Initial experience with an implantable hemodynamic monitor

被引:40
|
作者
Steinhaus, DM
Lemery, R
Bresnahan, DR
Handlin, L
Bennett, T
Moore, A
Cardinal, D
Foley, L
Levine, R
机构
[1] UNIV MISSOURI, SCH MED, MID AMER HEART INST, DEPT CARDIOL, KANSAS CITY, MO USA
[2] MEDTRONIC INC, MINNEAPOLIS, MN USA
关键词
monitors; hemodynamic; heart failure; devices; implantable;
D O I
10.1161/01.CIR.93.4.745
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background Measurement of intracardiac hemodynamic parameters has been limited to brief periods in the acute care setting. We developed and evaluated an implantable hemodynamic monitor that is capable of measuring chronic right ventricular oxygen saturation and pulmonary artery pressure. Methods and Results The device consists of an electronic controller placed subcutaneously and two transvenous leads placed in the right ventricle (reflectance oximeter) and pulmonary artery (variable capacitance pressure sensor). Implantation was performed in 10 patients with severe left ventricular dysfunction. Average implant pulmonary artery pressures were systolic, 52 +/- 16 mm Hg; diastolic, 29 +/- 11 mm Hg; and mean, 40 +/- 12 mm Hg. The mean right ventricular oxygen saturation at implant was 51%. Provocative maneuvers, including postural changes, sublingual nitroglycerin, and bicycle exercise, demonstrated expected changes in measured oxygen saturation and pulmonary artery pressures over time. At follow-up of 0.5 to 15.5 months, there were no significant differences between pulmonary artery pressures or oxygen saturation values transmitted from the device and simultaneous measurement with balloon flotation catheters. Four of the pulmonary artery leads dislodged and three demonstrated sensor drift, whereas two of the oxygen saturation sensors failed. Four patients died and four received transplants. Pathological study did not demonstrate injury to the right ventricular outflow tract or pulmonic valve. Conclusions Chronic measurement of hemodynamic parameters in the outpatient setting with implantable sensor technology appears to be feasible. The devices are well tolerated without significant untoward effects, and the sensors generally function well over time, providing reliable information. Clinical usefulness remains to be established.
引用
收藏
页码:745 / 752
页数:8
相关论文
共 50 条
  • [31] Implantable Hemodynamic Monitoring in Ambulatory Heart Failure: Who, When, Why, How?
    Desai, Akshay S.
    CURRENT CARDIOLOGY REPORTS, 2015, 17 (12)
  • [32] Postoperative Imaging Appearance of an Implantable Three-dimensional Bioabsorbable Tissue Marker After Breast Surgery: Initial Experience at a Single Institution
    Chen, Christine
    Dodelzon, Katerina
    Ginter, Paula S.
    Drotman, Michele B.
    Arleo, Elizabeth K.
    JOURNAL OF BREAST IMAGING, 2020, 2 (06) : 561 - 568
  • [33] Implantable Hemodynamic Monitoring in Ambulatory Heart Failure: Who, When, Why, How?
    Akshay S. Desai
    Current Cardiology Reports, 2015, 17
  • [34] An Ultra-Wide Bandwidth with Compact Size Implantable Antenna for Health Monitor Microsystem
    DucDung Nguyen
    Choi, YunSung
    Jang, WonHo
    Seo, Chulhun
    PROCEEDINGS OF 2024 IEEE WIRELESS POWER TECHNOLOGY CONFERENCE AND EXPO, WPTCE, 2024, : 497 - 500
  • [35] The Utility of a Wireless Implantable Hemodynamic Monitoring System in Patients Requiring Mechanical Circulatory Support
    Feldman, David S.
    Moazami, Nader
    Adamson, Philip B.
    Vierecke, Juliane
    Raval, Nir
    Shreenivas, Satya
    Cabuay, Barry M.
    Jimenez, Javier
    Abraham, William T.
    O'Connell, John B.
    Naka, Yoshifumi
    ASAIO JOURNAL, 2018, 64 (03) : 301 - 308
  • [36] Experience in totally implantable venous port catheter: Analysis of 3,000 patients in 12 years
    Yanik, Fazli
    Karamustafaoglu, Yekta Altemur
    Karatas, Adem
    Yoruk, Yener
    TURK GOGUS KALP DAMAR CERRAHISI DERGISI-TURKISH JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2018, 26 (03): : 422 - 428
  • [37] Safety and Efficacy of an Early Home-Based Walking Program After Receipt of an Initial Implantable Cardioverter-Defibrillator
    Lau, Emily T.
    Thompson, Elaine A.
    Burr, Robert L.
    Dougherty, Cynthia M.
    ARCHIVES OF PHYSICAL MEDICINE AND REHABILITATION, 2016, 97 (08): : 1228 - 1236
  • [38] Estimating changes in cardiac output using an implanted hemodynamic monitor in heart failure patients
    Stahlberg, Marcus
    Damgaard, Morten
    Ersgard, David
    Norsk, Peter
    Karamanoglu, Mustafa
    Linde, Cecilia
    Braunschweig, Frieder
    SCANDINAVIAN CARDIOVASCULAR JOURNAL, 2010, 44 (04) : 201 - 208
  • [39] Experience of elderly Spanish men with an implantable cardioverter-defibrillator
    Palacios-Cena, Domingo
    Elena Losa-Iglesias, Marta
    Salvadores-Fuentes, Paloma
    Alonso-Blanco, Cristina
    Fernandez-de-las-Penas, Cesar
    GERIATRICS & GERONTOLOGY INTERNATIONAL, 2011, 11 (03) : 320 - 327
  • [40] Innovation With Experience Using Implantable Left Ventricular Assist Devices
    Ascheim, Deborah D.
    Gelijns, Annetine C.
    Rose, Eric A.
    CIRCULATION-HEART FAILURE, 2009, 2 (01) : 1 - 2