Dual-Energy X-ray Imaging by Simultaneous Integration and Campbelling Readout

被引:0
|
作者
Roessl, Ewald [1 ]
Thran, Axel [1 ]
Martens, Gerhard [1 ]
Istel, Thomas [1 ]
Proksa, Roland [1 ]
Schlomka, Jens-Peter [2 ]
机构
[1] Philips Res Europe Hamburg, Digital Imaging, Rontgenstr 24-26, D-22335 Hamburg, Germany
[2] Philips Res Europe Aachen, X ray Imaging Syst, D-52066 Aachen, Germany
来源
2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC) | 2010年
关键词
dual-energy; computed tomography; Campbell-mode readout; DETECTOR CIRCUITS;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Dual-energy imaging is based on the acquisition of two spectrally distinct attenuation measurements. The two most common techniques are the dual-kV technique and the dual-detector layer technique. More recently, another technique was introduced based on the simultaneous acquisition of photon-counting and current integration data (CIX) by means of a dedicated detector readout ASIC. All of the above methods have certain advantages and disadvantages. While depending on the particular realization (e. g., kVp switching, dual-tube systems in computed tomography), the dual-kV technique suffers from a time-lag between the two acquisitions and as a result from susceptibility to motion. The dual-layer technique usually offers less spectral separation and requires a dedicated detection system. The CIX technique suffers from a reduced dynamic range due to the limiting rate performance in the counting channel. In this paper we introduce yet another dual-energy technique, similar to the CIX, however, without the restrictions coming from the limited count rate. We propose dual-energy imaging based on simultaneous acquisition of the mean (DC) and variance (AC) components of the electrical detector signal, i.e., simultaneous integration-mode readout and Campbell-mode readout (fluctuation mode). The method is based on "Campbell's theorem" which states that the variance of the deviations from the mean detector signal is proportional to the second moment of the incoming x-ray energy spectrum. We compare in simulations the dual-energy performance of the proposed new technique with conventional dual-kV and dual-crystal techniques and present first experimental Campbell-mode CT images obtained from a scintillator crystal coupled to a photomultiplier tube. Moreover, we demonstrate the feasibility of separating iodine from calcium with the novel technique and compare it to the separability achieved with a dual-kV technique.
引用
收藏
页码:2112 / 2115
页数:4
相关论文
共 50 条
  • [41] Stabilizing dual-energy x-ray computed tomography reconstructions using patch-based regularization
    Tracey, Brian H.
    Miller, Eric L.
    INVERSE PROBLEMS, 2015, 31 (10)
  • [42] A dual-energy material decomposition method for high-energy X-ray cargo inspection
    Lee, Jiseoc
    Lee, Yunjeong
    Cho, Seungryong
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2012, 61 (05) : 821 - 824
  • [43] Linear iterative near-field phase retrieval (LIPR) for dual-energy x-ray imaging and material discrimination
    Li, Heyang
    Kingston, Andrew M.
    Myers, Glenn R.
    Beeching, Levi
    Sheppard, Adrian P.
    JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2018, 35 (01) : A30 - A39
  • [44] Use of dual-energy x-ray absorptiometry in non-ruminant nutrition research
    Pomar, Candido
    Kipper, Marcos
    Marcoux, Marcel
    REVISTA BRASILEIRA DE ZOOTECNIA-BRAZILIAN JOURNAL OF ANIMAL SCIENCE, 2017, 46 (07): : 621 - 629
  • [45] Efficient material decomposition method for dual-energy X-ray cargo inspection system
    Lee, Donghyeon
    Lee, Jiseoc
    Min, Jonghwan
    Lee, Byungcheol
    Lee, Byeongno
    Oh, Kyungmin
    Kim, Jaehyun
    Cho, Seungryong
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2018, 884 : 105 - 112
  • [46] X-ray dual-energy calibration based on estimated spectral properties of the experimental system
    Hassler, U
    Garnero, L
    Rizo, P
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) : 1699 - 1712
  • [47] ESTIMATING INTRAABDOMINAL ADIPOSE-TISSUE IN WOMEN BY DUAL-ENERGY X-RAY ABSORPTIOMETRY
    TREUTH, MS
    HUNTER, GR
    KEKESSZABO, T
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1995, 62 (03) : 527 - 532
  • [48] Assessment of the possibilities of the dual-energy X-ray absorptiometry of multicomponent samples with variable content
    N. A. Antropov
    D. A. Karpov
    Yu. Yu. Kryuchkov
    Instruments and Experimental Techniques, 2012, 55 : 608 - 610
  • [49] Assessment of the possibilities of the dual-energy X-ray absorptiometry of multicomponent samples with variable content
    Antropov, N. A.
    Karpov, D. A.
    Kryuchkov, Yu Yu
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 2012, 55 (05) : 608 - 610
  • [50] Feasibility of dual-energy CBCT by spectral filtration of a dual-focus CNT x-ray source
    Li, Boyuan
    Spronk, Derrek
    Luo, Yueting
    Puett, Connor
    Inscoe, Christina R.
    Tyndall, Donald A.
    Lee, Yueh Z.
    Lu, Jianping
    Zhou, Otto
    PLOS ONE, 2022, 17 (02):