Patient dose from 3D rotational neurovascular studies

被引:31
作者
Bridcut, R. R.
Murphy, E.
Workman, A.
Flynn, P.
Winder, R. J.
机构
[1] Forster Green hosp, No Ireland Reg Med Phys Agcy, Belfast BT8 6HD, Antrim, North Ireland
[2] Royal Victoria Hosp, Dept Neuroradiol, Belfast BT12 6BA, Antrim, North Ireland
[3] Univ Ulster, Hlth & Rehabil Res Inst, Newtonabbey BT37 0QB, Ireland
关键词
D O I
10.1259/bjr/95349672
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The use of image-guided interventional radiological techniques is increasing in prevalence and complexity. Imaging system developments have helped improve the information available to interventionalists to plan and guide procedures. Information on doses to patients resulting from alternative imaging techniques or protocols is useful for both the process of justifying particular procedures and in optimizing the resultant exposures. Such information is not always available, especially for new or developing imaging techniques. We have undertaken a study of doses to patients associated with two alternative imaging methods for pre-intervention assessment of intracranial aneurysms. In the first technique the aneurysm is assessed from a series of digital subtraction angiography (DSA) runs taken at different imaging projections. The second technique involved acquiring images from one single image run while the imaging system rotated 180 degrees around the patient's head. In this technique, the aneurysm was then evaluated from a 3D reconstruction of the projection images. Effective doses were calculated using a computer model to simulate the exposure geometry and parameters. The mean dose from the DSA protocol used at our centre was 3.4 mSv and from the 3D rotational angiography (RA) technique was 0.20 mSv.
引用
收藏
页码:362 / 366
页数:5
相关论文
共 15 条
  • [1] [Anonymous], 1980, NUREGCR1159
  • [2] Intracranial aneurysms: Clinical value of 3D digital subtraction angiography in the therapeutic decision and endovascular treatment
    Anxionnat, R
    Bracard, S
    Ducrocq, X
    Trousset, Y
    Launay, L
    Kerrien, E
    Braun, M
    Vaillant, R
    Scomazzoni, F
    Lebedinsky, A
    Picard, L
    [J]. RADIOLOGY, 2001, 218 (03) : 799 - 808
  • [3] Cranley K., 1997, CATALOGUE DIAGNOSTIC
  • [4] Cristy M, 1980, MATH PHANTOMS REPRES
  • [5] PRACTICAL CONE-BEAM ALGORITHM
    FELDKAMP, LA
    DAVIS, LC
    KRESS, JW
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06): : 612 - 619
  • [6] ELECTROTHROMBOSIS OF SACCULAR ANEURYSMS VIA ENDOVASCULAR APPROACH .1. ELECTROCHEMICAL BASIS, TECHNIQUE, AND EXPERIMENTAL RESULTS
    GUGLIELMI, G
    VINUELA, F
    SEPETKA, I
    MACELLARI, V
    [J]. JOURNAL OF NEUROSURGERY, 1991, 75 (01) : 1 - 7
  • [7] HART D, 1994, NRPBR262
  • [8] Patient size and x-ray technique factors in head computed tomography examinations. I. Radiation doses
    Huda, W
    Lieberman, KA
    Chang, J
    Roskopf, ML
    [J]. MEDICAL PHYSICS, 2004, 31 (03) : 588 - 594
  • [9] International Commission on Radiological Protection, 1990, ICRP PUBL, V60, P21
  • [10] PATIENT AND STAFF DOSIMETRY IN NEURORADIOLOGICAL PROCEDURES
    MARSHALL, NW
    NOBLE, J
    FAULKNER, K
    [J]. BRITISH JOURNAL OF RADIOLOGY, 1995, 68 (809) : 495 - 501