Design of a Centrifugal Blood Pump: Heart Turcica Centrifugal

被引:10
|
作者
Demir, Onur [1 ]
Biyikli, Emre [1 ]
Lazoglu, Ismail [1 ]
Kucukaksu, Suha [2 ]
机构
[1] Koc Univ, Dept Mech Engn, Mfg & Automat Res Ctr, TR-34450 Istanbul, Turkey
[2] Florence Nightingale Hosp, Dept Cardiovasc Surg, Istanbul, Turkey
关键词
Blade height profile; Centrifugal blood pump; Design; Left ventricular assist device; Volute tongue profile; COMPUTATIONAL FLUID-DYNAMICS; FLOW VISUALIZATION; NUMERICAL-ANALYSIS; IMPELLER DESIGN; HEMOLYSIS TESTS; ESTABLISH;
D O I
10.1111/j.1525-1594.2010.01167.x
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A prototype of a new implantable centrifugal blood pump system named Heart Turcica Centrifugal (HTC) was developed as a left ventricular assist device (LVAD) for the treatment of end-stage cardiac failure. In the development of HTC, effects of blade height and volute tongue profiles on the hydraulic and hemolytic performances of the pump were investigated. As a result, the prototype was manufactured using the best blade height and volute tongue profiles. Performance of the prototype model was experimentally evaluated in a closed-loop flow system using water as the medium. The hydraulic performance requirement of an LVAD (5 L/min flow rate against a pressure difference of 100 mm Hg) was attained at 2800 rpm rotational speed.
引用
收藏
页码:720 / U138
页数:6
相关论文
共 50 条
  • [41] Remediation of stagnant zones in a centrifugal blood pump
    Miller, Gerald E.
    Holmes, Michelle
    American Society of Mechanical Engineers, Bioengineering Division (Publication) BED, 1999, 42 : 165 - 166
  • [42] Hemocompatibility of a hydrodynamic levitation centrifugal blood pump
    Yamane, Takashi
    Maruyama, Osamu
    Nishida, Masahiro
    Kosaka, Ryo
    Sugiyama, Daisuke
    Miyamoto, Yusuke
    Kawamura, Hiroshi
    Kato, Takahisa
    Sano, Takeshi
    Okubo, Takeshi
    Sankai, Yoshiyuki
    Shigeta, Osamu
    Tsutsui, Tatsuo
    JOURNAL OF ARTIFICIAL ORGANS, 2007, 10 (02) : 71 - 76
  • [43] Analysis of the Blood Flow Through the Centrifugal Pump
    Mikolajczak, Joanna
    Slowinski, Jakub
    Zawislak, Maciej
    PROCEEDINGS OF THE 13TH INTERNATIONAL SCIENTIFIC CONFERENCE: COMPUTER AIDED ENGINEERING, 2017, : 375 - 382
  • [44] Flow visualization studies in a centrifugal blood pump
    Ng, B.T.H.
    Chan, W.K.
    Yu, S.C.M.
    Chua, L.P.
    American Society of Mechanical Engineers, Bioengineering Division (Publication) BED, 1999, 42 : 163 - 164
  • [45] Channel Rotor Calculation for a Centrifugal Blood Pump
    A. P. Kuleshov
    G. P. Itkin
    Biomedical Engineering, 2019, 52 : 311 - 315
  • [46] Hybrid magnetic bearings for a centrifugal blood pump
    Ren, Zhaohui
    Jahanmir, Said
    Heshmat, Hooshang
    Walton, James
    PROCEEDINGS OF THE ASME/STLE INTERNATIONAL JOINT TRIBOLOGY CONFERENCE, PTS A AND B, 2008, : 997 - 999
  • [47] Design of a magnetic bearing-supported prototype centrifugal artificial heart Pump
    Allaire, PE
    Maslen, EH
    Kim, HC
    Bearnson, GB
    Olsen, DB
    TRIBOLOGY TRANSACTIONS, 1996, 39 (03): : 663 - 669
  • [48] Design of a magnetic bearing-supported prototype centrifugal artificial heart pump
    University of Virginia, Mechanical,, Aerospace, and Nuclear Engineering, Department, Charlottesville, Virginia 22901
    Tribol Trans, 3 (663-669):
  • [49] A new blood pump for cardiopulmonary bypass: The HiFlow centrifugal pump
    Gobel, C
    Eilers, R
    Reul, H
    Schwindke, P
    Jorger, M
    Rau, G
    ARTIFICIAL ORGANS, 1997, 21 (07) : 841 - 845
  • [50] A new blood pump for cardiopulmonary bypass: The HiFlow centrifugal pump
    Helmholtz-Inst. for Biomed. Eng., Aachen University of Technology, Aachen, Germany
    不详
    ARTIF. ORGANS, 7 (841-845):