Fiber-based polarization sensitive spectral domain optical coherence tomography using a fiber optic spectral polarimetry instrument

被引:0
作者
Kim, E [1 ]
Milner, TE [1 ]
机构
[1] Univ Texas, Dept Biomed Engn, Austin, TX 78712 USA
来源
Coherence Domain Optical Methods and Optical Coherence Tomography in Biomedicine IX | 2005年 / 5690卷
关键词
optical coherence tomography; polarization sensitive; spectral domain; spectral polarimeter;
D O I
10.1117/12.591150
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We present a novel fiber-based polarization sensitive spectral domain optical coherence tomography (SD-OCT) system using a fiber optic spectral polarimetry instrument (FOSPI), which provides measurement of depth-resolved polarization state of backscattered. light from a specimen with single optical frequency scanning. By inserting a FOSPI in the detection path of SD-OCT, the full set of Stokes parameters of light backscattered at the specific depth of a specimen can be obtained without any other polarization controlling components in the system and the prior knowledge of the polarization state of the light incident on the sample. The operation principle of the fiber-based polarization sensitive spectral domain optical coherence tomography is demonstrated with a Mica plate as birefringent sample.
引用
收藏
页码:283 / 291
页数:9
相关论文
共 3 条
[1]  
Hecht E., 1987, OPTICS
[2]  
KIM E, IN PRESS OPT COMMUN
[3]   Common-path interferometer for frequency-domain optical coherence tomography [J].
Vakhtin, AB ;
Kane, DJ ;
Wood, WR ;
Peterson, KA .
APPLIED OPTICS, 2003, 42 (34) :6953-6958