Monte-Carlo investigation of radiation beam quality of the CRNA neutron irradiator for calibration purposes

被引:8
作者
Mazrou, Hakim [1 ]
Sidahmed, Tassadit [1 ]
Allab, Malika [2 ]
机构
[1] CRNA, Algiers 16000, RP, Algeria
[2] USTHB, Fac Phys, Algiers 16111, Algeria
关键词
Neutron field; Beam quality; Area dosimeter; Ambient dose equivalent; Neutron fluence; MCNP5; SOURCE SPECTRUM; FIELD;
D O I
10.1016/j.apradiso.2010.04.022
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
An irradiation system has been acquired by the Nuclear Research Center of Algiers (CRNA) to provide neutron references for metrology and dosimetry purposes. It consists of an (241)Am-Be radionuclide source of 185 GBq (5 Ci) activity inside a cylindrical steel-enveloped polyethylene container with radially positioned beam channel. Because of its composition, filled with hydrogenous material, which is not recommended by ISO standards, we expect large changes in the physical quantities of primary importance of the source compared to a free-field situation. Thus, the main goal of the present work is to fully characterize neutron field of such special delivered set-up. This was conducted by both extensive Monte-Carlo calculations and experimental measurements obtained by using BF(3) and (3)He based neutron area dosimeters. Effects of each component present in the bunker facility of the Algerian Secondary Standard Dosimetry Laboratory (SSDL) on the energy neutron spectrum have been investigated by simulating four irradiation configurations and comparison to the ISO spectrum has been performed. The ambient dose equivalent rate was determined based upon a correct estimate of the mean fluence to ambient dose equivalent conversion factors at different irradiations positions by means of a 3-D transport code MCNP5. Finally, according to practical requirements established for calibration purposes an optimal irradiation position has been suggested to the SSDL staff to perform, in appropriate manner, their routine calibrations. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1915 / 1921
页数:7
相关论文
共 16 条
[1]  
AEA QSA-TECHNOLOGY/GmBH, 2000, SOURC MAN
[2]   Neutron spectrometry [J].
Alevra, A.V. .
Radioprotection, 1999, 34 (03) :305-333
[3]  
[Anonymous], LAUR031987 LOS AL NA
[4]  
[Anonymous], 2021, ISO8529-2
[5]  
BENMOKHTAR S, 2000, CALIBRATION AREA SUR
[6]   Characteristics of the neutron field of the facility at DIN-UPM [J].
Gallego, E ;
Lorente, A ;
Vega-Carrillo, HR .
RADIATION PROTECTION DOSIMETRY, 2004, 110 (1-4) :73-79
[7]  
*INT COMM RAD PROT, 1995, ICRP PUBL, V74
[8]  
International Organization for Standardization, 2001, 85291 ISO
[9]   Characterisation of the IRSN graphite moderated americium-beryllium neutron field [J].
Lacoste, V ;
Gressier, V ;
Muller, H ;
Lebreton, L .
RADIATION PROTECTION DOSIMETRY, 2004, 110 (1-4) :135-139
[10]   Monte Carlo simulation of the IRSN CANEL/T400 realistic mixed neutron-photon radiation field [J].
Lacoste, V ;
Gressier, V .
RADIATION PROTECTION DOSIMETRY, 2004, 110 (1-4) :123-127