Wear Analysis of Highly Cross-Linked Polyethylene in Young and Active Patients at Average Fourteen Years: A Concise Follow-Up of a Previous Report

被引:15
作者
Gaudiani, Michael A. [1 ]
Ranawat, Amar S. [1 ]
Ranawat, Chitranjan S. [1 ]
机构
[1] Hosp Special Surg, Dept Orthoped Surg, 535 East 70th St, New York, NY 10021 USA
关键词
highly cross-linked polyethylene; total hip arthroplasty; total hip liners; young and active; wear analysis; TOTAL HIP-ARTHROPLASTY; CLINICAL-PERFORMANCE; CONVENTIONAL POLYETHYLENE; DOUBLE-BLIND; OSTEOLYSIS; ASSOCIATION; COMPONENTS; OXIDATION; OUTCOMES;
D O I
10.1016/j.arth.2017.09.007
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Concerns persist regarding the oxidative stability of annealed highly cross-linked polyethylene (HXLPE) and hence its wear performance especially long term in young and active patients. The purpose of this study is to assess the wear rates and clinical outcomes of a first-generation annealed HXLPE. Methods: Forty-six patients (57 hips) with noncemented fixation, a mean age of 53 years, and a mean University of California Los Angeles activity score of 8 underwent primary total hip arthroplasty. A 28-mm femoral head against HXLPE liner was used in all cases. Linear wear rates were measured by 2 independent observers. Radiological and clinical assessments were recorded. Results: At a mean of 14 years (range 10-17), the wear rate was 0.03 mm/y (SD +/- 0.02). There were no revisions for wear-related complications. Using revision for wear-related complications as an endpoint, the survivorship was 100% and we observed no osteolysis. Western Ontario and McMaster Universities Osteoarthritis Index, patient administered questionnaire, and Hospital for Special Surgery hip scores were 91/100, 5/100, and 37/40 respectively. Conclusion: This study confirms that a first-generation annealed HXLPE shows excellent wear and clinical results at 10-17 years in young and active patients. Oxidative degradation of an annealed HXLPE does not adversely affect wear and shows no other wear-related complications. (C) 2017 Elsevier Inc. All rights reserved.
引用
收藏
页码:586 / 589
页数:4
相关论文
共 37 条
  • [1] AMSTUTZ HC, 1978, CLIN ORTHOP RELAT R, P87
  • [2] BELLAMY N, 1988, J RHEUMATOL, V15, P1833
  • [3] A Double-Blind, Prospective, Randomized Controlled Trial Comparing Highly Cross-Linked and Conventional Polyethylene in Primary Total Hip Arthroplasty
    Calvert, George T.
    Devane, Peter A.
    Fielden, Jan
    Adams, Kathryn
    Horne, J. Geoffrey
    [J]. JOURNAL OF ARTHROPLASTY, 2009, 24 (04) : 505 - 510
  • [4] Continued Improved Wear with an Annealed Highly Cross-linked Polyethylene
    Capello, William N.
    D'Antonio, James A.
    Ramakrishnan, Rama
    Naughton, Marybeth
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2011, 469 (03) : 825 - 830
  • [5] Five-year experience with Crossfire® highly cross-linked polyethylene
    D'Antonio, JA
    Manley, MT
    Capello, WN
    Bierbaum, BE
    Ramakrishnan, R
    Naughton, M
    Sutton, K
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (441) : 143 - 150
  • [6] DELEE JG, 1976, CLIN ORTHOP RELAT R, P20
  • [7] Clinical performance of a durasul highly cross-linked polyethylene acetabular liner for total hip arthroplasty at five years
    Dorr, LD
    Wan, ZN
    Shahrdar, C
    Sirianni, L
    Boutary, M
    Yun, A
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2005, 87A (08) : 1816 - 1821
  • [8] Characterization of long-term femoral-head-penetration rates - Association with and prediction of osteolysis
    Dowd, JE
    Sychterz, CJ
    Young, AM
    Engh, CA
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2000, 82A (08) : 1102 - 1107
  • [9] A literature review of the association between wear rate and osteolysis in total hip arthroplasty
    Dumbleton, JH
    Manley, MT
    Edidin, AA
    [J]. JOURNAL OF ARTHROPLASTY, 2002, 17 (05) : 649 - 661
  • [10] The creep and wear of highly cross-linked polyethylene - A three-year randomised, controlled trial using radiostereometric analysis
    Glyn-Jones, S.
    McLardy-Smith, P.
    Gill, H. S.
    Murray, D. W.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2008, 90B (05): : 556 - 561