Mechanism of cutting balloon angioplasty for in-stent restenosis: An intravascular ultrasound study

被引:8
|
作者
Montorsi, P [1 ]
Galli, S [1 ]
Fabbiocchi, F [1 ]
Loaldi, A [1 ]
Trabattoni, D [1 ]
Grancini, L [1 ]
Cozzi, S [1 ]
Ravagnani, P [1 ]
Parodi, O [1 ]
Bartorelli, AL [1 ]
机构
[1] Univ Milan, Inst Cardiol, Ctr Cardiol Monzino, IRCCS, I-20138 Milan, Italy
关键词
cutting balloon; in-stent restenosis; intravascular ultrasound;
D O I
10.1002/ccd.10191
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
We investigated by intravascular ultrasound (IVUS) the mechanism of action of cutting balloon (CB) angioplasty in patients with in-stent restenosis. Seventy-one consecutive restenotic lesions of 66 patients were studied by quantitative coronary angiography (QCA) and IVUS before, immediately after, and, in 20 cases, at 24-hr time interval after CB. CB was selected according to 1:1 CB-to-stent ratio and inflated at 8 atm for 60-90 sec. Both IVUS planar and volumetric (Simpson's rule, 25 patients) analysis were carried out. IVUS measurements included external elastic membrane area (EEMA), stent area (SA), minimal lumen area (MLA), and restenosis area (RA). Following CB, OCA analysis showed increase of minimal lumen diameter (1.17 +/- 0.46 vs. 2.45 +/- 0.51 mm; P < 0.0001) and decrease of diameter stenosis (64% +/- 13% vs. 21% +/- 9%; P < 0.0001). IVUS measurements showed a significant increase of MLA (2.18 +/- 0.80 vs. 7.31 +/- 1.8 mm(2); p < 0.0001), SA (9.62 +/- 2.6 vs. 10.7 +/- 2.75 mm(2); p < 0.0001), and EEMA (117.27 +/- 5 vs. 18.1 +/- 5 mm(2); P < 0.0001) and a decrease of RA (7.43 +/- 2.63 vs. 3.45 +/- 1.39 mm(2); p < 0.0001). No significant change was observed in the original plaque + media area (7.65 +/- 3 vs. 7.38 +/- 2.9 mm(2); p = NS). Thus, of the total lumen enlargement (5.13 +/- 1.85 mm(2)), 23% was the result of increase in mean SA, whereas 77% was the result of a decrease in mean RA. These changes were associated with a 5% increase in EEMA. IVUS volumetric changes paralleled planar variations. Angiographic and IVUS changes were well maintained at 24 hr. CB enlarges coronary lumen mainly by in-stent tissue reduction associated with a moderate degree of additional stent expansion. Favorable QCA and IVUS acute results are maintained at 24 hr. (C) 2002 Wiley-Liss, Inc.
引用
收藏
页码:166 / 173
页数:8
相关论文
共 50 条
  • [1] Cutting Balloon angioplasty for in-stent restenosis
    Kurbaan, AS
    Foale, RA
    Sigwart, U
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2000, 50 (04) : 480 - 483
  • [2] Intravascular ultrasound-guided balloon angioplasty for treatment of in-stent restenosis
    Sakamoto, T
    Kawarabayashi, T
    Taguchi, H
    Tanaka, A
    Nishida, Y
    Shimada, K
    Yoshikawa, J
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 1999, 47 (03) : 298 - 303
  • [3] Cutting Balloon angioplasty for the treatment of in-stent restenosis
    Albiero, R
    Nishida, T
    Karvouni, E
    Corvaja, N
    Vaghetti, M
    Di Mario, C
    Colombo, A
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2000, 50 (04) : 452 - 459
  • [4] Preliminary experience with cutting balloon angioplasty for iliac artery in-stent restenosis
    Tsetis, Dimitrios
    Belli, Anna Maria
    Morgan, Robert
    Basile, Antonio
    Kostas, Theodoros
    Manousaki, Eirini
    Katsamouris, Asterios
    Gourtsoyiannis, Nicholas
    JOURNAL OF ENDOVASCULAR THERAPY, 2008, 15 (02) : 193 - 202
  • [5] Cutting balloon angioplasty for carotid in-stent restenosis: Case reports and review of the literature
    Shah, Qaisar A.
    Georgiadis, Alexandros L.
    Suri, M. Fareed K.
    Rodriguez, Gustavo J.
    Qureshi, Adnan I.
    JOURNAL OF NEUROIMAGING, 2008, 18 (04) : 428 - 432
  • [6] Soft plaque detected on intravascular ultrasound is the strongest predictor of in-stent restenosis: an intravascular ultrasound study
    Sahara, M
    Kirigaya, H
    Oikawa, Y
    Yajima, J
    Nagashima, K
    Hara, H
    Ogasawara, K
    Aizawa, T
    EUROPEAN HEART JOURNAL, 2004, 25 (22) : 2026 - 2033
  • [7] The challenge of in-stent restenosis: insights from intravascular ultrasound
    Spanos, V
    Stankovic, G
    Tobis, J
    Colombo, A
    EUROPEAN HEART JOURNAL, 2003, 24 (02) : 138 - 150
  • [8] Cutting balloon angioplasty vs. conventional balloon angioplasty in patients receiving intracoronary brachytherapy for the treatment of in-stent restenosis
    Fasseas, P
    Orford, JL
    Lennon, R
    O'Neill, J
    Denktas, AE
    Panetta, CJ
    Berger, PB
    Holmes, DR
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 63 (02) : 152 - 157
  • [9] Comparison of Angiographic Result and Long-Term Outcome in Patients With In-Stent Restenosis Treated With Cutting Balloon or With Scoring Balloon Angioplasty
    Leick, Juergen
    Rheude, Tobias
    Cassese, Salvatore
    Krause, Tobias
    Gjata, Anida
    Saad, Louai
    Lindner, Michael
    Steinbach, Mirjam
    Kastrati, Adnan
    Werner, Nikos
    JOURNAL OF INVASIVE CARDIOLOGY, 2024, 36 (10):
  • [10] Drug-Eluting Balloon Angioplasty for Carotid In-Stent Restenosis
    Liistro, Francesco
    Porto, Italo
    Grotti, Simone
    Ventoruzzo, Giorgio
    Vergallo, Rocco
    Bellandi, Guido
    Bolognese, Leonardo
    JOURNAL OF ENDOVASCULAR THERAPY, 2012, 19 (06) : 729 - 733