Microfluidic packaging of high-density CMOS electrode array for labon-a-chip applications

被引:13
作者
Chung, Jaehoon [1 ]
Hwang, How Yuan [1 ]
Chen, Yu [1 ]
Lee, Tae Yoon [1 ,2 ,3 ]
机构
[1] ASTAR, Inst Microelect, 11 Sci Pk Rd,Singapore Sci Pk 2, Singapore 117685, Singapore
[2] Chungnam Natl Univ, Dept Technol Educ, 99 Daehak Ro, Daejeon 34134, South Korea
[3] Chungnam Natl Univ, Dept Biomed Engn, 99 Daehak Ro, Daejeon 34134, South Korea
基金
新加坡国家研究基金会;
关键词
Microfluidic system; Lab-on-a-chip; CMOS packaging; Microfluidic packaging; Impedance detection; Dielectrophoresis; MICROELECTRODE ARRAY; SYSTEM; MANIPULATION; MICROPARTICLES; FABRICATION; TECHNOLOGY; SEPARATION; SENSORS; DESIGN;
D O I
10.1016/j.snb.2017.07.122
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Effective packaging and integration of microfluidic components with a small electronic chip, such as a complementary metal-oxide-semiconductor (CMOS) chip, is crucial for the ultimate realization of lab-on-a-chip (LOC) devices. Hence, the methods and materials used for the integration should provide good reproducibility, reliability, biocompatibility, and capability for mass production. This paper introduces a new concept that allows the miniaturization of microfluidic packaging and the integration of CMOS chips. The photosensitive polymer material used in this study serves not only as a photoresist to pattern diverse shapes of microfluidic structures at the wafer level, but also as an adhesive to bond the Indium-Tin-Oxide (ITO) coated glass chip to the CMOS die. The patterning process, using conventional photolithography, was demonstrated with a wide range of thicknesses from 10 mu m to 80 mu m, and reliable seamless bonding was achieved with a conventional flip-chip bonder. We also validated that the proposed packaging can be utilized in biological experiments by culturing live cells (MCF-7) for three days and by measuring the auto-fluorescence from the polymer material in the test-vehicles. In addition, the performance of the proposed packaging method was demonstrated by applying it to the microfluidic/CMOS hybrid chip for preliminary electrochemical testing (impedance measurement), and particle entrapment testing (dielectrophoresis experiment). The proposed fabrication procedure is expected to facilitate the wide adoption of CMOS technology in LOC applications. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:542 / 550
页数:9
相关论文
共 33 条
  • [1] CMOS Cell Sensors for Point-of-Care Diagnostics
    Adiguzel, Yekbun
    Kulah, Haluk
    [J]. SENSORS, 2012, 12 (08) : 10042 - 10066
  • [2] Polymer microfluidic devices
    Becker, H
    Locascio, LE
    [J]. TALANTA, 2002, 56 (02) : 267 - 287
  • [3] LOW-TEMPERATURE SI3N4 DIRECT BONDING
    BOWER, RW
    ISMAIL, MS
    ROBERDS, BE
    [J]. APPLIED PHYSICS LETTERS, 1993, 62 (26) : 3485 - 3487
  • [4] CMOS high density electrical impedance biosensor array for tumor cell detection
    Chen, Yu
    Wong, Chee Chung
    Pui, Tze Sian
    Nadipalli, Revanth
    Weerasekera, Roshan
    Chandran, Jegatha
    Yu, Hao
    Rahman, Abdur R. A.
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2012, 173 : 903 - 907
  • [5] Microfluidic system for dielectrophoretic separation based on a trapezoidal electrode array
    Choi, S
    Park, JK
    [J]. LAB ON A CHIP, 2005, 5 (10) : 1161 - 1167
  • [6] Design, Fabrication, and Testing of a Hybrid CMOS/PDMS Microsystem for Cell Culture and Incubation
    Christen, Jennifer Blain
    Andreou, Andreas G.
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2007, 1 (01) : 3 - 18
  • [7] D-Chuadhuri T., 2016, IEEE T BIOMED CIRC S, V10, P1129
  • [8] Packaging commercial CMOS chips for lab on a chip integration
    Datta-Chaudhuri, Timir
    Abshire, Pamela
    Smela, Elisabeth
    [J]. LAB ON A CHIP, 2014, 14 (10) : 1753 - 1766
  • [9] Microelectronic system for high-resolution mapping of extracellular electric fields applied to brain slices
    Frey, U.
    Egert, U.
    Heer, F.
    Hafizovic, S.
    Hierlemann, A.
    [J]. BIOSENSORS & BIOELECTRONICS, 2009, 24 (07) : 2191 - 2198
  • [10] Switch-Matrix-Based High-Density Microelectrode Array in CMOS Technology
    Frey, Urs
    Sedivy, Jan
    Heer, Flavio
    Pedron, Rene
    Ballini, Marco
    Mueller, Jan
    Bakkum, Douglas
    Hafizovic, Sadik
    Faraci, Francesca D.
    Greve, Frauke
    Kirstein, Kay-Uwe
    Hierlemann, Andreas
    [J]. IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2010, 45 (02) : 467 - 482