Measurement of dose given by Co-60 in radiotherapy with TLD-500

被引:6
作者
Tanir, Gunes [1 ]
Cengiz, Ferhat [1 ]
Bolukdemir, M. Hicabi [1 ]
机构
[1] Gazi Univ, Fac Sci, TR-06500 Ankara, Turkey
关键词
Optically stimulated luminescence; Dose measurements; Percentage depth dose; TLD-500; Medical dosimeter;
D O I
10.1016/j.radphyschem.2011.12.001
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
The uses of dosimeters based on optically stimulated luminescence technique have become widespread in clinical applications. In the present study, the dose values given by Cobalt-60 radiotherapy machine were measured with optically stimulated luminescence (OSL) technique using TLD-500 and compared with those of commonly used ionization chamber dosimeter system. The percentage depth dose (DD%) values and graphs were formed. OSL system with TLD-500 can be reliably used as medical and personal dosimeter. (C) 2011 Elsevier Ltd. All rights reserved.