A rapid and simple liquid chromatography-tandem mass spectrometry method for the measurement of testosterone, androstenedione, and dehydroepiandrosterone in human serum

被引:11
作者
Xu, Wenjuan [1 ]
Li, Huijun [1 ]
Guan, Qing [1 ]
Shen, Ying [1 ]
Cheng, Liming [1 ]
机构
[1] Huazhong Univ Sci & Technol, Tongji Med Coll, Tongji Hosp, Dept Lab Med, Wuhan, Hubei, Peoples R China
基金
中国国家自然科学基金;
关键词
clinical application; LC-MS; MS; quantification; serum; steroid; SOLID-PHASE EXTRACTION; LC-MS/MS; REFERENCE INTERVALS; GAS-CHROMATOGRAPHY; STEROIDS; HORMONES; PLASMA; DHEA; 17-ALPHA-HYDROXYPROGESTERONE; DELTA-4-ANDROSTENEDIONE;
D O I
10.1002/jcla.22102
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BackgroundThe assay of androgens plays an important role for the differential diagnosis of androgen-related endocrine diseases. We developed a rapid and simple liquid chromatography tandem mass spectrometry (LC-MS/MS) method with small sample volume and simple sample preparation for simultaneous determination of serum testosterone, androstenedione and dehydroepiandrosterone (DHEA) in adults. MethodsWe spiked 100L of serum with 200L of internal standard solution, which was prepared in methanol, for matrix effect relief and protein precipitation. The obtained supernatant was used directly for LC-MS/MS analysis. ResultsThe validated method exhibits excellent linearity for each analyte withlinear correlation coefficient above 0.992. The total precisions at three concentrations were 15.66%, 8.81% and 4.34% for testosterone, 13.60%, 6.62% and 5.96% for androstenedione, 14.35%, 15.34% and 13.92% for DHEA, respectively. Recovery, carryover, matrix effect and analytical specificity were also performed for well validation of this LC-MS/MS method with satisfied results. Reference intervals (5th-95th centile) were also established for adult men and women. DiscussionThis method offers simpler implement for quantification of steroids in clinical laboratories with minimal manual sample manipulation and minimized sample volume requirements. ConclusionIt is demonstrated that a reliable, simple, sensitive and potential method for steroids assay in clinical laboratories had been successfully developed by LC-MS/MS.
引用
收藏
页数:7
相关论文
共 24 条
[1]   A new reliable sample preparation for high throughput focused steroid profiling by gas chromatography-mass spectrometry [J].
Anizan, Sebastien ;
Bichon, Emmanuelle ;
Monteau, Fabrice ;
Cesbron, Nora ;
Antignac, Jean-Philippe ;
Le Bizec, Bruno .
JOURNAL OF CHROMATOGRAPHY A, 2010, 1217 (43) :6652-6660
[2]   Simultaneous measurement of testosterone, androstenedione and dehydroepiandrosterone (DHEA) in serum and plasma using Isotope-Dilution 2-Dimension Ultra High Performance Liquid-Chromatography Tandem Mass Spectrometry (ID-LC-MS/MS) [J].
Buttler, Rahel M. ;
Martens, Frans ;
Kushnir, Mark M. ;
Ackermans, Mariette T. ;
Blankenstein, Marinus A. ;
Heijboer, Annemieke C. .
CLINICA CHIMICA ACTA, 2015, 438 :157-159
[3]   Rapid quantification of steroid patterns in human serum by on-line solid phase extraction combined with liquid chromatography-triple quadrupole linear ion trap mass spectrometry [J].
Ceglarek, Uta ;
Kortz, Linda ;
Leichtle, Alexander ;
Fiedler, Georg M. ;
Kratzsch, Juergen ;
Thiery, Joachim .
CLINICA CHIMICA ACTA, 2009, 401 (1-2) :114-118
[4]   Urinary metanephrines by liquid chromatography tandem mass spectrometry: Using multiple quantification methods to minimize interferences in a high throughput method [J].
Clark, Zlatuse D. ;
Frank, Elizabeth L. .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2011, 879 (31) :3673-3680
[5]   Serum steroid profiling by isotopic dilution-liquid chromatography-mass spectrometry: Comparison with current immunoassays and reference intervals in healthy adults [J].
Fanelli, Flaminia ;
Belluomo, Ilaria ;
Di Lallo, Valentina D. ;
Cuomo, Gaia ;
De Iasio, Rosaria ;
Baccini, Margherita ;
Casadio, Elena ;
Casetta, Bruno ;
Vicennati, Valentina ;
Gambineri, Alessandra ;
Grossi, Gabriele ;
Pasquali, Renato ;
Pagotto, Uberto .
STEROIDS, 2011, 76 (03) :244-253
[6]   Development and validation of a serum total testosterone liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay calibrated to NIST SRM 971 [J].
French, Deborah .
CLINICA CHIMICA ACTA, 2013, 415 :109-117
[7]   Ultra high performance liquid chromatography tandem mass spectrometry determination and profiling of prohibited steroids in human biological matrices. A review [J].
Gosetti, Fabio ;
Mazzucco, Eleonora ;
Gennaro, Maria Carla ;
Marengo, Emilio .
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES, 2013, 927 :22-36
[8]   Simultaneous determination of 12 steroids by isotope dilution liquid chromatography-photo spray ionization tandem mass spectrometry [J].
Guo, Tiedong ;
Taylor, Robert L. ;
Singh, Ravinder J. ;
Soldin, Steven J. .
CLINICA CHIMICA ACTA, 2006, 372 (1-2) :76-82
[9]   Novel liquid chromatography tandem mass spectrometry (LC-MS/MS) methods for measuring steroids [J].
Keevil, Brian G. .
BEST PRACTICE & RESEARCH CLINICAL ENDOCRINOLOGY & METABOLISM, 2013, 27 (05) :663-674
[10]   Measurement of reference intervals for urinary free adrenal steroid levels in Japanese newborn infants by using stable isotope dilution gas chromatography/mass spectrometry [J].
Koyama, Yuhei ;
Homma, Keiko ;
Miwa, Masayuki ;
Ikeda, Kazushige ;
Murata, Mitsuru ;
Hasegawa, Tomonobu .
CLINICA CHIMICA ACTA, 2013, 415 :302-305