Comparing signal-based,and case-based methodologies for CAD assessment in a detection task

被引:1
作者
Paquerault, Sophie [1 ]
Samuelson, Frank W. [1 ]
Petrick, Nicholas [1 ]
Myers, Kyle J. [1 ]
机构
[1] US FDA, CDRH Lab Assessment Med Imaging Syst, NIBIB, Rockville, MD 20857 USA
来源
MEDICAL IMAGING 2008: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT | 2008年 / 6917卷
关键词
computer-aided detection; multiple reader multiple case (MRMC) analysis; free-response receiver operating characteristics; receiver operating characteristics;
D O I
10.1117/12.771498
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We are investigating the potential for differences in study conclusions when assessing the estimated impact of a computer-aided detection (CAD) system on readers' performance using a signal-based performance analysis derived from Free-response Receiver Operating Characteristics (FROC) versus a case-based performance analysis derived from Receiver Operating Characteristics (ROC) analysis. To consider this question, we utilized reader data from a CAD assessment study based on 100 mammographic background images to which fixed-size and fixed-intensity Gaussian signals were added, generating a low- and high-intensity set. The study thus allowed CAD assessment in two situations: when CAD sensitivity was 1) superior or 2) equivalent or lower than the average reader. Seven readers were asked to review each set using CAD in both second-reader and concurrent modes. Signal-based detection results were analyzed using the area under the FROC curve below 0.5 false positives per image. Case-based decision results were analyzed using the area under the parametric ROC curve. The results were consistent between the signal-based and case-based analyses for the low-intensity set, suggesting that CAD in both reading modes can increase reader signal-based detection and case-based decision accuracies. For the high-intensity set, the signal-based and case-based analysis suggested different conclusions regarding the utility of CAD, although neither analysis resulted in statistical significance.
引用
收藏
页数:9
相关论文
共 19 条
[1]   Variability in the interpretation of screening mammograms by US radiologists - Findings from a national sample [J].
Beam, CA ;
Layde, PM ;
Sullivan, DC .
ARCHIVES OF INTERNAL MEDICINE, 1996, 156 (02) :209-213
[2]   Improvement in sensitivity of screening mammography with computer-aided detection: A multiinstitutional trial [J].
Brem, RF ;
Baum, J ;
Lechner, M ;
Kaplan, S ;
Souders, S ;
Naul, LG ;
Hoffmeister, J .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2003, 181 (03) :687-693
[3]  
Burhenne LJW, 2000, RADIOLOGY, V215, P554
[4]  
Chakraborty D. P., 2000, HDB MED IMAGING, P771
[5]   Observer studies involving detection and localization: Modeling, analysis, and validation [J].
Chakraborty, DP ;
Berbaum, KS .
MEDICAL PHYSICS, 2004, 31 (08) :2313-2330
[6]  
Efron B., 1986, STAT SCI, V1, P54, DOI [DOI 10.1214/SS/1177013815, 10.1214/ss/1177013815]
[7]   Influence of computer-aided detection on performance of screening mammography [J].
Fenton, Joshua J. ;
Taplin, Stephen H. ;
Carney, Patricia A. ;
Abraham, Linn ;
Sickles, Edward A. ;
D'Orsi, Carl ;
Berns, Eric A. ;
Cutter, Gary ;
Hendrick, R. Edward ;
Barlow, William E. ;
Elmore, Joann G. .
NEW ENGLAND JOURNAL OF MEDICINE, 2007, 356 (14) :1399-1409
[8]   Screening mammography with computer-aided detection: Prospective study of 12,860 patients in a community breast center [J].
Freer, TW ;
Ulissey, MJ .
RADIOLOGY, 2001, 220 (03) :781-786
[9]   THE EFFICACY OF DIAGNOSTIC-IMAGING [J].
FRYBACK, DG ;
THORNBURY, JR .
MEDICAL DECISION MAKING, 1991, 11 (02) :88-94
[10]   Changes in breast cancer detection and mammography recall rates after the introduction of a computer-aided detection system [J].
Gur, D ;
Sumkin, JH ;
Rockette, HE ;
Ganott, M ;
Hakim, C ;
Hardesty, L ;
Poller, WR ;
Shah, R ;
Wallace, L .
JOURNAL OF THE NATIONAL CANCER INSTITUTE, 2004, 96 (03) :185-190