Tubing Erosion of an Inflatable Penile Prosthesis Long after Implantation

被引:5
|
作者
Morales, Alvaro [1 ]
机构
[1] Queens Univ, Dept Urol, Kingston, ON K7M 6R4, Canada
来源
SEXUAL MEDICINE | 2014年 / 2卷 / 02期
关键词
Inflatable Penile Prosthesis; Complications; Erosion; Tubal Erosion; COMPLICATIONS; RELIABILITY; EXPERIENCE; RESERVOIR;
D O I
10.1002/sm2.30
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction. Erosion through skin of connecting tubing of an inflatable penile prosthesis (IPP) has not been previously reported. Aim. The aim of this study was to present a case of tubing erosion, review the pertinent literature, and discuss the possible causes and management options, including preservation of the device and its components. Methods. A 42-year-old male failing to respond to medical treatment for erectile dysfunction underwent insertion of anAMS700 IPP in 1986. Six years later, a revision was necessary because of a leak in the right cylinder and 4 years after, the pump was replaced. Fourteen years after the original implant, he presented with a portion of the tube connecting the pump to the right cylinder eroding through the skin. There was no infection. The skin area involved was resected and the original pump and tubing were buried in a new scrotal pocket after thorough irrigation. Results. The IPP remained in place, allowing vaginal penetration and without infection for another 11 years. Three years later, it was de-functionalized, converted into a fixed volume device. It eventually was replaced 25 years after originally implanted with a semirigid prosthesis because it did not provide sufficient rigidity and because of concerns about the presence of "screws" detected during pelvic imaging. Conclusions. Mechanical failures in the early IPP models, as illustrated in this case, were expected. However, the long survival of the device is remarkable. Erosion of the connecting tubing through the skin is unique and, under exceptional circumstances, may be managed conservatively without replacing components of the IPP. Clinicians unfamiliar with procedures involving inflatable devices need to be aware of "foreign bodies" visible in radiological examinations in men who have had revisions of an IPP.
引用
收藏
页码:103 / 106
页数:4
相关论文
共 50 条
  • [21] Utility of magnetic resonance imaging in evaluating inflatable penile prosthesis malfunction and complaints
    D D Thiel
    G A Broderick
    M Bridges
    International Journal of Impotence Research, 2003, 15 : S155 - S161
  • [22] Utility of magnetic resonance imaging in evaluating inflatable penile prosthesis malfunction and complaints
    Thiel, DD
    Broderick, GA
    Bridges, M
    INTERNATIONAL JOURNAL OF IMPOTENCE RESEARCH, 2003, 15 (Suppl 5) : S155 - S161
  • [23] Tips and Tricks of Inflatable Penile Prosthesis Reservoir Placement: A Case Presentation and Discussion
    Simon, Ross
    Hakky, Tariq S.
    Henry, Gerard
    Perito, Paul
    Martinez, Daniel
    Parker, Justin
    Carrion, Rafael E.
    JOURNAL OF SEXUAL MEDICINE, 2014, 11 (05): : 1325 - 1333
  • [24] Long Term Outcomes and Couples' Satisfaction with Three-Piece Inflatable Penile Prosthesis Surgery in the Treatment of Erectile Dysfunction
    Efesoy, Ozan
    Ozbay, Engin
    Tek, Mesut
    Akbay, Erdem
    Cayan, Selahittin
    TURKIYE KLINIKLERI TIP BILIMLERI DERGISI, 2010, 30 (06): : 1965 - 1970
  • [25] Management of the complications of penile prosthesis implantation
    Natali, Alessandro
    JOURNAL OF MENS HEALTH, 2010, 7 (03) : 193 - 198
  • [26] Reservoir Placement Considerations During Inflatable Penile Prosthesis Surgery
    Steven J. Hudak
    Current Urology Reports, 2019, 20
  • [27] Reservoir Placement Considerations During Inflatable Penile Prosthesis Surgery
    Hudak, Steven J.
    CURRENT UROLOGY REPORTS, 2019, 20 (02)
  • [28] Inflatable penile prosthesis: Results of 150 cases
    Garber, BB
    BRITISH JOURNAL OF UROLOGY, 1996, 78 (06): : 933 - 935
  • [29] Inflatable Penile Prosthesis Complication in the Emergency Department
    Sedaghat, Farzad
    Kamona, Aws
    HEALTHCARE, 2024, 12 (10)
  • [30] Antibiotic Patterns with Inflatable Penile Prosthesis Insertion
    Wosnitzer, Matthew S.
    Greenfield, Jason M.
    JOURNAL OF SEXUAL MEDICINE, 2011, 8 (05): : 1521 - 1528