Quantifying signal quality in scanning transmission X-ray microscopy

被引:3
|
作者
Watts, Benjamin [1 ]
Finizio, Simone [1 ]
Raabe, Jorg [1 ]
机构
[1] Paul Scherrer Inst, Swiss Light Source, Forschungsstr 111, CH-5232 Villigen, Switzerland
关键词
transmission; STXM; uncertainty; NEXAFS; SPECTROSCOPY; ERROR;
D O I
10.1107/S1600577522004210
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
While the general effects of experimental conditions such as photon flux and sample thickness on the quality of data acquired by scanning transmission X-ray microscopy (STXM) are widely known at a basic level, the specific details are rarely discussed. This leaves the community open to forming misconceptions that can lead to poor decisions in the design and execution of STXM measurements. A formal treatment of the uncertainty and distortions of transmission signals (due to dark counts, higher-order photons and poor spatial or spectral resolution) is presented here to provide a rational basis for the pursuit of maximizing data quality in STXM experiments. While we find an optimum sample optical density of 2.2 in ideal conditions, the distortions considered tend to have a stronger effect for thicker samples and so similar to 1 optical density at the analytical energy is recommended, or perhaps even thinner if significant distortion effects are expected (e.g. lots of higher-order light is present in the instrument). (Note that X-ray absorption calculations based on simple elemental composition do not include near-edge resonances and so cannot accurately represent the spectral resonances typically employed for contrast in STXM.) Further, we present a method for objectively assessing the merits of higher-order suppression in terms of its impact on the quality of transmission measurements that should be useful for the design of synchrotron beamlines.
引用
收藏
页码:1054 / 1064
页数:11
相关论文
共 50 条
  • [31] Qualitative detection of single submicron and nanoparticles in human skin by scanning transmission x-ray microscopy
    Graf, Christina
    Meinke, Martina
    Gao, Qi
    Hadam, Sabrina
    Raabe, Joerg
    Sterry, Wolfram
    Blume-Peytavi, Ulrike
    Lademann, Juergen
    Ruehl, Eckart
    Vogt, Annika
    JOURNAL OF BIOMEDICAL OPTICS, 2009, 14 (02)
  • [32] Probing the electronic structure of graphene sheets with various thicknesses by scanning transmission X-ray microscopy
    Bai, Lili
    Liu, Jinyin
    Zhao, Guanqi
    Gao, Jing
    Sun, Xuhui
    Zhong, Jun
    APPLIED PHYSICS LETTERS, 2013, 103 (25)
  • [33] New online beam intensity synchronous monitoring system in scanning transmission X-ray microscopy
    Jiao, Yuchen
    Zhang, Xiangzhi
    Xu, Zijian
    Yao, Zhen
    Sun, Tianxiao
    Zhang, Yufei
    Zhao, Bo
    Guo, Zhi
    Wang, Yong
    Zhen, Xiangjun
    Liu, Haigang
    Liang, Shasha
    Li, Haitao
    Zhao, Xuanyu
    He, Jian
    Tai, Renzhong
    JOURNAL OF SYNCHROTRON RADIATION, 2025, 32 : 424 - 431
  • [34] Simultaneous soft X-ray transmission and emission microscopy
    Gianoncelli, A.
    Kaulich, B.
    Alberti, R.
    Klatka, T.
    Longoni, A.
    de Marco, A.
    Marcello, A.
    Kiskinova, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2009, 608 (01) : 195 - 198
  • [35] Experimental investigation of beam heating in a soft X-ray scanning transmission X-ray microscope
    Leontowich, Adam F. G.
    Hitchcock, Adam P.
    ANALYST, 2012, 137 (02) : 370 - 375
  • [36] Composition analysis of a polymer electrolyte membrane fuel cell microporous layer using scanning transmission X-ray microscopy and near edge X-ray absorption fine structure analysis
    George, Michael G.
    Wang, Jian
    Banerjee, Rupak
    Bazylak, Aimy
    JOURNAL OF POWER SOURCES, 2016, 309 : 254 - 259
  • [37] A new bend-magnet beamline for scanning transmission X-ray microscopy at the Advanced Light Source
    Warwick, T
    Ade, H
    Kilcoyne, D
    Kritscher, M
    Tylisczcak, T
    Fakra, S
    Hitchcock, A
    Hitchcock, P
    Padmore, H
    JOURNAL OF SYNCHROTRON RADIATION, 2002, 9 : 254 - 257
  • [38] Monitoring the fate of copper nanoparticles in river biofilms using scanning transmission X-ray microscopy (STXM)
    Lawrence, John R.
    Dynes, James J.
    Korber, Darren R.
    Swerhone, George D. W.
    LeppardD, Gary G.
    Hitchcock, Adam P.
    CHEMICAL GEOLOGY, 2012, 329 : 18 - 25
  • [39] Time-of-arrival detection for time-resolved scanning transmission X-ray microscopy imaging
    Finizio, Simone
    Mayr, Sina
    Raabe, Jorg
    JOURNAL OF SYNCHROTRON RADIATION, 2020, 27 : 1320 - 1325
  • [40] Scanning transmission X-ray microscopy of nano structured thin film catalysts for proton-exchange-membrane fuel cells
    Lee, Vincent
    Berejnov, Viatcheslav
    West, Marcia
    Kundu, Sumit
    Susac, Darija
    Stumper, Juergen
    Atanasoski, Radoslav T.
    Debe, Mark
    Hitchcock, Adam P.
    JOURNAL OF POWER SOURCES, 2014, 263 : 163 - 174