Dual-Energy X-ray Imaging by Simultaneous Integration and Campbelling Readout

被引:0
|
作者
Roessl, Ewald [1 ]
Thran, Axel [1 ]
Martens, Gerhard [1 ]
Istel, Thomas [1 ]
Proksa, Roland [1 ]
Schlomka, Jens-Peter [2 ]
机构
[1] Philips Res Europe Hamburg, Digital Imaging, Rontgenstr 24-26, D-22335 Hamburg, Germany
[2] Philips Res Europe Aachen, X ray Imaging Syst, D-52066 Aachen, Germany
来源
2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC) | 2010年
关键词
dual-energy; computed tomography; Campbell-mode readout; DETECTOR CIRCUITS;
D O I
暂无
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Dual-energy imaging is based on the acquisition of two spectrally distinct attenuation measurements. The two most common techniques are the dual-kV technique and the dual-detector layer technique. More recently, another technique was introduced based on the simultaneous acquisition of photon-counting and current integration data (CIX) by means of a dedicated detector readout ASIC. All of the above methods have certain advantages and disadvantages. While depending on the particular realization (e. g., kVp switching, dual-tube systems in computed tomography), the dual-kV technique suffers from a time-lag between the two acquisitions and as a result from susceptibility to motion. The dual-layer technique usually offers less spectral separation and requires a dedicated detection system. The CIX technique suffers from a reduced dynamic range due to the limiting rate performance in the counting channel. In this paper we introduce yet another dual-energy technique, similar to the CIX, however, without the restrictions coming from the limited count rate. We propose dual-energy imaging based on simultaneous acquisition of the mean (DC) and variance (AC) components of the electrical detector signal, i.e., simultaneous integration-mode readout and Campbell-mode readout (fluctuation mode). The method is based on "Campbell's theorem" which states that the variance of the deviations from the mean detector signal is proportional to the second moment of the incoming x-ray energy spectrum. We compare in simulations the dual-energy performance of the proposed new technique with conventional dual-kV and dual-crystal techniques and present first experimental Campbell-mode CT images obtained from a scintillator crystal coupled to a photomultiplier tube. Moreover, we demonstrate the feasibility of separating iodine from calcium with the novel technique and compare it to the separability achieved with a dual-kV technique.
引用
收藏
页码:2112 / 2115
页数:4
相关论文
共 50 条
  • [31] K-edge eliminated material decomposition method for dual-energy X-ray CT
    Zhao, Tiao
    Li, Liang
    Chen, Zhiqiang
    APPLIED RADIATION AND ISOTOPES, 2017, 127 : 231 - 236
  • [32] Single-Shot Quantitative X-ray Imaging from Simultaneous Scatter and Dual Energy Measurements: A Simulation Study
    Wang, Adam S.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [33] Validity and reliability of dual-energy X-ray absorptiometry for the assessment of abdominal adiposity
    Glickman, SG
    Marn, CS
    Supiano, MA
    Dengel, DR
    JOURNAL OF APPLIED PHYSIOLOGY, 2004, 97 (02) : 509 - 514
  • [34] Representation of Object Classification Curve by Dual-Energy X-ray Transmission Technology
    Sun Li-na
    VACUUM TECHNOLOGY AND SURFACE ENGINEERING - PROCEEDINGS OF THE 9TH VACUUM METALLURGY AND SURFACE ENGINEERING CONFERENCE, 2009, : 467 - 471
  • [35] Application of conventional- and dual-energy X-ray tomography in process engineering
    Gehrke, S
    Wirth, KE
    IEEE SENSORS JOURNAL, 2005, 5 (02) : 183 - 187
  • [36] Dynamic material decomposition method for MeV dual-energy X-ray CT
    Zhao, Tiao
    Li, Liang
    Chen, Zhiqiang
    APPLIED RADIATION AND ISOTOPES, 2018, 140 : 55 - 62
  • [37] Grade of fine composite mineral particles by dual-energy X-ray radiography
    Schena, G
    Chiaruttini, C
    Dreossi, D
    Olivo, A
    Pani, S
    INTERNATIONAL JOURNAL OF MINERAL PROCESSING, 2002, 67 (1-4) : 101 - 122
  • [38] Dual-energy CBCT by Spectral Filtration of a Dual-focus CNT X-ray Source
    Li, Boyuan
    Hu, Yuanming
    Inscoe, Christina R.
    Xu, Shuang
    Tyndall, Donald A.
    Lee, Yueh Z.
    Lu, Jianping
    Zhou, Otto
    MEDICAL IMAGING 2024: PHYSICS OF MEDICAL IMAGING, PT 1, 2024, 12925
  • [39] Design and implementation of a dual-energy X-ray imaging system for organic material detection in an airport security application
    Macdonald, RDR
    MACHINE VISION APPLICATIONS IN INDUSTRIAL INSPECTION IX, 2001, 4301 : 31 - 41
  • [40] A dual-energy material decomposition method for high-energy X-ray cargo inspection
    Jiseoc Lee
    Yunjeong Lee
    Seungryong Cho
    Byung-Cheol Lee
    Journal of the Korean Physical Society, 2012, 61 : 821 - 824