Malleostapedotomy Prosthesis Size and Shape: Key Measurements From a Temporal Bone Study

被引:11
作者
Gluth, Michael B. [1 ,2 ,3 ]
Cohen, Mauricio A. [2 ,3 ,4 ]
Friedland, Peter L. [2 ,3 ]
Dornhoffer, John L. [1 ]
Atlas, Marcus D. [2 ,3 ]
机构
[1] Univ Arkansas Med Sci, Dept Otolaryngol Head & Neck Surg, Little Rock, AR 72205 USA
[2] Univ Western Australia, Ear Sci Inst Australia, Perth, WA 6009, Australia
[3] Univ Western Australia, Ear Sci Ctr, Perth, WA 6009, Australia
[4] Univ Chile, Dept Otolaryngol, Clin Las Condes, Santiago, Chile
关键词
Conductive hearing loss; Middle ear; Otosclerosis; Stapedotomy;
D O I
10.1097/MAO.0b013e31824b7882
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Hypothesis: By determining the dimensions of middle ear anatomic relationships pertinent to malleostapedotomy (MS), a simplified, yet optimized scheme for sizing, shaping, and placing a prosthesis can be generated. Methods: Surgical dissection of 20 fresh (nonpreserved) cadaveric temporal bones was undertaken. Needlepoint calipers were used to manually measure pertinent distances between the ossicles. These measurements were then used to calculate the dimensions of anatomic triangles within the middle ear that were applied to the process of MS prosthesis sizing, shaping, and placement. Results: Mean distances were 6.3 mm (range, 5.75-7.0 mm) between the usual MS crimp site and stapedotomy site, 3.6 mm (range, 3.00-4.25 mm) between the crimp site and the lateral edge of the distal incus long process, 4.9 mm (range, 4.50-5.00) between the lateral edge of the distal incus long process and the stapedotomy site, 3.7 mm (range, 3.25-4.00) between the crimp site and the umbo, and 3.3 mm (range, 2.75-3.75) between the umbo and lateral edge of the distal incus long process. Conclusion: A total crimped prosthesis length of at least 7.75 mm is required to accommodate the range of interossicular measurements, yet production of longer prostheses is wise owing to the limited number of temporal bones studied. Furthermore, the use of a single-length prosthesis for all MS cases will not suffice except when cutting a sufficiently long prosthesis to its desired length. Understanding of the ossicular anatomy can be used to bend the MS piston to favorably affect the orientation of the prosthesis barrel into the vestibule.
引用
收藏
页码:518 / 522
页数:5
相关论文
共 9 条
[1]   Malleostapedotomy -: The Marburg experience [J].
Dalchow, C. V. ;
Duenne, A. A. ;
Sesterhenn, A. ;
Teymoortash, A. ;
Werner, J. A. .
OTOSCLEROSIS AND STAPES SURGERY, 2007, 65 :215-221
[2]   Malleostapedotomy in revision surgery for otosclerosis [J].
Fisch, U ;
Acar, GÖ ;
Huber, AM .
OTOLOGY & NEUROTOLOGY, 2001, 22 (06) :776-785
[3]  
Fisch U, 2008, TYMPANOPLASTY MASTOI
[4]   Incus Replacement Malleostapedotomy in Quiescent Chronic Otitis Media With a Mobile Stapes Footplate: An Alternative to TORP in Select Cases [J].
Gluth, Michael B. ;
Motakef, Shahpar ;
Friedland, Peter L. ;
Atlas, Marcus D. .
OTOLOGY & NEUROTOLOGY, 2011, 32 (02) :242-245
[5]  
Häusler R, 2000, FUNCTION AND MECHANICS OF NORMAL, DISEASED AND RECONSTRUCTED MIDDLE EARS, P261
[6]   Morphology of the Malleus Handle and the Comparison of Different Prostheses for Malleostapedotomy [J].
Kwok, Pingling ;
Fisch, Ugo ;
Nussbaumer, Markus ;
Herkenhoff, Sven ;
Strutz, Juergen .
OTOLOGY & NEUROTOLOGY, 2009, 30 (08) :1175-1185
[7]   MALLEUS GRIP PROSTHESIS [J].
SCHUKNECHT, HF ;
BARTLEY, ML .
ANNALS OF OTOLOGY RHINOLOGY AND LARYNGOLOGY, 1986, 95 (05) :531-534
[8]  
SHEA JJ, 1983, LARYNGOSCOPE, V93, P989
[9]  
SHEEHY JL, 1982, LARYNGOSCOPE, V92, P258