Elimination of sulfide interference by sodium hypochlorite solution in the cold vapor atomic absorption spectrometric determination of mercury using tin(II) reduction in alkaline medium

被引:3
|
作者
Kagaya, S [1 ]
Nishimura, T [1 ]
Hasegawa, K [1 ]
机构
[1] Toyama Univ, Fac Engn, Dept Chem & Biochem Engn, Gofuku, Toyama 9308555, Japan
来源
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY | 1999年 / 363卷 / 01期
关键词
Sodium; Sulfide; Mercury; Atomic Absorption; Sample Solution;
D O I
10.1007/s002160051149
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
An addition of sodium hypochlorite solution effectively eliminated the interference by sulfide in the cold vapor atomic absorption spectrometric determination of mercury using tin(II) reduction in alkaline medium. Mercury ranging from 0.01 to 0.5 mu g could be determined within +/-10% error in 100 mt of sample solution containing 10 mg of sulfide by the addition of 1 mL of 10.4%(w/v) sodium hypochlorite solution. The proposed method is rapid (5 min per determination) and has good reproducibility (3.0%).