Cryogen-free automated gas chromatograph system for monitoring of halocarbons in the atmosphere at background concentration levels

被引:5
|
作者
Toyoda, S [1 ]
Tominaga, T [1 ]
Makide, Y [1 ]
机构
[1] Univ Tokyo, Radioisotope Ctr, Tokyo 1130032, Japan
关键词
atmospheric concentration; continuous monitoring; halocarbon; chlorofluorocarbon; stratospheric ozone depletion; global warming; gas chromatography; electron capture detector; cryogen-free;
D O I
10.2116/analsci.14.917
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An automated gas chromatograph/electron capture detector (GC/ECD) system is developed for the monitoring of major seven halocarbons in the atmosphere (CCl2F2: CFC-12, CCl3F: CFC-11, CCl2FCClF2: CFC-113, CH3CCl3, CCl4, CHCl=CCl2, and CCl2=CCl2) at the tropospheric background concentration levels. By using two columns in series for each of two groups divided by the volatility of the compounds, the system requires neither cryogen for preconcentration nor any physical/chemical dryer for water removing. A sample amount of ca. 18 ml STP is sufficient to detect these species, and analysis of sub-ambient atmospheric-pressure sample in a canister is also possible. Monitoring of these halocarbon concentrations was carried out at 90-min intervals in the urban area.