Time-dependent fixation and implantation forces for a femoral knee component-An in vitro study

被引:8
作者
Burgers, Travis [1 ]
Mason, Jim [2 ]
Squire, Matthew [3 ]
Ploeg, Heidi-Lynn [1 ,4 ]
机构
[1] Univ Wisconsin, Dept Mech Engn, Madison, WI 53706 USA
[2] Van Andel Inst, Lab Orthopaed Cell & Tissue Mech, Grand Rapids, MI USA
[3] Univ Wisconsin, Dept Orthoped & Rehabil, Madison, WI USA
[4] Univ Wisconsin, Dept Biomed Engn, Madison, WI USA
关键词
Distal femur; Femoral knee component; Long term fixation; Implant loosening; Viscoelasticity; In vitro testing; Impact forces; Implantation; Press fit; Cementless; Biomechanics; CANCELLOUS BONE; TRABECULAR BONE; SKULL FRACTURE; CORTICAL BONE; HEAD; ARTHROPLASTY; BIOMECHANICS; BEHAVIOR;
D O I
10.1016/j.medengphy.2010.06.011
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Implant survival rate is a primary concern for individuals receiving a primary total knee arthroplasty Loosening is the primary reason for revision surgery and was therefore the focus of the current study To better understand the mechanics of implant fixation the time-dependent fixation of a femoral knee component was measured in vitro on three cadaveric femurs The fixation of each femoral knee component was measured with strain gauged implants for at least 10 min on each femoral component Additionally impaction forces were measured during the implantation of each component These forces were 2-6 times less than previously reported The implantation impact forces were higher for the bones with higher bone density Power law regressions were fit to the absolute value of the principal strains measured on the components over time to quantify the relaxation of the bone The average power coefficient value for the three bones was lower for the bones with higher bone density The average power coefficient value for the maximum principal strains was significantly higher than that of the minimum principal strains in each bone The results were extrapolated to approximate the fixation strength at 9 months after implantation In this time period the strain was predicted to decrease to between 78 and 91% of the strain 1 s after implantation where those with lower bone density will have decreased fixation strength (C) 2010 IPEM Published by Elsevier Ltd All rights reserved
引用
收藏
页码:968 / 973
页数:6
相关论文
共 25 条
[1]   Dissociation of modular humeral head components: A biomechanical and implant retrieval study [J].
Blevins, FT ;
Deng, XH ;
Torzilli, PA ;
Dines, D ;
Warren, RF .
JOURNAL OF SHOULDER AND ELBOW SURGERY, 1997, 6 (02) :113-124
[2]   The effect of damage on the viscoelastic behavior of human vertebral trabecular bone [J].
Bredbenner, Todd L. ;
Davy, Dwight T. .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 2006, 128 (04) :473-480
[3]  
BROWN T, 1994, 40 ANN ORTH RES SOC
[4]  
Burgers Travis A., 2009, International Journal of Experimental and Computational Biomechanics, V1, P23, DOI 10.1504/IJECB.2009.022857
[5]   Post-yield relaxation behavior of bovine cancellous bone [J].
Burgers, Travis A. ;
Lakes, Roderic S. ;
Garcia-Rodriguez, Sylvana ;
Piller, Geoffrey R. ;
Ploeg, Heidi-Lynn .
JOURNAL OF BIOMECHANICS, 2009, 42 (16) :2728-2733
[6]   STRESS-RELAXATION BEHAVIOR OF TRABECULAR BONE SPECIMENS [J].
DELIGIANNI, DD ;
MARIS, A ;
MISSIRLIS, YF .
JOURNAL OF BIOMECHANICS, 1994, 27 (12) :1469-1476
[7]   Biomechanics of frontal skull fracture [J].
Delye, Hans ;
Verschueren, Peter ;
Depreitere, Bart ;
Verpoest, Ignaas ;
Berckmans, Daniel ;
Sloten, Jos Vander ;
Van der Perre, Georges ;
Goffin, Jan .
JOURNAL OF NEUROTRAUMA, 2007, 24 (10) :1576-1586
[8]   Lateral head impacts and protection of the temporal area by bicycle safety helmets [J].
Depreitere, Bart ;
Van Lierde, Carl ;
Vander Sloten, Jos ;
Van der Perre, Georges ;
Van Audekercke, Remy ;
Plets, Christiaan ;
Goffin, Jan .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 2007, 62 (06) :1440-1445
[9]   Progression of human bone ingrowth into porous-coated implants - Rate of bone ingrowth in humans [J].
Hofmann, AA ;
Bloebaum, RD ;
Bachus, KN .
ACTA ORTHOPAEDICA SCANDINAVICA, 1997, 68 (02) :161-166
[10]   In vivo skeletal responses to porous-surfaced implants subjected to small induced motions [J].
Jasty, M ;
Bragdon, C ;
Burke, D ;
OConnor, D ;
Lowenstein, J ;
Harris, WH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1997, 79A (05) :707-714