A Hysteresis Current Control based Shunt Current Compensation Scheme for Power Quality Improvement in High Power Radiology Applications

被引:0
作者
Angadi, Shridevi. R. [1 ]
Nanjundaswamy, R. [2 ]
Srinivas, N. R. [3 ]
Murthy, R. S. Ananda [1 ]
机构
[1] SJCE, Dept EEE, Mysore, Karnataka, India
[2] SIT, Dept EEE, Tumkur, Karnataka, India
[3] Skanray Technol, Mysore, Karnataka, India
来源
PROCEEDINGS OF THE 2015 INTERNATIONAL CONFERENCE ON POWER AND ADVANCED CONTROL ENGINEERING (ICPACE) | 2015年
关键词
Hysteresis Current Control; Active Power Filter; THD reduction;
D O I
暂无
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
In this paper, a shunt active power filter scheme is proposed for a three phase system feeding a non-linear load. The system represents the HV generator used in high power radiology equipment. Hysteresis current control is used, which controls the operation such that a sinusoidal current is drawn from the source. The proposed method also reduces the THD, thus improving the power quality. The compensating currents are injected in phase opposition at the point of common contact to cancel out the harmonic currents in ac lines. The Hysteresis current control is used because it is simple and easy to implement. The simulation is carried out using MATLAB-SIMULINK software, whose results are displayed.
引用
收藏
页码:53 / 58
页数:6
相关论文
共 17 条
  • [1] Acordi E. J., 2012, 2012 INT C REN EN PO, P1
  • [2] [Anonymous], 1992, 5191992 IEEE
  • [3] Charles S., 2010, INT J COMPUTER ELECT, V2, P175, DOI DOI 10.7763/IJCEE.2010.V2.132
  • [4] Chen Lezhu, 2010 IEEE C IND EL A, P948
  • [5] Dhanasekaran S., 2014, INT J ADV RES ELECT, V1, P6717
  • [6] Exposto B, 2012 ANN C IEEE IND, P1274
  • [7] Garlapati S, 2012 IND INT C POW E, P1
  • [8] SURVEY OF ACTIVE POWER-LINE CONDITIONING METHODOLOGIES
    GRADY, WM
    SAMOTYJ, MJ
    NOYOLA, AH
    [J]. IEEE TRANSACTIONS ON POWER DELIVERY, 1990, 5 (03) : 1536 - 1542
  • [9] Gyugyi L., 1976 IEEE IAS ANN M, P529
  • [10] Kusuma Latha Y., 2011, INT J ELECT COMPUTER, V1, P31