Experimental evaluation of pencil beam algorithm by measurements of dose distributions of protons traversing an L-shaped phantom

被引:9
作者
Kohno, R [1 ]
Takada, Y
Sakae, T
Nohtomi, A
Terunuma, T
Yasuoka, K
机构
[1] Univ Tsukuba, Inst Appl Phys, Tsukuba, Ibaraki 305, Japan
[2] Univ Tsukuba, Inst Clin Med, Tsukuba, Ibaraki 305, Japan
[3] Univ Tsukuba, Proton Med Res Ctr, Tsukuba, Ibaraki 3050801, Japan
[4] Univ Tsukuba, Inst Basic Med Sci, Tsukuba, Ibaraki 305, Japan
来源
JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS | 2001年 / 40卷 / 01期
关键词
proton; dose distribution; pencil-beam algorithm; broad-beam algorithm; silicon semiconductor detector;
D O I
10.1143/JJAP.40.441
中图分类号
O59 [应用物理学];
学科分类号
摘要
A proton dose-calculation method has been developed on the basis of the pencil-beam algorithm (PBA), which is expected to improve the precision of the conventional method based on the broad-beam algorithm (BBA). In order to verify the accuracy of calculation by the PEA, dose distributions formed by a horizontal proton beam traversing a phantom with an L-shaped horizontal cross section were measured using a silicon semiconductor detector (SSD) in a water vessel. The results of the measured dose distributions agree well with the ones calculated casing the PEA within an rms error of 2.3%. Therefore, the dose-calculation method by the PEA is useful and applicable to actual treatment planning of the proton therapy.
引用
收藏
页码:441 / 445
页数:5
相关论文
共 10 条
[1]  
CARLI C, 1991, EULIMA FEASIBILI FEB, P160
[2]   Monte Carlo and analytical calculation of proton pencil beams for computerized treatment plan optimization [J].
Carlsson, AK ;
Andreo, P ;
Brahme, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1997, 42 (06) :1033-1053
[3]  
GOITEIN M, 1978, Medical Physics (Woodbury), V5, P258, DOI 10.1118/1.594507
[4]  
GOITEIN M, 1978, Medical Physics (Woodbury), V5, P265, DOI 10.1118/1.594509
[5]   MULTIPLE COULOMB SCATTERING OF 160 MEV PROTONS [J].
GOTTSCHALK, B ;
KOEHLER, AM ;
SCHNEIDER, RJ ;
SISTERSON, JM ;
WAGNER, MS .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 1993, 74 (04) :467-490
[6]   A pencil beam algorithm for proton dose calculations [J].
Hong, L ;
Goitein, M ;
Bucciolini, M ;
Comiskey, R ;
Gottschalk, B ;
Rosenthal, S ;
Serago, C ;
Urie, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (08) :1305-1330
[7]   DIFFERENTIAL-PENCIL-BEAM DOSE CALCULATIONS FOR CHARGED-PARTICLES [J].
PETTI, PL .
MEDICAL PHYSICS, 1992, 19 (01) :137-149
[8]   Dose calculation models for proton treatment planning using a dynamic beam delivery system: an attempt to include density heterogeneity effects in the analytical dose calculation [J].
Schaffner, B ;
Pedroni, E ;
Lomax, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1999, 44 (01) :27-41
[9]   DEGRADATION OF THE BRAGG PEAK DUE TO INHOMOGENEITIES [J].
URIE, M ;
GOITEIN, M ;
HOLLEY, WR ;
CHEN, GTY .
PHYSICS IN MEDICINE AND BIOLOGY, 1986, 31 (01) :1-15
[10]   COMPENSATING FOR HETEROGENEITIES IN PROTON RADIATION-THERAPY [J].
URIE, M ;
GOITEIN, M ;
WAGNER, M .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (05) :553-566