A hybrid method for calculating absolute peak efficiency of germanium detectors

被引:35
作者
Jiang, SH
Liang, JH
Chou, JT
Lin, UT
Yeh, WW
机构
[1] Natl Tsing Hua Univ, Dept Engn & Syst Sci, Hsinchu 30043, Taiwan
[2] Inst Nucl Energy Res, Div Hlth Phys, Lungtan 32500, Taiwan
[3] Taiwan Power Co, Radiat Lab, Taipei, Taiwan
关键词
D O I
10.1016/S0168-9002(98)00563-4
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We propose a hybrid method using the Moens concept of effective solid angle for calculating absolute peak efficiency of germanium detectors for extended sources. The effective solid angle is evaluated according to Monte Carlo integration instead of conventional numerical integration. The proposed hybrid method was tested for point, disk, cylinder, and Marinelli-beaker sources with essentially identical results as compared with those calculated using conventional numerical integration, The off-axis effect of source positions and the absolute peak efficiency for cuboid sources were also analyzed to demonstrate the powerful capability of the proposed hybrid method. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:281 / 292
页数:12
相关论文
共 50 条
  • [21] On the absolute efficiency of large neutron detectors for μCF
    V.R. Bom
    V.V. Filchenkov
    Hyperfine Interactions, 1999, 119 : 365 - 367
  • [22] A METHOD FOR EXPERIMENTAL ABSOLUTE DETERMINATION OF DETECTION EFFICIENCY OF TIME-OF-FLIGHT NEUTRON DETECTORS
    ADELFAWZY, M
    FORTSCH, H
    MITTAG, S
    PILZ, W
    SCHMIDT, D
    SEELIGER, D
    STREIL, T
    KERNENERGIE, 1981, 24 (03): : 107 - 109
  • [23] ON THE PHOTOPEAK EFFICIENCY OF GERMANIUM GAMMA-RAY DETECTORS
    JACKEL, B
    WESTMEIER, W
    PATZELT, P
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1987, 261 (03) : 543 - 548
  • [24] AUTOMATIC METHOD TO RESET PEAK DETECTORS
    MAHLER, Y
    ADLER, D
    ELECTRONIC ENGINEERING, 1978, 50 (609): : 26 - &
  • [25] NUCLEAR REACTION EFFICIENCY CORRECTION FOR SILICON AND GERMANIUM DETECTORS
    MAKINO, MQ
    WADDELL, CN
    EISBERG, RM
    NUCLEAR INSTRUMENTS & METHODS, 1968, 60 (01): : 109 - &
  • [26] PROPERTIES OF INTRINSIC GERMANIUM DETECTORS AT LOW ENERGIES BY ESCAPE PEAK SPECTROSCOPY
    ROSNER, B
    MINGAY, DW
    X-RAY SPECTROMETRY, 1983, 12 (02) : 82 - 86
  • [27] A probabilistic γ-ray tracking method for germanium detectors
    Piqueras, I
    Beck, FA
    Pachoud, E
    Duchêne, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 516 (01) : 122 - 133
  • [28] A new method of calculating the diffraction efficiency for diffraction/refraction infrared hybrid system
    Wang, Tao
    Bai, Jian
    Liang, Yiyong
    Hou, Xiyun
    HOLOGRAPHY, DIFFRACTIVE OPTICS, AND APPLICATIONS IV, 2010, 7848
  • [29] Modifications of the Method for Calculating Absolute Drug Bioavailability
    Jiang, Wei
    Yang, Juan-juan
    Cao, Lei
    Xiao, Xue
    Shi, Xiao-lian
    Cao, Yong-xiao
    JOURNAL OF PHARMACY AND PHARMACEUTICAL SCIENCES, 2016, 19 (02): : 181 - 187
  • [30] Mathematical method to calculate full-energy peak efficiency of detectors based on transfer technique
    Gouda, M. M.
    Hamzawy, A.
    Badawi, M. S.
    El-Khatib, A. M.
    Thabet, A. A.
    Abbas, M. I.
    INDIAN JOURNAL OF PHYSICS, 2016, 90 (02) : 201 - 210