Quantification of measurement error effects on conductivity reconstruction in electrical impedance tomography

被引:2
|
作者
Sun, Xiang [1 ]
Lee, Eunjung [1 ]
Choi, Jung-Il [1 ]
机构
[1] Yonsei Univ, Dept Computat Sci & Engn, Seoul 03722, South Korea
基金
新加坡国家研究基金会;
关键词
Uncertainty quantification; electrical impedance tomography; polynomial chaos; measurement error; sensitivity analysis; POLYNOMIAL CHAOS; SENSITIVITY-ANALYSIS; ELECTRODE MODELS; PROPAGATION; BOUNDARY; REGULARIZATION; UNCERTAINTY; INFORMATION; COLLOCATION;
D O I
10.1080/17415977.2020.1762595
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
Electrical impedance tomography (EIT) is a boundary measurement inverse technique targeting reconstruction of the conductivity distribution of the interior of a physical body based on boundary measurement data. Typically, the measured data are uncertain because of various error sources; thus, there are many uncertainties in the reconstructed image. This study attempts to quantify the effects of these measurement errors on EIT reconstruction. A comprehensive framework that combines uncertainty quantification techniques and EIT reconstruction techniques is proposed. In this framework, a polynomial chaos expansion method is used to construct a surrogate model of the conductivity field with respect to the measurement errors. Two shape detection indices are introduced to show the EIT reconstruction quality. Finally, under certain detection index constraints, statistical and sensitivity analyses are performed using the properties of the surrogate model. Several EIT problems are examined in this study, involving one or two anomalies in a circular domain or two asymmetric anomalies in a body-like domain. The results show that the proposed framework can quantify the effects of measurement errors on EIT reconstruction at reasonable cost. Further, for the test cases, the measurement errors at the electrodes close to the anomalies are shown to have the greatest influence on the image reconstruction.
引用
收藏
页码:1669 / 1693
页数:25
相关论文
共 50 条
  • [31] Shape Reconstruction Using Boolean Operations in Electrical Impedance Tomography
    Liu, Dong
    Gu, Danping
    Smyl, Danny
    Deng, Jiansong
    Du, Jiangfeng
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2020, 39 (09) : 2954 - 2964
  • [32] A Conditional Diffusion Model for Electrical Impedance Tomography Image Reconstruction
    Shi, Shuaikai
    Kang, Ruiyuan
    Liatsis, Panos
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2025, 74
  • [33] On the measurement of conductivity distribution of the human head using magnetic resonance electrical impedance tomography
    Gao, N
    Zhu, SN
    He, B
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 4443 - 4446
  • [34] A direct reconstruction algorithm for electrical impedance tomography
    Mueller, JL
    Siltanen, S
    Isaacson, D
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (06) : 555 - 559
  • [35] Sparsity Constrained Reconstruction for Electrical Impedance Tomography
    Theertham, Ganesh Teja
    Varanasi, Santhosh Kumar
    Jampana, Phanindra
    IFAC PAPERSONLINE, 2020, 53 (02): : 362 - 367
  • [36] A hybrid reconstruction algorithm for electrical impedance tomography
    Hu, Li
    Wang, Huaxiang
    Zhao, Bo
    Yang, Wuqiang
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2007, 18 (03) : 813 - 818
  • [37] Adaptive techniques in electrical impedance tomography reconstruction
    Li, Taoran
    Isaacson, David
    Newell, Jonathan C.
    Saulnier, Gary J.
    PHYSIOLOGICAL MEASUREMENT, 2014, 35 (06) : 1111 - 1124
  • [38] An Algorithm for Cavity Reconstruction in Electrical Impedance Tomography
    FENG TIAN-HONG AND MA FU-MING (School of Mathematics
    Communications in Mathematical Research, 2011, 27 (03) : 278 - 287
  • [39] Reconstruction of conductivity distribution variation with Tikhonov and wavelet frame combined method for electrical impedance tomography
    Shi, Yanyan
    Zhang, Yuhang
    Wang, Meng
    Lou, Yajun
    Zheng, Shuo
    TRANSACTIONS OF THE INSTITUTE OF MEASUREMENT AND CONTROL, 2023, 45 (14) : 2658 - 2668
  • [40] A Parametric Level Set Method for Electrical Impedance Tomography
    Liu, Dong
    Khambampati, Anil Kumar
    Du, Jiangfeng
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2018, 37 (02) : 451 - 460