Validation of Ankle Strength Measurements by Means of a Hand-Held Dynamometer in Adult Healthy Subjects

被引:8
作者
Ancillao, Andrea [1 ]
Palermo, Eduardo [1 ]
Rossi, Stefano [2 ]
机构
[1] Sapienza Univ Rome, Dept Mech & Aerosp Engn, Via Eudossiana 18, I-00184 Rome, Italy
[2] Univ Tuscia, Dept Econ & Management Ind Engn DEIM, Via Paradiso 47, I-01100 Viterbo, Italy
关键词
MUSCLE STRENGTH; GAIT PATTERN; RELIABILITY; KNEE; VALIDITY; QUALITY; CHILDREN; POSTURE; TORQUE; MOTION;
D O I
10.1155/2017/5426031
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Uniaxial Hand-Held Dynamometer (HHD) is a low-cost device widely adopted in clinical practice to measure muscle force. HHD measurements depend on operator's ability and joint movements. The aim of the work is to validate the use of a commercial HHD in both dorsiflexion and plantarflexion ankle strength measurements quantifying the effects of HHD misplacements and unwanted foot's movements on the measurements. We used an optoelectronic system and a multicomponent load cell to quantify the sources of error in the manual assessment of the ankle strength due to both the operator's ability to hold still the HHD and the transversal components of the exerted force that are usually neglected in clinical routine. Results showed that foot's movements and angular misplacements of HHD on sagittal and horizontal planes were relevant sources of inaccuracy on the strength assessment. Moreover, ankle dorsiflexion and plantarflexion force measurements presented an inaccuracy less than 2% and higher than 10%, respectively. In conclusion, the manual use of a uniaxial HHD is not recommended for the assessment of ankle plantarflexion strength; on the contrary, it can be allowed asking the operator to pay strong attention to the HHD positioning in ankle dorsiflexion strength measurements.
引用
收藏
页数:8
相关论文
共 33 条
  • [1] RELIABILITY OF MEASUREMENTS OF MUSCLE STRENGTH AND VOLUNTARY ACTIVATION USING TWITCH INTERPOLATION
    ALLEN, GM
    GANDEVIA, SC
    MCKENZIE, DK
    [J]. MUSCLE & NERVE, 1995, 18 (06) : 593 - 600
  • [2] Analysis of Knee Strength Measurements Performed by a Hand-Held Multicomponent Dynamometer and Optoelectronic System
    Ancillao, Andrea
    Rossi, Stefano
    Cappa, Paolo
    [J]. IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2017, 66 (01) : 85 - 92
  • [3] Ancillao A, 2015, IEEE INT SYM MED MEA, P595, DOI 10.1109/MeMeA.2015.7145273
  • [4] Linear correlation between fractal dimension of surface EMG signal from Rectus Femoris and height of vertical jump
    Ancillao, Andrea
    Galli, Manuela
    Rigoldi, Chiara
    Albertini, Giorgio
    [J]. CHAOS SOLITONS & FRACTALS, 2014, 66 : 120 - 126
  • [5] An optoelectronic based approach for handwriting capture
    Ancillao, Andrea
    Galli, Manuela
    Vimercati, Sara Laura
    Albertini, Giorgio
    [J]. COMPUTER METHODS AND PROGRAMS IN BIOMEDICINE, 2013, 111 (02) : 357 - 365
  • [6] Measuring muscular strength of the lower limbs by hand-held dynamometer: A standard protocol
    Bandinelli, S
    Benvenuti, E
    Del Lungo, I
    Baccini, M
    Benvenuti, F
    Di Iorio, A
    Ferrucci, L
    [J]. AGING-CLINICAL AND EXPERIMENTAL RESEARCH, 1999, 11 (05) : 287 - 293
  • [7] Berry Emily T, 2004, Pediatr Phys Ther, V16, P191, DOI 10.1097/01.PEP.0000145932.21460.61
  • [8] MAKE TESTS AND BREAK TESTS OF ELBOW FLEXOR MUSCLE STRENGTH
    BOHANNON, RW
    [J]. PHYSICAL THERAPY, 1988, 68 (02): : 193 - 194
  • [9] INTERRATER RELIABILITY OF HAND-HELD DYNAMOMETRY
    BOHANNON, RW
    ANDREWS, AW
    [J]. PHYSICAL THERAPY, 1987, 67 (06): : 931 - 933
  • [10] HAND-HELD COMPARED WITH ISOKINETIC DYNAMOMETRY FOR MEASUREMENT OF STATIC KNEE EXTENSION TORQUE (PARALLEL RELIABILITY OF DYNAMOMETERS)
    BOHANNON, RW
    [J]. CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1990, 11 (03): : 217 - 222