Asymptomatic right ventricular perforation by an implantable cardioverter defibrillator lead detected by home monitoring system

被引:7
作者
Migliore, Federico [1 ]
Leoni, Loira [1 ]
Torregrossa, Gianluca [1 ]
Guglielmi, Cosimo [1 ]
Tarantini, Giuseppe [1 ]
Buja, Gianfranco [1 ]
Iliceto, Sabino [1 ]
Corrado, Domenico [1 ]
机构
[1] Univ Padua, Dept Cardiac Thorac & Vasc Sci, Padua, Italy
关键词
Right ventricular perforation; Implantable cardioverter defibrillator (ICD) lead perforation; Home monitoring (HM) system; Brugada syndrome; ICD LEAD; PACEMAKER;
D O I
10.1016/j.jelectrocard.2010.07.022
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Right ventricular lead perforation is a rare but serious and potentially life-threatening complication of pacemaker or defibrillator lead implantation. This report describes a patient with Brugada syndrome in whom the diagnosis of asymptomatic right ventricular perforation by an implantable cardioverter defibrillator lead was detected 12 days after implantation, thanks to a report from home monitoring system. The patient was admitted to our institution, where the lead was explanted and replaced. This case illustrates the potential lifesaving benefit of the home monitoring system in patients with implantable cardioverter defibrillator. (C) 2010 Elsevier Inc. All rights reserved.
引用
收藏
页码:673 / 675
页数:3
相关论文
共 11 条
  • [1] Increased incidence of subacute lead perforation noted with one implantable cardioverter-defibrillator
    Danik, Stephan B.
    Mansour, Moussa
    Singh, Jagmeet
    Reddy, Vivek Y.
    Ellinor, Patrick T.
    Milan, David
    Heist, E. Kevin
    d'Avila, Andre
    Ruskin, Jeremy N.
    Mela, Theofanie
    [J]. HEART RHYTHM, 2007, 4 (04) : 439 - 442
  • [2] Fisher JD, 2008, PACE, V31, P7, DOI 10.1111/j.1540-8159.2007.00918.x
  • [3] GREYFMAN G, 2007, PACING CLIN ELECTROP, V30, P498
  • [4] Hirschl DA, 2007, PACE, V30, P28
  • [5] Delayed lead perforation: A disturbing trend
    Khan, MN
    Joseph, G
    Khaykin, Y
    Ziada, KM
    Wilkoff, BL
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2005, 28 (03): : 251 - 253
  • [6] Home monitoring remote control of pacemaker and implantable cardioverter defibrillator patients in clinical practice: impact on medical management and health-care resource utilization
    Ricci, Renato Pietro
    Morichetli, Loredana
    Santini, Massimo
    [J]. EUROPACE, 2008, 10 (02): : 164 - 170
  • [7] Satpathy R, 2008, PACE, V31, P10, DOI 10.1111/j.1540-8159.2007.00919.x
  • [8] SCHOLTEN MF, 2004, PACING CLIN ELECTROP, V27, P11512
  • [9] Remote detection of incessant slow VT with an ICD capable of home monitoring
    Siaplaouras, Stephanos
    Buob, Axel
    Neuberger, Hans Ruprecht
    Mewis, Christian
    [J]. EUROPACE, 2006, 8 (07): : 512 - 514
  • [10] Severe pacemaker lead perforation detected by an automatic home-monitoring system
    Spencker, Sebastian
    Mueller, Dirk
    Marek, Andrea
    Zabel, Markus
    [J]. EUROPEAN HEART JOURNAL, 2007, 28 (12) : 1432 - 1432