Optical molecular imaging-guided radiation therapy part 1: Integrated x-ray and bioluminescence tomography

被引:21
作者
Shi, Junwei [1 ]
Udayakumar, Thirupandiyur S. [1 ]
Wang, Zhiqun [1 ]
Dogan, Nesrin [1 ]
Pollack, Alan [1 ]
Yang, Yidong [1 ]
机构
[1] Univ Miami, Sch Med, Dept Radiat Oncol, Miami, FL 33136 USA
关键词
bioluminescence tomography; cone-beam computed tomography; in vivo mice experiments; orthotopic prostate tumor; radiation guidance; SYSTEM; GUIDANCE; ALGORITHMS; QUALITY;
D O I
10.1002/mp.12415
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: X-ray CT faces challenges in differentiating tumors from surrounding healthy tissues. A bioluminescence tomography (BLT) system which can directly reconstruct the internal luminescent tumors, was developed and integrated with CT system to accurately guide radiation close delivery. Methods: The BLT system, employing a lens-coupled CCD camera, was physically registered with an onboard cone beam CT system in an image-guided small animal arc radiation treatment system (iSMAART). The onboard CT provides animal anatomy and accurate surface contour used to construct the three-dimensional mesh for the BLT reconstruction. Bioluminescence projections were captured from multiple angles, once every 90 degrees rotation. The BLT reconstruction was performed on an orthotopic prostate tumor model to evaluate its robustness and accuracy in locating and delineating bioluminescent tumors. The location and volume of the tumor identified from iodinated contrast CT was used to validate the BLT performance. Phantom experiment was also conducted to confirm the precision of BLT-guided radiation. Results: The BLT was able to accurately locate the bioluminescent tumors with < 0.5 mm error. The tumor volume in BLT was significantly correlated with that in the iodinated contrast CT (R-2 = 0.81, P < 0.001). Phantom experiments further validated that BLT can be used to guide radiation with sub millimeter accuracy. Conclusion: Together with CT, BLT can provide precision radiation guidance and robust minor volume assessment in small animal cancer research. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:4786 / 4794
页数:9
相关论文
共 36 条
[1]   Tomographic bioluminescence imaging by use of a combined optical-PET (OPET) system: a computer simulation feasibility study [J].
Alexandrakis, G ;
Rannou, FR ;
Chatziioannou, AF .
PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (17) :4225-4241
[2]   CT Guidance is Needed to Achieve Reproducible Positioning of the Mouse Head for Repeat Precision Cranial Irradiation [J].
Armour, M. ;
Ford, E. ;
Iordachita, I. ;
Wong, J. .
RADIATION RESEARCH, 2010, 173 (01) :119-123
[3]   Optical tomography in medical imaging [J].
Arridge, SR .
INVERSE PROBLEMS, 1999, 15 (02) :R41-R93
[4]   Tumor imaging in small animals with a combined micro-CT/micro-DSA system using iodinated conventional and blood pool contrast agents [J].
Badea, Cristian T. ;
Hedlund, Laurence W. ;
De Lin, Ming ;
Mackel, Julie F. Boslego ;
Johnson, G. Allan .
CONTRAST MEDIA & MOLECULAR IMAGING, 2006, 1 (04) :153-164
[5]   Tracer kinetic modelling of tumour angiogenesis based on dynamic contrast-enhanced CT and MRI measurements [J].
Brix, Gunnar ;
Griebel, Juergen ;
Kiessling, Fabian ;
Wenz, Frederik .
EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2010, 37 :S30-S51
[6]   Characterization of image quality and image-guidance performance of a preclinical microirradiator [J].
Clarkson, R. ;
Lindsay, P. E. ;
Ansell, S. ;
Wilson, G. ;
Jelveh, S. ;
Hill, R. P. ;
Jaffray, D. A. .
MEDICAL PHYSICS, 2011, 38 (02) :845-856
[7]   A Born-type approximation method for bioluminescence tomography [J].
Cong, WX ;
Durairaj, K ;
Wang, LV ;
Wang, G .
MEDICAL PHYSICS, 2006, 33 (03) :679-686
[8]   Advances in vivo bioluminescence imaging of gene expression [J].
Contag, CH ;
Bachmann, MH .
ANNUAL REVIEW OF BIOMEDICAL ENGINEERING, 2002, 4 :235-260
[9]   Small animal fluorescence and bioluminescence tomography: a review of approaches, algorithms and technology update [J].
Darne, Chinmay ;
Lu, Yujie ;
Sevick-Muraca, Eva M. .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (01) :R1-R64
[10]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619