Effects of Image Postprocessing in Digital Radiography to Detect Wooden, Soft Tissue Foreign Bodies

被引:0
|
作者
Kleinfelder, Thomas R. [1 ]
Ng, Curtise K. C. [2 ]
机构
[1] Royal Perth Hosp, Perth, WA, Australia
[2] Curtin Univ, Dept Med Radiat Sci, Perth, WA, Australia
关键词
digital radiography; edge enhancement; foreign body; image postprocessing; noise reduction; sharpening; DIAGNOSTIC-ACCURACY; CHEST RADIOGRAPHY; ENHANCEMENT;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To investigate the effects of image postprocessing functions (ie, edge enhancement, noise reduction, and sharpening) that are available on digital radiography systems, including computed radiography (CR) and direct digital radiography (DDR), for detection of wooden, soft tissue foreign bodies. Methods Dorsoplantar and lateral porcine foot radiographs with 4 lengths of wooden foreign bodies (no foreign bodies, 2 mm, 5 mm, and 10 mm) placed 1 mm (superficial) and 1 cm (deep) below the skin were acquired by CR and DDR systems using 10 exposure factors. Images were postprocessed to produce 960 images, including original CR, original DDR, sharpened CR, sharpened DDR, edge-enhanced DDR, and noised-reduced DDR images. Contrast-to-noise ratios (CNR) were used for objective assessments of foreign body visibility on the images. Six Australian radiologic technologists were recruited to review selected images. Australia allows radiologic technologists to provide initial comments on plain radiographs with the supervision of a radiologist. Technologists rated the visibility of foreign bodies using a 4-point scale to determine diagnostic performances of different image receptor and postprocessing types. Means, standard deviations, analyses of variance, and intraclass correlation coefficients were calculated for statistical analyses. Results Among the CR and DDR images with and without postprocessing, the edge-enhanced DDR images had the highest overall mean CNR value (3.39, P = .003) and sensitivity (35.13%). The sensitivity of the edge-enhanced DDR images for detecting the 10 mm foreign body was 43.33%. Discussion Edge-enhanced DDR can be considered an additional tool for suspected wooden, soft tissue foreign body diagnoses in rural areas where digital radiography is the only available imaging modality. This would allow some patients in rural areas to avoid long-distance travel to access sonography or computed tomography to detect foreign bodies, which could minimize emotional, financial, and social costs. Conclusion This study shows that the image postprocessing function of the DDR system can detect wooden, soft tissue foreign bodies. Edge enhancement, specifically, can improve wooden, soft tissue foreign body detection, especially for large foreign bodies >= 10 mm).
引用
收藏
页码:544 / 554
页数:11
相关论文
共 18 条
  • [1] Computed radiography versus indirect digital radiography for the detection of glass soft-tissue foreign bodies
    Sheridan, N.
    McNulty, J. P.
    RADIOGRAPHY, 2016, 22 (03) : 223 - 227
  • [2] Detection of soft tissue foreign bodies in the presence of soft tissue gas
    Lyon, M
    Brannam, L
    Johnson, D
    Blaivas, M
    Duggal, S
    JOURNAL OF ULTRASOUND IN MEDICINE, 2004, 23 (05) : 677 - 681
  • [3] Ultrasound in the emergency department: Detection of wooden foreign bodies in the soft tissues
    Graham, DD
    JOURNAL OF EMERGENCY MEDICINE, 2002, 22 (01) : 75 - 79
  • [4] Diagnostic Modalities for the Detection of Soft Tissue Foreign Bodies
    Shepherd, Matt
    Lee, Jonathan
    McGahon, Michele C.
    ADVANCED EMERGENCY NURSING JOURNAL, 2007, 29 (04) : 297 - 308
  • [5] Photoacoustic tomography of foreign bodies in soft biological tissue
    Cai, Xin
    Kim, Chulhong
    Pramanik, Manojit
    Wang, Lihong V.
    JOURNAL OF BIOMEDICAL OPTICS, 2011, 16 (04)
  • [6] Photoacoustic Tomography of Foreign Bodies in Soft Biological Tissue
    Cai, Xin
    Kim, Chulhong
    Pramanik, Manojit
    Wang, Lihong V.
    PHOTONS PLUS ULTRASOUND: IMAGING AND SENSING 2010, 2010, 7564
  • [7] DIAGNOSIS OF GASTRO-INTESTINAL FOREIGN BODIES IN DOGS USING DIGITAL (COMPUTED) RADIOGRAPHY
    Palecha, Sakar
    Gahlot, T. K.
    VETERINARY PRACTITIONER, 2010, 11 (01): : 30 - 33
  • [8] BEDSIDE ULTRASOUND FOR THE DETECTION OF SOFT TISSUE FOREIGN BODIES: A CADAVERIC STUDY
    Crystal, Chad S.
    Masneri, David A.
    Hellums, John S.
    Kaylor, David W.
    Young, Scott E.
    Miller, Michael A.
    Levsky, Marc E.
    JOURNAL OF EMERGENCY MEDICINE, 2009, 36 (04) : 377 - 380
  • [9] Scanning Equalization Digital Radiography (SEDR) - effects of exposure equalization on image processing
    Altunbas, MC
    Shaw, CC
    Liu, XM
    Medical Imaging 2005: Physics of Medical Imaging, Pts 1 and 2, 2005, 5745 : 1130 - 1137
  • [10] Radiographic griding of subcutaneous soft tissue metallic foreign bodies in emergency department
    Agarwal, Akhilesh Kr
    De, Utpal
    Ray, Udipta
    Agarwal, Anshu
    Singodia, Pankaj
    INDIAN JOURNAL OF SURGERY, 2010, 72 (02) : 130 - 132