Device-less patent foramen ovale closure by radiofrequency thermal energy

被引:0
作者
Walpoth, Nazan B. [1 ]
Habermacher, Kathrin [1 ]
Moarof, Igal [1 ]
Watson, Sandy [1 ]
Wahl, Andreas [1 ]
Windecker, Stephan [1 ]
Schoenenberger, Christa [1 ]
Meier, Bernhard [1 ]
机构
[1] Univ Hosp Bern, Cardiovasc Dept, CH-3010 Bern, Switzerland
关键词
patent foramen ovale; cardiac catheterisation; device-less closure; radiofrequency energy;
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The goal of this study was to assess the feasibility, safety and success of a system which uses radiofrequency energy (RFE) rather than a device for percutaneous closure of patent foramen ovale (PFO). Methods: Sixteen patients (10 men, 6 women, mean age 50 years) were included in the study. All of them had a proven PFO with documented right-to-left shunt (RLS) after Valsalva manoeuvre (VM) during transoesophageal echocardiography (TEE). The patients had an average PFO diameter of 6 +/- 2 mm at TEE and an average of 23 +/- 4 microembolic signals (MES) in power M-mode transcranial Doppler sonography (pm-TCD), measured over the middle cerebral artery. An atrial septal aneurysm (ASA) was present in 7 patients (44%). Balloon measurement, performed in all patients, revealed a stretched PFO diameter of 8 +/- 3 mm. In 2 patients (stretched diameter 11 and 14 mm respectively, both with ASA >10 mm), radiofrequency was not applied (PFO too large) and the PFO was closed with an Amplatzer PFO occluder instead. A 6-month follow-up TEE was performed in all patients. Results: There were no serious adverse events during the procedure or at follow-up (12 months average). TEE 6 months after the first RFE procedure showed complete closure of the PFO in 50% of the patients (7/14). Closure appeared to be influenced by PFO diameter, complete closure being achieved in 89% (7/8) with a balloon-stretched diameter: 7 mm but in none of the patients >7 mm. Only one of the complete closure patients had an ASA. Of the remainder, 4 (29%) had an ASA. Although the PFO was not completely closed in this group, some reduction in the diameter of the PFO and in MES was documented by TEE and pm-TCD with VM. Five of the 7 residual shunt patients received an Amplatzer PFO occluder. Except for one patient with a minimal residual shunt, all showed complete closure of PFO at 6-month follow-up TEE and pm-TCD with VM. The other two refused a closure device. Conclusions: The results confirm that radiofrequency closure of the PFO is safe albeit less efficacious and more complex than device closure. The technique in its current state should not be attempted in patients with a balloon-stretched PFO diameter >7 mm and an ASA.
引用
收藏
页码:108 / 113
页数:6
相关论文
共 50 条
  • [1] Patent Foramen Ovale Stroke and Device Closure
    Suradi, Hussam S.
    Hijazi, Ziyad M.
    CARDIOLOGY CLINICS, 2016, 34 (02) : 231 - +
  • [2] Thrombosis of a patent foramen ovale closure device: Thrombolytic management
    Vanderheyden, M
    Willaert, W
    Claessens, P
    Branders, I
    Ex, P
    Vermeersch, P
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2002, 56 (04) : 522 - 526
  • [3] Patent foramen ovale: diagnostic evaluation and the role of device closure
    Gonnah, Ahmed R.
    Bharadwaj, Mahima S.
    Nassar, Hassan
    Abdelaziz, Hesham K.
    Roberts, David Hesketh
    CLINICAL MEDICINE, 2022, 22 (05) : 441 - 448
  • [4] Patent foramen ovale morphology and impact on percutaneous device closure
    El Said, HG
    McMahon, CJ
    Mullins, CE
    Pignatelli, RH
    Grifka, RG
    Nihill, MR
    Vincent, JA
    PEDIATRIC CARDIOLOGY, 2005, 26 (01) : 62 - 65
  • [5] Patent Foramen Ovale Morphology and Impact on Percutaneous Device Closure
    H.G. El Said
    C.J. McMahon
    C.E. Mullins
    R.H. Pignatelli
    R.G. Grifka
    M.R. Nihill
    J.A. Vincent
    Pediatric Cardiology, 2005, 26 : 62 - 65
  • [6] Pacman Sign during device closure of the patent foramen ovale
    Meier, B
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 60 (02) : 221 - 223
  • [7] Transcatheter closure of patent foramen ovale during a radiofrequency ablation procedure
    Manolis, Antonis S.
    Andrikopoulos, George
    Tsagou, Vassiliki
    Pyrros, John
    Kranidis, Athanasios
    CLINICAL CARDIOLOGY, 2006, 29 (08) : 369 - 371
  • [8] Atrial fibrillation following device closure of patent foramen ovale
    Burow, Annika
    Schwerzmann, Markus
    Wallmann, Dieter
    Tanner, Hildegard
    Sakata, Takao
    Windecker, Stephan
    Meier, Bernhard
    Delacretaz, Etienne
    CARDIOLOGY, 2008, 111 (01) : 47 - 50
  • [9] Xtrace: Novel Bioresorbable Device for Patent Foramen Ovale Closure
    Abu Ajamieh, Sara
    Mindroc-Filimon, Diana
    Mozo, Irene
    Rocha, Isabel
    BIODEVICES: PROCEEDINGS OF THE 13TH INTERNATIONAL JOINT CONFERENCE ON BIOMEDICAL ENGINEERING SYSTEMS AND TECHNOLOGIES, VOL 1: BIODEVICES, 2020, 2020, : 344 - 349
  • [10] Device closure of residual shunt after percutaneous closure of patent foramen ovale
    Majunke, Nicolas
    Wallenborn, Julia
    Baranowski, Andreas
    Wunderlich, Nina
    Sievert, Horst
    EUROINTERVENTION, 2010, 5 (07) : 833 - 837