement. The analyzed data included the PDDs, profiles, penumbra, out-of-field doses, surface doses, output factors, head and phantom scatter factors, and MLC transmissions for both FFF and flattened beams of 6 MV and 10 MV energy from an Elekta Versa HD linac. The 6MVFFF and 10MVFFF beams had an equivalent mean energy to the flattened beams and showed less PDD variations with the field sizes. Compared with their corresponding flattened beams, D-max was deeper for FFF beams for all field sizes; the ionization ratio variations with the field size were lower for FFF beams; the out-of-field doses were lower and the penumbras were sharper for the FFF beams; the off-axis profile variations with the depths were lesser for the FFF beams. Further, the 6MVFFF and 10MVFFF beams had 35.7% and 40.9% less variations in output factor with the field size, respectively. The collimator exchange effect was reduced in the FFF mode. The head scatter factor showed 59.1% and 73.6% less variations, on average, for the 6MVFFF and 10MVFFF beams, respectively; the variations in the phantom scatter factor were also smaller. The surface doses for all beams increased linearly with the field size. The 6MVFFF and 10MVFFF beams had higher surface doses than the corresponding flattened beams for field sizes of up to 10x10cm(2) but had lower surface doses for larger fields. Both FFF beams had lower average MLC transmissions than the flattened beams. The finding that the FFF beams were of equivalent quality to the corresponding flattened beams indicates a significant difference from the data on unmatched FFF beams.
引用
收藏
页码:358 / 370
页数:13
相关论文
共 50 条
  • [21] Commissioning Measurements of Flattening Filter and Flattening Filter Free Photon Beams Using a TrueBeam Stx® Linear Accelerator
    Saidi K.
    El Baydaoui R.
    El Gouach H.
    Mesradi M.R.
    Naim A.
    Iranian Journal of Medical Physics, 2021, 18 (01) : 49 - 62
  • [22] Secondary MU Calculation for Flattening Filter Free (FFF) Photon Beams
    Kim, M.
    Zhu, T.
    MEDICAL PHYSICS, 2015, 42 (06) : 3495 - 3495
  • [23] The Use of Photon Beams of a Flattening Filter-free Linear Accelerator for Hypofractionated Volumetric Modulated Arc Therapy in Localized Prostate Cancer
    Zwahlen, Daniel R.
    Lang, Stephanie
    Hrbacek, Jan
    Glanzmann, Christoph
    Kloeck, Stephan
    Najafi, Yousef
    Streller, Tino
    Studer, Gabriela
    Zaugg, Kathrin
    Luetolf, Urs M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (05): : 1655 - 1660
  • [24] Definition of parameters for quality assurance of flattening filter free (FFF) photon beams in radiation therapy
    Fogliata, A.
    Garcia, R.
    Knoos, T.
    Nicolini, G.
    Clivio, A.
    Vanetti, E.
    Khamphan, C.
    Cozzi, L.
    MEDICAL PHYSICS, 2012, 39 (10) : 6455 - 6464
  • [25] When to Use Flattening Filter Free(FFF) Mode for Elekta Versa HD Delivered with VMAT
    Zhang, R.
    Zhang, W.
    Gao, Y.
    Bai, W.
    Dong, Z.
    Wang, B.
    Zhang, H.
    Zhao, D.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [26] Treatment vault shielding for a flattening filter-free medical linear accelerator
    Kry, Stephen F.
    Howell, Rebecca M.
    Polf, Jerimy
    Mohan, Radhe
    Vassiliev, Oleg N.
    PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (05): : 1265 - 1273
  • [27] Comparison between flattening filter-free (FFF) and flattened photon beam VMAT plans for the whole brain radiotherapy (WBRT) with hippocampus sparing
    Ji, Tianlong
    Sun, Lu
    Cai, Feng
    Li, Guang
    ASIA-PACIFIC JOURNAL OF CLINICAL ONCOLOGY, 2022, 18 (05) : E263 - E267
  • [28] Dosimetric Comparison of Unmatched Flattening Filter-free and Flattened Beams in Volumetric Arc Therapy Plans for Head-and-neck Cancer
    Kumar, Arvind
    Sharma, Kiran
    Shan, Chandi Prasad
    Garg, Abhishek
    JOURNAL OF MEDICAL PHYSICS, 2023, 48 (04) : 338 - 344
  • [29] Effect of dose-delivery time for flattened and flattening filter-free photon beams based on microdosimetric kinetic model
    Nakano, Hisashi
    Kawahara, Daisuke
    Ono, Kaoru
    Akagi, Yukio
    Hirokawa, Yutaka
    PLOS ONE, 2018, 13 (11):
  • [30] Dosimetric Impact of Intrafraction Motion During RapidArc Stereotactic Vertebral Radiation Therapy Using Flattened and Flattening Filter-Free Beams
    Ong, Chin Loon
    Dahele, Max
    Cuijpers, Johan P.
    Senan, Suresh
    Slotman, Ben J.
    Verbakel, Wilko F. A. R.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 86 (03): : 420 - 425