Swept-source endoscopic optical coherence tomography real-time imaging system based on GPU acceleration for axial megahertz high-speed scanning

被引:0
|
作者
Lyu, J. [1 ,2 ,3 ]
Ren, L. [4 ,5 ]
Liu, Q-Y [2 ,3 ]
Wang, Y. [2 ,3 ]
Zhou, Z-Q [1 ,2 ,3 ]
Chen, Y-Y [2 ,3 ]
Jia, H-B [1 ,2 ,3 ]
Tang, Y-G [1 ,2 ,3 ]
Li, M. [1 ,2 ,3 ]
机构
[1] Univ Sci & Technol China, Sch Biomed Engn Suzhou, Div Life Sci & Med, Suzhou, Peoples R China
[2] Chinese Acad Sci, Suzhou Inst Biomed Engn & Technol, Suzhou, Peoples R China
[3] Chinese Acad Sci, Suzhou Inst Biomed Engn & Technol, Jiangsu Key Lab Med Opt, Suzhou, Peoples R China
[4] Guangxi Univ, Sch Phys Sci & Technol, Nanning, Peoples R China
[5] Guangxi Univ, Adv Inst Brain & Intelligence, Nanning, Peoples R China
关键词
Swept-source; Optical coherence tomography; GPU acceleration; Axial megahertz scanning; Field-programmable gate array; Suppress the image drift; EN-FACE OCT; ANGIOGRAPHY; SEGMENTATION; DISTORTION; DESIGN;
D O I
暂无
中图分类号
R9 [药学];
学科分类号
1007 ;
摘要
OBJECTIVE: In order to solve the problem of image real-time processing and correction for high-speed endoscopic swept-source optical coherence tomography (SS-OCT), we highly optimize a computer-unified device architecture-based platform and use a field-programmable gate array to summarize the application experience. MATERIALS AND METHODS: We use the Half-Sync/Half-Asyn mode to optimize memory in order to build a high-throughput data thread pool for CPU. We use asynchronous streaming architecture to multiplex multiple threads at high speed to accelerate data processing. At the same time, we design a rotary scanning position information encoding feedback module to suppress image drift, which can realize 25ns logic-timing sequence synchronization control through FPGA 40MHz clock. RESULTS: The maximum complete attainable axial-scan-processing rate (including memory transfer and display of B-scan frames) is 3.52 MHz for a 16-bit pixel depth and A-scans/s of 1024 pixels. To our knowledge, this is the fastest processing rate reported to date with a single-chip graphical processing unit for SS-OCT. Finally, the established high-speed SS-OCT is used to image mouse esophagus and human fingers, and the output images are stable. When the image size is 1024 x 1024 pixels, the real-time imaging rate is 200 frames per second. CONCLUSIONS: This paper develops a real-time image processing and reconstruction technology suitable for high-throughput SS-OCT systems, which can have high-density operation and efficient parallelism, while suppressing high-speed image drift. It lays the foundation for the non-destructive, in vivo, non-staining, fast and convenient early tumor diagnosis of high-speed endoscopic SS-OCT.
引用
收藏
页码:7349 / 7358
页数:10
相关论文
共 50 条
  • [1] Real-time and high-performance calibration method for high-speed swept-source optical coherence tomography
    Azimi, Ehsan
    Liu, Bin
    Brezinski, Mark E.
    JOURNAL OF BIOMEDICAL OPTICS, 2010, 15 (01)
  • [2] High-speed imaging of human retina in vivo with swept-source optical coherence tomography
    Lim, H.
    Mujat, M.
    Kerbage, C.
    Lee, E. C. W.
    Chen, Y.
    Chen, Teresa C.
    de Boer, J. F.
    OPTICS EXPRESS, 2006, 14 (26): : 12902 - 12908
  • [3] DEVELOPMENT OF HIGH-SPEED SWEPT-SOURCE OPTICAL COHERENCE TOMOGRAPHY SYSTEM AT 1320 nm
    Wu, Tong
    Ding, Zhihua
    Chen, Minghui
    Xu, Lei
    Shi, Guohua
    Zhang, Yudong
    JOURNAL OF INNOVATIVE OPTICAL HEALTH SCIENCES, 2009, 2 (01) : 117 - 122
  • [4] Generic real-time uniform K-space sampling method for high-speed swept-Source optical coherence tomography
    Xi, Jiefeng
    Huo, Li
    Li, Jiasong
    Li, Xingde
    OPTICS EXPRESS, 2010, 18 (09): : 9511 - 9517
  • [5] Normalization detection method for high-speed swept-source optical coherence tomography
    Moon, Sucbei
    Kim, Dug Young
    2007 PACIFIC RIM CONFERENCE ON LASERS AND ELECTRO-OPTICS, VOLS 1-4, 2007, : 175 - 176
  • [6] High-speed KTN optical beam deflector for swept-source optical coherence tomography
    Sasaki, Yuzo
    Okabe, Yuichi
    Ueno, Masahiro
    Sakamoto, Takashi
    Toyoda, Seiji
    Kobayashi, Junya
    Yagi, Shogo
    Naganuma, Kazunori
    2013 IEEE 3RD CPMT SYMPOSIUM JAPAN (ICSJ 2013), 2013,
  • [7] Concept for high speed vocal cord imaging with swept-source optical coherence tomography
    Zabic, Miroslav
    Sharifpourboushehri, Sara
    Mueller-Wirts, Lennart
    Benecke, Hannes
    Heisterkamp, Alexander
    Meyer, Heiko
    Ripken, Tammo
    OPTICAL IMAGING, THERAPEUTICS, AND ADVANCED TECHNOLOGY IN HEAD AND NECK SURGERY AND OTOLARYNGOLOGY 2019, 2019, 10853
  • [8] Real-time in vivo imaging by high-speed spectral optical coherence tomography
    Wojtkowski, M
    Bajraszewski, T
    Targowski, P
    Kowalczyk, A
    OPTICS LETTERS, 2003, 28 (19) : 1745 - 1747
  • [9] Megahertz all-optical swept-source optical coherence tomography based on broadband amplified optical time-stretch
    Xu, Jingjiang
    Zhang, Chi
    Xu, Jianbing
    Wong, K. K. Y.
    Tsia, K. K.
    OPTICS LETTERS, 2014, 39 (03) : 622 - 625
  • [10] Subgingival calculus imaging based on swept-source optical coherence tomography
    Hsieh, Yao-Sheng
    Ho, Yi-Ching
    Lee, Shyh-Yuan
    Lu, Chih-Wei
    Jiang, Cho-Pei
    Chuang, Ching-Cheng
    Wang, Chun-Yang
    Sun, Chia-Wei
    JOURNAL OF BIOMEDICAL OPTICS, 2011, 16 (07)