63Ni, its half-life and standardization:: Revisited

被引:16
作者
Colle, R. [1 ]
Zimmerman, B. E. [1 ]
Cassette, P. [2 ]
Laureano-Perez, L. [1 ]
机构
[1] NIST, Phys Lab, Gaithersburg, MD 20899 USA
[2] CEA Saclay, Lab Natl Henri Becquerel, F-91191 Gif Sur Yvette, France
关键词
efficiency tracing; liquid scintillation; LNHB; nickel-63; NIST; radioactivity; standards; TDCR;
D O I
10.1016/j.apradiso.2007.07.017
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Recent liquid scintillation (LS) measurements at the National Institute of Standards and Technology (NIST) and at the Laboratoire National Henri Becquerel (LNHB) on a standardized Ni-63 solution that has been tracked for nearly 40 years have resulted in several important findings: (i) a Ni-63 half-life value of 101.2 +/- 1.5 a has been determined with the present decay data. This value is consistent with a previous specific activity determination and with an earlier value from decay measurements; and it appears to be more satisfactory than a recent data evaluator's recommended value of 98.7 a. (ii) All solution standards of Ni-63 as disseminated by NIST for the past 38(+) years are internally consistent with past and recent standardizations. (iii) Primary LS standardizations of Ni-63 by the triple-to-double coincidence ratio (TDCR) method and by CIEMAT/NIST H-3-standard efficiency tracing (CNET) appear to be comparable, although the latter methodology is believed to be inherently inferior. (iv) There is excellent measurement agreement between NIST and LNHB for Ni-63 primary standardizations. (C) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:60 / 68
页数:9
相关论文
共 39 条