Graphene oxide-regulated low-background aptasensor for the "turn on" detection of tetracycline

被引:17
|
作者
Xu, Jin [1 ]
Qing, Taiping [1 ]
Jiang, Zixin [1 ]
Zhang, Peng [1 ]
Feng, Bo [1 ]
机构
[1] Xiangtan Univ, Coll Environm & Resources, Xiangtan 411105, Hunan, Peoples R China
基金
中国国家自然科学基金;
关键词
Low background; Graphene oxide; Tetracycline; Turn-on; Aptamer; ANTIBIOTICS RESIDUES; APTAMERS; IMMUNOASSAY; ANTIBODIES;
D O I
10.1016/j.saa.2021.119898
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Tetracyclines (TC) are a common antibiotic for using in livestock breeding and healthcare; however, due to the inappropriate application of TCs, more than 75% of TCs are excreted and released into the environment in an active form through human and animal urine and feces, which results in high levels of TCs in the ecological system, causing adverse effects on the food safety and human health. Thus, the high-performance monitoring of TC pollution is necessary. In this work, a highly sensitive fluorescent aptasensor was developed that was based on graphene oxide (GO) regulation of low background signal and target-induced fluorescence restoration. In the absence of analyte, the DNA probe (TC aptamer) was adsorbed completely by GO and failed to enhance the fluorescence of SYBR gold (SG), thereby resulting in a low background signal. When the TC-included samples were added, the DNA probe formed an aptamer-TC complex, thereby separating from the surface of the GO and inducing the fluorescence of SG. Under optimal conditions, the proposed strategy could detect TC concentrations of less than 6.2 x 10(-3) ng mL(-1), which is four orders of magnitude better than the detection limit of the "turn off" mode (53.9511 ng mL(-1)). Moreover, this aptasensor has been used to detect TC from milk samples and wastewater samples, and its satisfactory performances demonstrate that the proposed strategy can be applied in practice for TC monitor in food safety and environmental protection. Therefore, we believe that this work is meaningful in pollution monitoring, environment restoration and emergency treatment. (C) 2021 Elsevier B.V. All rights reserved.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] A graphene oxide-based fluorescent aptasensor for the turn-on detection of epithelial tumor marker mucin 1
    He, Yue
    Lin, Yi
    Tang, Hongwu
    Pang, Daiwen
    NANOSCALE, 2012, 4 (06) : 2054 - 2059
  • [22] A Highly Selective Low-Background Fluorescent Imaging Agent for Nitric Oxide
    Yang, Youjun
    Seidlits, Stephanie K.
    Adams, Michelle M.
    Lynch, Vincent M.
    Schmidt, Christine E.
    Anslyn, Eric V.
    Shear, Jason B.
    JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2010, 132 (38) : 13114 - 13116
  • [23] A Split G-Quadruplex and Graphene Oxide-Based Low-Background Platform for Fluorescence Authentication of Pseudostellaria heterophylla
    Zheng, Zhenzhu
    Hu, Juan
    He, Zhaodong
    SENSORS, 2014, 14 (12) : 22971 - 22981
  • [24] Novel Electrochemical Aptasensor Based on Floral Zinc Oxide/Reduced Graphene Oxide/Colloidal Gold Composite for the Detection of Tetracycline, Doxycycline and Oxytetracycline in Milk
    Guan, Fukai
    Dong, Haowei
    Xiang, Yaodong
    Zhang, Mei
    Huang, Jingcheng
    Wang, Guanjie
    Shen, Zheng
    Xu, Deyan
    Sun, Xia
    Guo, Yemin
    Zhao, Shancang
    JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 2022, 169 (12)
  • [25] A turn-on graphene quantum dot and graphene oxide based fluorometric aptasensor for the determination of telomerase activity
    Kazemi, Elahe
    Bagheri, Habib
    Norouzian, Dariush
    MICROCHIMICA ACTA, 2019, 186 (12)
  • [26] A turn-on graphene quantum dot and graphene oxide based fluorometric aptasensor for the determination of telomerase activity
    Elahe Kazemi
    Habib Bagheri
    Dariush Norouzian
    Microchimica Acta, 2019, 186
  • [27] Low-background X-ray detection with Micromegas for axion research
    Garcia, J. A.
    Dafni, T.
    Diago, A.
    Ferrer-Ribas, E.
    Galan, J.
    Gardikiotis, A.
    Giomataris, I.
    Garza, J. G.
    Iguaz, F. J.
    Irastorza, I. G.
    Ortega, I.
    Papaevangelou, T.
    Rodriguez, A.
    Tomas, A.
    Vafeiadis, T.
    Yildiz, S. C.
    SIXTH SYMPOSIUM ON LARGE TPCS FOR LOW ENERGY RARE EVENT DETECTION, 2013, 460
  • [28] On-chip graphene oxide aptasensor for multiple protein detection
    Ueno, Yuko
    Furukawa, Kazuaki
    Matsuo, Kota
    Inoue, Suzuyo
    Hayashi, Katsuyoshi
    Hibino, Hiroki
    ANALYTICA CHIMICA ACTA, 2015, 866 : 1 - 9
  • [29] A novel technology on infrared multi-band low-background detection
    Zhen Z.
    Wang Y.
    Ou W.
    Zhou J.
    Li A.
    Hongwai yu Jiguang Gongcheng/Infrared and Laser Engineering, 2020, 49 (05):
  • [30] Low-background detection of fission neutrons produced by pulsed neutron interrogation
    Ruddy, Frank H.
    Flammang, Robert W.
    Seidel, John G.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 598 (02): : 518 - 525