Optical coherence tomography for shape and radius of curvature measurements of deeply curved machined metallic surfaces: a comparison with two-beam laser interferometry

被引:15
|
作者
Rao, K. Divakar [1 ]
Udupa, D. V. [2 ]
Prathap, C. [1 ]
Rathod, A. [3 ]
Balasubramaniam, R. [3 ]
Sahoo, N. K. [2 ]
机构
[1] Bhabha Atom Res Ctr Facil, Photon & Nanotechnol Sect, Visakhapatnam 530012, Andhra Pradesh, India
[2] Bhabha Atom Res Ctr, Atom & Mol Phys Div, Bombay 400085, Maharashtra, India
[3] Bhabha Atom Res Ctr, Div Precis Engn, Bombay 400085, Maharashtra, India
关键词
Optical coherence tomography; Radius of curvature; Fizeau interferometry; Curved surfaces;
D O I
10.1016/j.optlaseng.2014.09.012
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
An application of optical coherence tomography (OCT) for measuring shape and the radius of curvature of deeply curved metallic surfaces in precision mechanical components is demonstrated. Measurements carried out by OCT technique on concave and convex hemispherical metallic surfaces generated by diamond turning process is presented. The technique is compared with a two-beam laser interferometric method for determinations of shape errors and radius of curvature. It is shown that OCT is a useful tool for three-dimensional shape determinations for deeply curved surfaces such as hemispheres and has advantages over two-beam laser interferometric method. (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:204 / 209
页数:6
相关论文
共 1 条
  • [1] Comparison of radius of anterior lens surface curvature measurements in vivo using the anterior segment optical coherence tomography and Scheimpflug imaging
    Liu, Zhenzhen
    Ruan, Xiaoting
    Wang, Wei
    Liu, Juzi
    Meng, Yijing
    Gu, Xiaoxun
    Fu, Jun
    Luo, Lixia
    Liu, Yizhi
    ANNALS OF TRANSLATIONAL MEDICINE, 2020, 8 (05)